Skip to content

HAMILTON-H900 950004Recalls & reported problems

Hamilton Medical AG · Class 2

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630002801546), received between Feb 11, 2025 and Feb 11, 2025 — classified as 1 death.

The most frequently reported specific device problem is “Contamination” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 5, 2021Completed

    When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm will be delayed and medical staff are unaware that there is no water in the chamber.

    Z-1385-2022 · Root cause: Employee error · FDA record ↗

Most reported problems

Device problems

  • Contamination1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathFeb 11, 2025Contamination; Contamination

    INFANT WAS DIAGNOSED WITH A BACTERIAL INFECTION PSEUDOMONAS AERUGINOSA. IT WAS BELIEVED THAT THE ORIGIN WAS FROM THE HAMILTON H-900 DUE TO THE RAIN OUT. THE H900 WAS LATER DEEMED NOT TO BE THE SOURCE OF THE BACTERIA AND WAS NOT LIABLE. NO CULTURES ON PRODUCT, TESTING WAS COMPLETED (CONTEXT OF TESTING IS BIOBURDEN TESTING).

    Patient problems: Insufficient Information

    MAUDE report 21359268 ↗

Other Humidifier, Respiratory Gas, (Direct Patient Interface) devices

All Humidifier, Respiratory Gas, (Direct Patient Interface) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.