HAMILTON-H900 Humidifier, Models: 950001, 950004 recallZ-1385-2022 · Hamilton Medical AG · initiated Jan 5, 2021

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hamilton Medical AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm will be delayed and medical staff are unaware that there is no water in the chamber.

FDA root cause category: Employee error

What the company is doing

On 01/05/21, Field Safety Corrective Action notices were sent to customers who were asked to do the following: Follow the instructions in the HAMILTON-H900 Operator s Manual 624431 section 3.4. Before connecting a new patient to the humidifier; " Verify that the alarms operate correctly. " Ensure that the water supply to the humidifier chamber is functioning properly. Distribute this notification to all end users within your organization. A Medical Representative will be in contact to schedule service with a software update. Customers with additional questions or concerns were encouraged to call Technical Support at 1-800-426-6331, option #2.

Recall details

Recall number
Z-1385-2022
Status
Completed
Initiated by the company
Jan 5, 2021
Recalling firm
Hamilton Medical AG
Quantity
203
Product code
BTT
FDA submissions
K163283
Distribution
US: WI, FL, CA, NV, PA, TX, PA

Product description

HAMILTON-H900 Humidifier, Models: 950001, 950004

Code information (models, lots, serial numbers)

UDI:07630002801546. All devices with software version 1.10c