Elecsys HIV Duo 08836973162Recalls & reported problems
Roche Diagnostics GmbH · Class U
All models of Elecsys HIV Duo (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613336183120), received between Feb 18, 2026 and Jul 29, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- False Positive Result5
- Incorrect, Inadequate or Imprecise Result or Readings2
- Non Reproducible Results2
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 29, 2026False Positive Result; Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50067037 ↗MalfunctionJul 29, 2026False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Non Reproducible Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50066870 ↗MalfunctionMay 29, 2026False Positive Result
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS HIV DUO AND ELECSYS HBSAG II RESULTS FOR 1 PATIENT ON A COBAS E 801 ANALYTICAL UNIT. THE INITIAL ELECSYS HIV DUO RESULT WAS REACTIVE. ON (B)(6) 2026, THE SAMPLE WAS SENT FOR CONFIRMATORY TESTING. THE CONFIRMATORY HIV (GEENIUS METHOD) TEST RESULT WAS "CONFIRMED NON-REACTIVE". THE INITIAL ELECSYS HBSAG II RESULT WAS REACTIVE. ON (B)(6) 2026, THE SAMPLE WAS SENT FOR CONFIRMATORY TESTING. THE CONFIRMATORY (ROCHE E801 ECLIA METHOD) TEST RESULT WAS "CONFIRMED NON-REACTIVE". THE SAMPLE WAS REPEATED BECAUSE OF THE INITIALLY REACTIVE RESULTS. THIS MEDWATCH WILL APPLY TO THE ELECSYS HIV DUO ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE ELECSYS HBSAG II ASSAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333754 ↗MalfunctionMay 29, 2026False Positive Result
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS HIV DUO AND ELECSYS HBSAG II RESULTS FOR 1 PATIENT ON A COBAS E 801 ANALYTICAL UNIT. THE INITIAL ELECSYS HIV DUO RESULT WAS REACTIVE. ON(B)(6) 2026, THE SAMPLE WAS SENT FOR CONFIRMATORY TESTING. THE CONFIRMATORY HIV (GEENIUS METHOD) TEST RESULT WAS "CONFIRMED NON-REACTIVE". THE INITIAL ELECSYS HBSAG II RESULT WAS REACTIVE. ON (B)(6) 2026, THE SAMPLE WAS SENT FOR CONFIRMATORY TESTING. THE CONFIRMATORY (ROCHE E801 ECLIA METHOD) TEST RESULT WAS "CONFIRMED NON-REACTIVE". THIS MEDWATCH WILL APPLY TO THE ELECSYS HIV DUO ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER(B)(6) FOR INFORMATION RELATED TO THE ELECSYS HBSAG II.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333752 ↗MalfunctionFeb 18, 2026False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 8 PATIENTS' SAMPLES TESTED WITH ELECSYS HIV DUO ASSAYS ON A COBAS E 801 ANALYTICAL UNIT. SAMPLE 1 AND SAMPLE 2 WERE: INITIALLY TESTED FOR HIV ON ABBOTT ALINITY, AND THE RESULTS WERE NEGATIVE. ELECSYS HIV DUO RESULTS: POSITIVE. GEENIUS HIV ½ CONFIRMATORY TEST RESULT: NEGATIVE. SAMPLE 3, SAMPLE 4, SAMPLE 5, SAMPLE 6, SAMPLE 7, AND SAMPLE 8 WERE: ELECSYS HIV DUO RESULTS: POSITIVE. GEENIUS HIV ½ CONFIRMATORY TEST RESULTS: NEGATIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24379356 ↗
Other Serology tests used for infectious disease screening of blood and plasma donations devices
- Elecsys HBsAg II 08814848162Roche Diagnostics GmbH · Serology tests used for infectious disease screening of blood and plasma donations5 reports
- Elecsys HTLV-I/II 09015272162Roche Diagnostics GmbH · Serology tests used for infectious disease screening of blood and plasma donations2 reports
- Elecsys Anti-HBc II 09014926162Roche Diagnostics GmbH · Serology tests used for infectious disease screening of blood and plasma donations2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page