Atellica CH UCFP 11097543Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Atellica CH UCFP Reagent · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 00630414279206), received between Mar 10, 2023 and Jul 31, 2026 — classified as 19 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (19 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 11, 2025Class IIOpen, Classified
Potential for falsely depressed patient, quality control (QC), and/or calibration results or delayed results.
Z-2542-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 29, 2024Class IIOpen, Classified
The potential for biased quality control (QC) and patient results when using Atellica CH UCFP lot 130414 and ADVIA Chemistry UCFP lot 140 reagents.
Z-3063-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Low Test Results19
Patient problems
- No Clinical Signs, Symptoms or Conditions19
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 31, 2026Low Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50097973 ↗MalfunctionOct 15, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED TO SIEMENS THEY OBTAINED AN ERRONEOUSLY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULT ON ONE (1) PATIENT SAMPLE ON THE ATELLICA CH 930 ANALYZER. THE ERRONEOUSLY DEPRESSED RESULT WAS REPORTED TO THE PHYSICIAN AND NOT QUESTIONED. THE SAME SAMPLE WAS REPROCESSED ON THE ORIGINAL ATELLICA CH 930 ANALYZER. A HIGHER RESULT WAS OBTAINED, CONSIDERED CORRECT, AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) PATIENT RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23279025 ↗MalfunctionOct 9, 2024Low Test Results; Low Test Results
THE CUSTOMER CONTACTED SIEMENS REGARDING ONE (1) ERRONEOUSLY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) PATIENT SAMPLE RESULT OBTAINED ON AN ATELLICA ® CH 930 ANALYZER. THE ERRONEOUSLY DEPRESSED PATIENT RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED ON AN ALTERNATE ATELLICA ® CH 930 ANALYZER. A HIGHER RESULT WAS OBTAINED, CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUSLY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20410112 ↗MalfunctionJun 25, 2024Low Test Results; Low Test Results
THE CUSTOMER REPORTED FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS WERE OBTAINED ON FOUR (4) PATIENT SAMPLES ON AN ATELLICA CH 930 INSTRUMENT (S/N (B)(6) ). THE ERRONEOUS RESULTS WERE REPORTED TO THE PHYSICIAN(S) AND WERE NOT QUESTIONED. THE SAME SAMPLES WERE REPROCESSED ON AN ALTERNATE ATELLICA CH 930 INSTRUMENT (S/N (B)(6) ). THE REPROCESSED HIGHER RESULTS OBTAINED WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19605916 ↗MalfunctionJun 17, 2024Low Test Results; Low Test Results
THE CUSTOMER REPORTED FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS WERE OBTAINED ON ONE (1) PATIENT SAMPLE ON AN ATELLICA CH 930 INSTRUMENT S/N (B)(6). THE ERRONEOUS RESULT WAS REPORTED TO THE PHYSICIAN AND WAS NOT QUESTIONED. THE SAME SAMPLE WAS REPROCESSED ON THE INITIAL ATELLICA CH 930 INSTRUMENT WITH AN ALTERNATE REAGENT LOT. A HIGHER REPROCESSED RESULT WAS OBTAINED, CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUS DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19549865 ↗MalfunctionJun 17, 2024Low Test Results; Low Test Results
THE CUSTOMER REPORTED FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS WERE OBTAINED ON TWO (2) PATIENT SAMPLES ON AN ATELLICA CH 930 INSTRUMENT (S/N (B)(6)). THE ERRONEOUS RESULTS WERE REPORTED TO THE PHYSICIAN AND WERE NOT QUESTIONED. THE SAME SAMPLES WERE REPROCESSED ON THE INITIAL ATELLICA CH 930 INSTRUMENT WITH AN ALTERNATE REAGENT LOT. THE REPROCESSED SAMPLES OBTAINED HIGHER UCFP RESULTS THAT WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUS DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19549052 ↗MalfunctionJun 17, 2024Low Test Results; Low Test Results
THE CUSTOMER REPORTED FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS WERE OBTAINED ON TWO (2) PATIENT SAMPLES ON AN ATELLICA CH 930 INSTRUMENT S/N (B)(6). THE ERRONEOUS RESULTS WERE REPORTED TO THE PHYSICIAN AND WERE NOT QUESTIONED. THE SAME SAMPLES WERE REPROCESSED ON THE INITIAL ATELLICA CH 930 INSTRUMENT WITH AN ALTERNATE REAGENT LOT. THE REPROCESSED SAMPLES OBTAINED HIGHER UCFP RESULTS THAT WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUS DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19548469 ↗MalfunctionMay 2, 2024Low Test Results; Low Test Results
