Skip to content

Elecsys HBsAg II 08814848162Recalls & reported problems

Roche Diagnostics GmbH · Class U

All models of Elecsys HBsAg II (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613336183076), received between Nov 17, 2025 and Sep 23, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Positive Result5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026False Positive Result

    There was an allegation of questionable results with the Elecsys HBsAg II assay for a donor sample tested on the Roche instrument (details have been requested). It was reported that the patient has always been non-reactive on a non-Roche instrument. Since switching to Roche instruments, the last two times the patient has been reactive for HBsAg but then confirmed negative for HBsAg.The test was repeated as the reactive result did not match previous history.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870259 ↗
  • MalfunctionMay 29, 2026False Positive Result

    THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS HBSAG II AND ELECSYS HIV DUO RESULTS FOR 1 PATIENT ON A COBAS E 801 ANALYTICAL UNIT. THE INITIAL ELECSYS HIV DUO RESULT WAS REACTIVE. ON (B)(6) 2026, THE SAMPLE WAS SENT FOR CONFIRMATORY TESTING. THE CONFIRMATORY HIV (GEENIUS METHOD) TEST RESULT WAS "CONFIRMED NON-REACTIVE". THE INITIAL ELECSYS HBSAG II RESULT WAS REACTIVE. ON (B)(6) 2026, THE SAMPLE WAS SENT FOR CONFIRMATORY TESTING. THE CONFIRMATORY (ROCHE E801 ECLIA METHOD) TEST RESULT WAS "CONFIRMED NON-REACTIVE". THE SAMPLE WAS REPEATED BECAUSE OF THE INITIALLY REACTIVE RESULTS. THIS MEDWATCH WILL APPLY TO THE ELECSYS HBSAG II ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE ELECSYS HIV DUO ASSAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25333764 ↗
  • MalfunctionMay 29, 2026False Positive Result

    THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS HBSAG II AND ELECSYS HIV DUO RESULTS FOR 1 PATIENT ON A COBAS E 801 ANALYTICAL UNIT. THE INITIAL ELECSYS HIV DUO RESULT WAS REACTIVE. ON (B)(6) 2026, THE SAMPLE WAS SENT FOR CONFIRMATORY TESTING. THE CONFIRMATORY HIV (GENIUS METHOD) TEST RESULT WAS "CONFIRMED NON-REACTIVE". THE INITIAL ELECSYS HBSAG II RESULT WAS REACTIVE. ON(B)(6) 2026, THE SAMPLE WAS SENT FOR CONFIRMATORY TESTING. THE CONFIRMATORY (ROCHE E801 ECLIA METHOD) TEST RESULT WAS "CONFIRMED NON-REACTIVE". THIS MEDWATCH WILL APPLY TO THE ELECSYS HBSAG II ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE ELECSYS HIV DUO ASSAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25333761 ↗
  • MalfunctionNov 17, 2025False Positive Result; False Positive Result

    THERE WAS AN ALLEGATION OF FALSE-POSITIVE ELECSYS HBSAG II RESULTS FOR 1 PATIENT SAMPLE ON A COBAS E 801 ANALYTICAL UNIT. THE INITIAL RESULT WAS 0.993 COI (REACTIVE). THE SAMPLE WAS REPEATED, AND THE RESULTS WERE 0.962 COI (REACTIVE) AND 0.907 COI (REACTIVE). ON (B)(6) 2025, CONFIRMATORY TESTING CONFIRMED THAT THE SAMPLE WAS NON-REACTIVE. A NEW SAMPLE WAS OBTAINED FROM THE PATIENT, AND THE ELECSYS HBSAG II RESULT WAS NON-REACTIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23571006 ↗
  • MalfunctionNov 17, 2025False Positive Result; False Positive Result

    THERE WAS AN ALLEGATION OF FALSE-POSITIVE ELECSYS HBSAG II RESULTS FOR 1 PATIENT SAMPLE ON A COBAS E 801 ANALYTICAL UNIT. THE INITIAL RESULT WAS 1.02 COI (REACTIVE). THE SAMPLE WAS REPEATED, AND THE RESULTS WERE 1.01 COI (REACTIVE) AND 0.998 COI (REACTIVE). ON (B)(6) 2025, CONFIRMATORY TESTING CONFIRMED THAT THE SAMPLE WAS NON-REACTIVE. A NEW SAMPLE WAS OBTAINED FROM THE PATIENT, AND THE ELECSYS HBSAG II RESULT WAS NON-REACTIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23570375 ↗

Other Serology tests used for infectious disease screening of blood and plasma donations devices

All Serology tests used for infectious disease screening of blood and plasma donations devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page