SCHILLER BP-200 plusRecalls & reported problems

Schiller AG · The BP-200 plus is intended to be used as an adjunct to exercise stress testing devices. It is intended to measure and display diastolic and systolic blood pressure, heart rate, percentage of oxygen saturationin arterial blood (SpO2) and pulse rate in adult or adolescent patients during stress tests. The BP-200 plus can be used for patients of both sexes and all races. · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613365000122), received between Oct 30, 2024 and Mar 21, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Unable to Obtain Readings2
  • High Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 21, 2025Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT THE DEVICE FUNCTIONS NORMALLY DURING RESTING BLOOD PRESSURE MEASUREMENTS. HOWEVER, DURING STRESS TESTING, THE DEVICE FAILS TO DISPLAY BLOOD PRESSURE VALUES AFTER STAGE 2, AND OCCASIONALLY DURING STAGE 2 AS WELL. THE USER ALSO EXPERIENCED ERROR CODES 180, 181, AND 182, ALONG WITH OCCASIONAL ABNORMAL BLOOD PRESSURE READINGS DURING STRESS TESTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21661633 ↗
  • MalfunctionMar 21, 2025Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT THE DEVICE FUNCTIONS NORMALLY DURING RESTING BLOOD PRESSURE MEASUREMENTS. HOWEVER, DURING STRESS TESTING, THE DEVICE FAILS TO DISPLAY BLOOD PRESSURE VALUES AFTER STAGE 2, AND OCCASIONALLY DURING STAGE 2 AS WELL. THE USER ALSO EXPERIENCED ERROR CODES 180, 181, AND 182, ALONG WITH OCCASIONAL ABNORMAL BLOOD PRESSURE READINGS DURING STRESS TESTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21661632 ↗
  • MalfunctionOct 30, 2024High Readings; High Readings

    DURING A STRESS TEST THE DEVICE SHOWED VERY HIGH BLOOD PRESSURE VALUES.

    Patient problems: Insufficient Information

    MAUDE report 20565839 ↗

Other System, Measurement, Blood-Pressure, Non-Invasive devices

All System, Measurement, Blood-Pressure, Non-Invasive devices →