THE CUSTOMER REPORTED A DISCORDANT DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CH 930 INSTRUMENT. THE DISCORDANT RESULT WAS REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULT. THE SAMPLE WAS REPEATED ON AN ALTERNATE ATELLICA CH 930 INSTRUMENT. THE REPEAT RESULT WAS THE SAME AS THE DISCORDANT RESULT. THE PATIENT SAMPLE WAS REPEATED ON AN ALTERNATE NON-SIEMENS METHOD AT AN ALTERNATE SITE AND THE RESULT WAS HIGHER. THE CUSTOMER IS NOT SURE WHICH RESULT IS CORRECT AND SENT BOTH RESULTS TO THE PHYSICIAN WITHOUT ISSUING A CORRECTED REPORT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT DEPRESSED UCFP RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19232474 ↗MalfunctionApr 1, 2024Low Test Results; Low Test Results
FIVE (5) FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) PATIENT SAMPLE RESULTS WERE OBTAINED ON AN ATELLICA ® CH 930 ANALYZER. THE FALSELY DEPRESSED PATIENT SAMPLE RESULTS WERE REPORTED TO THE PHYSICIAN AND UNKNOWN IF QUESTIONED. THE SAME SAMPLES WERE REPROCESSED ON THE SAME ATELLICA ® CH 930 ANALYZER UTILIZING AN ALTERNATE REAGENT LOT. THE REPEATED SAMPLE RESULTS WITH THE ALTERNATE REAGENT LOT WERE CONSIDERED CORRECT AND REPORTED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19010412 ↗MalfunctionMar 13, 2024Low Test Results; Low Test Results
TWO (2) FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) PATIENT SAMPLE RESULTS WERE OBTAINED ON AN ATELLICA ® CH 930 ANALYZER. THE FALSELY DEPRESSED PATIENT SAMPLE RESULTS WERE REPORTED TO THE PHYSICIAN AND NOT QUESTIONED. THE SAME SAMPLES WERE REPROCESSED ON AN ALTERNATE ATELLICA ® CH 930 ANALYZER. HIGHER RESULTS OBTAINED WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18894622 ↗MalfunctionMar 11, 2024Low Test Results; Low Test Results
FIVE (5) FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) PATIENT SAMPLE RESULTS WERE OBTAINED ON AN ATELLICA ® CH 930 ANALYZER. THE FALSELY DEPRESSED PATIENT SAMPLE RESULTS WERE REPORTED TO THE PHYSICIAN, AND IT IS UNKNOWN IF THEY WERE QUESTIONED. THE SAME SAMPLES WERE REPROCESSED ON ALTERNATE ATELLICA ® CH 930 ANALYZERS. THE HIGHER RESULTS OBTAINED WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18874951 ↗MalfunctionMar 8, 2024Low Test Results; Low Test Results
SIX (6) FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) PATIENT SAMPLE RESULTS WERE OBTAINED ON AN ATELLICA ® CH 930 ANALYZER. THE FALSELY DEPRESSED PATIENT RESULTS WERE NOT REPORTED TO THE PHYSICIAN. THE SAME SAMPLES WERE REPROCESSED ON AN ALTERNATE ATELLICA ® CH 930 ANALYZER. THE HIGHER RESULTS OBTAINED WERE CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18864569 ↗MalfunctionFeb 6, 2024Low Test Results; Low Test Results
A DISCORDANT FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULT ON A PATIENT SAMPLE WAS OBTAINED ON AN ATELLICA CH 930 ANALYZER. THE FALSELY DEPRESSED RESULT WAS REPORTED TO THE PHYSICIAN(S), AND THE RESULT WAS QUESTIONED. THE SAME SAMPLE WAS REPROCESSED ON THE ORIGINAL ATELLICA CH 930 ANALYZER. A HIGHER RESULT, CONSIDERED CORRECT, WAS OBTAINED. A CORRECTED REPORT WAS ISSUED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18649120 ↗MalfunctionOct 23, 2023Low Test Results; Low Test Results
TWO (2) DISCORDANT FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS WERE OBTAINED ON A PATIENT SAMPLE ON AN ATELLICA CH 930 ANALYZER. THE FALSELY DEPRESSED RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED ON THE SAME ATELLICA CH 930 ANALYZER WITH 1:10 DILUTION AND OBTAINED A FALSELY DEPRESSED RESULT. THE SAME SAMPLE WAS REPROCESSED ON AN ALTERNATE NON-SIEMENS INSTRUMENT. A HIGHER RESULT CONSIDERED CORRECT WAS OBTAINED AND REPORTED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17986451 ↗MalfunctionAug 30, 2023Low Test Results; Low Test Results
