Skip to content

MiniMAX Stems 01.13.102RRecalls & reported problems

Medacta International SA · miniMAX Stem - Size 2 Right · Class 2 · Implant

All models of MiniMAX Stems (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630030803185), received between Jul 15, 2019 and Mar 13, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (2 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 15, 2017Terminated

    Complaint monitoring noted a trend of packaging being breached for the MiniMAX Stem.

    Z-0196-2019 · Root cause: Package design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Osseointegration Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1

Patient problems

  • Inadequate Osseointegration2
  • Joint Disorder2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 13, 2025Osseointegration Problem; Osseointegration Problem

    THE PATIENT HAD A PRIMARY HIP SURGERY ON (B)(6) 2023. SUBSEQUENTLY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM. ON (B)(6) 2024, THE SURGEON REVISED THE MEDACTA STEM, HEAD AND LINER WITH ANOTHER MEDACTA STEM, HEAD AND LINER. PRESENTLY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM. THE SURGEON REVISED THE MEDACTA STEM AND HEAD WITH A COMPETITOR'S COMPONENTS AND REVISED THE MEDACTA LINER WITH ANOTHER MEDACTA LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21591415 ↗
  • InjuryJul 11, 2024Osseointegration Problem; Osseointegration Problem

    MINIMAX STEM MOBILIZED AND ROTATED IN THE PATIENT'S FEMUR. STEM AND HEAD REVISED AT ABOUT 3 MONTHS FROM THE PREVIOUS REVISION SURGERY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19719882 ↗
  • InjuryJul 3, 2020Device Dislodged or Dislocated

    THE PATIENT CAME IN REPORTING A LEG LENGTH DISCREPANCY THAT WAS CAUSED BY A SUBSIDED STEM. THE SURGEON REVISED THE STEM, HEAD AND LINER 7 MONTHS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder

    MAUDE report 10229609 ↗
  • InjuryJul 15, 2019Adverse Event Without Identified Device or Use Problem

    ON (B)(4) 2019 WE WERE ALERTED OF A REVISION PERFORMED ON (B)(6) 2019. REVISION SURGERY HAS BEEN PERFORMED 6 MONTHS AFTER PRIMARY DUE TO LEG LENGTH DISCREPANCY, 15 MM (APPARENTLY DUE TO SEVERE SCOLIOSIS OF THE SPINE). THROUGH A POSTERIOR APPROACH THE SURGEON REMOVED THE STEM, LINER AND HEAD. HE REPLACED THESE EXPLANTED ITEMS WITH AN MPACT HOODED LINER AND A COMPETITOR REVISION STEM AND HEAD.

    Patient problems: Joint Disorder

    MAUDE report 8789471 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page