Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613336159781), received between Aug 5, 2025 and Aug 5, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results

    THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS ANTI-HBS RESULTS FROM TWO PATIENT SAMPLES TESTED ON THE COBAS E 602 MODULE. ONE EXAMPLE OF DISCREPANT RESULTS WAS PROVIDED: THE INITIAL RESULT WAS 9.58 IU/L (INDETERMINATE). THE FIRST REPEAT RESULT WAS 3.50 IU/ML WITH A DATA FLAG (NEGATIVE). THE SECOND REPEAT RESULT WAS 32.29 WITH A DATA FLAG IU/ML (POSITIVE). THE INITIAL RESULT WAS NOT REPORTED OUTSIDE THE LABORATORY, AS PATIENT SAMPLES WITH "INDETERMINATE" INITIAL RESULTS ARE RERUN TWICE. THE RESULT WAS REPORTED AS "NEGATIVE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22715427 ↗