A DISCORDANT FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) PATIENT RESULT WAS OBTAINED ON AN ATELLICA® CH 930 ANALYZER. CUSTOMER STATED THE URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) IS INACCURATE, FALSELY DEPRESSED AND INCONSISTENT WITH THE MICROALBUMIN_2 (ALB_2) RESULT. THE FALSELY DEPRESSED PATIENT RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS NOT REPROCESSED. CUSTOMER STATED THEY DO NOT KNOW THE CORRECT RESULT BECAUSE THERE IS NO UCFP REAGENT FROM OTHER MANUFACTURERS TO COMPARE. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) PATIENT RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17651910 ↗MalfunctionAug 14, 2023Low Test Results; Low Test Results
ONE (1) FALSELY DEPRESSED URINE PATIENT SAMPLE RESULT WAS OBTAINED WITH URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP, LOT NUMBER 120243) WHEN PROCESSED ON AN ATELLICA CH 930 ANALYZER (S/N: (B)(6)). THE RESULT WAS DEPRESSED COMPARED TO THE ATELLICA CH MICROALBUMIN (UALB_2) RESULT AND A RESULT FROM A ROUTINE URINE TEST METHOD. THE INITIAL UCFP RESULT WAS NOT REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED UCFP RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17546065 ↗MalfunctionAug 11, 2023Low Test Results; Low Test Results
THE CUSTOMER INFORMED SIEMENS OF DISCORDANT LOW ATELLICA CH 930 URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP), LOT 120243 RESULTS ON A URINE SAMPLE. THE RETEST RESULT ON AN ALTERNATE METHOD WAS HIGHER. THE DISCORDANT RESULTS WERE NOT REPORTED. THE SAMPLE WAS RETESTED USING AN ALTERNATE UCFP METHOD AND THE RESULT WAS REPORTED AS CORRECT AND NOT QUESTIONED BY THE PHYSICIAN. QC WERE ALL WITHIN THE NORMAL RANGES. THE PATIENT HAS TYPE 2 DIABETES, MEDICATION IS UNKNOWN. THERE ARE NO ALLEGATIONS OF PATIENT OR USER INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17509507 ↗MalfunctionJun 19, 2023Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) RESULTS ON TWO PATIENT SAMPLES FROM AN ATELLICA CH 930 ANALYZER. THE FALSELY DEPRESSED RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE PATIENT SAMPLES WERE REPEATED ON AN ALTERNATE TESTING PLATFORM. THE REPEAT RESULTS WERE HIGHER THAN THE FALSELY DEPRESSED RESULTS. THE REPEAT RESULTS WERE REPORTED AS CORRECT TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED UCFP RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17151331 ↗MalfunctionMar 10, 2023Low Test Results; Low Test Results
THE CUSTOMER OBTAINED A FALSELY DEPRESSED URINARY/CEREBROSPINAL FLUID PROTEIN (UCFP) URINE SAMPLE RESULT USING LOT 120243 ON AN ATELLICA CH ANALYZER THAT DID NOT MATCH WITH THE PROFILE OF THE OTHER TESTS PERFORMED ON THIS PATIENT. A RETEST WAS PERFORMED ON THE SAME SAMPLE AND ANALYZER AND THE RESULT WAS SIMILAR TO THE INITIAL RESULT. THE ATELLICA CH UCFP RESULTS WERE NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS ALSO PROCESSED ON ATELLICA CH ANALYZERS WITH OPEN CHANNEL NON-SIEMENS URINE TOTAL PROTEIN REAGENT. THE CUSTOMER CONSIDERED THESE RESULTS TO BE FALSELY DEPRESSED AS WELL. RESULTS WERE NOT REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT UCFP RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16523611 ↗
Other BIURET (COLORIMETRIC), TOTAL PROTEIN devices
- ADVIA Chemistry UCFP 11319151Siemens Healthcare Diagnostics Inc. · BIURET (COLORIMETRIC), TOTAL PROTEIN2 recalls10 reports
- Atellica™ CH TP 11097604Siemens Healthcare Diagnostics Inc. · BIURET (COLORIMETRIC), TOTAL PROTEIN1 recall0 reports
- VITROS 8392292Ortho-Clinical Diagnostics, Inc. · BIURET (COLORIMETRIC), TOTAL PROTEIN1 recall0 reports
- Total Protein Gen.2 08058652190Roche Diagnostics GmbH · BIURET (COLORIMETRIC), TOTAL PROTEIN13 reports
- Total Protein Gen2 3183734190Roche Diagnostics GmbH · BIURET (COLORIMETRIC), TOTAL PROTEIN11 reports
- Total Protein Gen2 5171385190Roche Diagnostics GmbH · BIURET (COLORIMETRIC), TOTAL PROTEIN5 reports
- Atellica CH UPro 11097524Siemens Healthcare Diagnostics Inc. · BIURET (COLORIMETRIC), TOTAL PROTEIN1 report
- Total Protein Gen2 monocromatic 4657586190Roche Diagnostics GmbH · BIURET (COLORIMETRIC), TOTAL PROTEIN1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page