Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613336175293), received between Nov 9, 2023 and Apr 13, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- False Positive Result4
- Inadequate or Insufficient Training1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 13, 2026Inadequate or Insufficient Training
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS HBSAG CONFIRMATORY TEST RESULTS FROM THE COBAS 6000 E601 MODULE. THE INITIAL SCREENING RESULT WAS 110.6 COI REACTIVE). THE CUSTOMER DID NOT REPEAT THE SCREENING ASSAY. THE CUSTOMER THEN TESTED THE SAMPLE USING ELECSYS HBSAG CONFIRMATORY TEST USING THE INCORRECT SAMPLE PREPARATION OPTION AND DETERMINED THE RESULT WAS POSITIVE (74.03 COI). ON (B)(6) 2026, THE CUSTOMER PERFORMED TWO MORE CONFIRMATORY TESTS FOR THIS SAMPLE USING THE OTHER SAMPLE PREPARATION OPTIONS. USING THE SECOND INCORRECT SAMPLE PREPARATION OPTION, THE CONCLUSION WAS NEGATIVE (234.4 COI). USING THE CORRECT SAMPLE PREPARATION OPTION, BUT AN INCORRECT DILUTION OF 1:10 INSTEAD OF 1:20, THE CONCLUSION WAS NEGATIVE (2374 COI). THE CUSTOMER THEN PERFORMED THE CONFIRMATORY TEST FOR THIS SAMPLE USING THE CORRECT SAMPLE PREPARATION OPTION AND CORRECT DILUTION. THE CONCLUSION WAS POSITIVE (4.39 COI).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24858187 ↗MalfunctionSep 16, 2025False Positive Result; False Positive Result
THE INITIAL REPORTER ALLEGED DISCREPANT RESULTS WITH THE HBSAG CONFIRMATORY ELECSYS COBAS E V2 ASSAY. THE PATIENT HAD A HISTORY OF SLIGHTLY POSITIVE HBSAG II RESULTS AND NEGATIVE OR SLIGHTLY POSITIVE HBSAG CONFIRMATORY TEST RESULTS IN PRIOR YEARS. ON (B)(6)2022, THE PATIENT SAMPLE WAS TESTED FOR HBSAG II, YIELDING THE FOLLOWING RESULTS: RUN 1: 1.16 COI; RUN 2 (AFTER CENTRIFUGATION): 1.16 COI; RUN 3: 1.28 COI; RUN 4 (USING A NEW BLOOD COLLECTION TUBE): 1.18 COI; AND RUN 5: 1.22 COI. ALL RESULTS WERE REPEATEDLY REACTIVE. THE CUSTOMER PERFORMED HBSAG CONFIRMATORY TESTING USING A NEW REAGENT LOT. THE CONFIRMATORY TEST RESULTS WERE AS FOLLOWS: POSITIVE CONTROL PLUS CONFIRMATORY REAGENT (C): 0.49 COI; POSITIVE CONTROL PLUS COMPARATOR REAGENT (D): 3.76 COI. THE CALCULATED RATIO (C/D*100) WAS 13.0%, WHICH CONFIRMED THE SAMPLE AS POSITIVE. HOWEVER, THE CUSTOMER USED 180 L + 20 L INSTEAD OF THE RECOMMENDED 270 L + 30 L VOLUMES FOR THE CONFIRMATORY TEST. DESPITE THE CONFIRMED POSITIVE RESULT, THE CUSTOMER REPORTED THE SAMPLE AS NEGATIVE BASED ON RESULTS FROM AN OUTSOURCED LABORATORY. NO NEW BLOOD SAMPLE COULD BE COLLECTED FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23067273 ↗MalfunctionMay 14, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE INITIAL REPORTER QUESTIONED NEGATIVE RESULTS FOR 4 PATIENT SAMPLES TESTED FOR ELECSYS HBSAG CONFIRMATORY TESTS (HBSAG CONFIRMATORY) ON A COBAS E 801 ANALYTICAL UNIT WHERE THE INITIAL HBSAG RESULTS WERE HIGH (POSITIVE). OF THE DATA PROVIDED, DISCREPANT RESULTS WERE IDENTIFIED FOR 1 PATIENT SAMPLE BETWEEN 2 DIFFERENT HBSAG CONFIRMATORY REAGENTS. THIS MEDWATCH WILL COVER HBSAG CONFIRMATORY REAGENT WITH MATERIAL NUMBER 09127127190. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE HBSAG CONFIRMATORY REAGENT WITH MATERIAL NUMBER 11820648122. REFER TO THE ATTACHED DATA FOR THE PATIENT RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19311831 ↗MalfunctionMar 26, 2024False Positive Result; False Positive Result
THE INITIAL REPORTER RECEIVED A QUESTIONABLE HBSAG CONFIRMATORY ELECSYS RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS E 801 ANALYTICAL UNIT. THE QUESTIONABLE RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY. THE RESULT OF HBSAG CONFIRMATORY TEST WAS "REACTIVE". THE PATIENT SAMPLE WAS SENT TO ANOTHER LABORATORY. THE HBSAG RESULT FROM THE OTHER LABORATORY WAS "NEGATIVE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18977392 ↗MalfunctionMar 26, 2024False Positive Result; False Positive Result
THE INITIAL REPORTER RECEIVED QUESTIONABLE HBSAG CONFIRMATORY ELECSYS RESULTS FROM TWO PATIENT SAMPLES TESTED ON THE COBAS E 801 MODULE. THE REPORTER STATED THAT THE PATIENTS ARE CONSIDERED TO BE NEGATIVE FOR HEPATITIS B. SAMPLE 1: ON (B)(6) 2023: PATIENT 1: THE HBSAG CONFIRMATORY RESULT WAS 50.500% (POSITIVE). THE RESULT OF THE HEPATITIS B VIRUS POLYMERASE CHAIN REACTION: HEPATITIS B VIRUS - DEOXYRIBOSE NUCLEIC ACID (HBVPCR: HBV-DNA) TEST WAS "NOT DETECTABLE". ON (B)(6) 2023: THE HBSAG CONFIRMATORY RESULT WAS 21.200% (POSITIVE). SAMPLE 2: ON (B)(6) 2023: THE HBSAG CONFIRMATORY RESULT WAS 28.500% (POSITIVE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18977159 ↗MalfunctionNov 9, 2023False Positive Result; False Positive Result
THERE WAS AN ALLEGATION OF A QUESTIONABLE HBSAG CONFIRMATORY ELECSYS RESULT FROM THE COBAS E 801 MODULE. THE RESULT FOR THE HBSAG CONFIRMATORY TEST WAS "POSITIVE" AND WAS REPORTED OUTSIDE OF THE LABORATORY. THE PATIENT WENT TO ANOTHER MEDICAL INSTITUTION WHERE THE HBSAG TEST RESULT WAS "NEGATIVE". THE METHODOLOGY USED WAS NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18103665 ↗
Other Test, hepatitis b (b core, be antigen, be antibody, b core igm) devices
- Elecsys HBsAg II 08814848190Roche Diagnostics GmbH · Test, hepatitis b (b core, be antigen, be antibody, b core igm)93 reports
- ARCHITECT 4P53-25Abbott Ireland Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall16 reports
- VITROS 6801925Ortho Clinical Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall15 reports
- ARCHITECT 1L82-27Abbott Ireland Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall4 reports
- VITROS 6801926Ortho Clinical Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall0 reports
- ARCHITECT 4P54-25Abbott Ireland Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)27 reports
- ADVIA Centaur® HBsII 10492138Siemens Healthcare Diagnostics Inc. · Test, Hepatitis B (B Core, Be Antigen, Be Antibody, B Core Igm)18 reports
- Atellica™ IM HBsII 10995604Siemens Healthcare Diagnostics Inc. · Test, hepatitis b (b core, be antigen, be antibody, b core igm)12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page