MiniMAX Stems 01.13.102RRecalls & reported problems
Medacta International SA · miniMAX Stem - Size 2 Right · Class 2 · Implant
All models of MiniMAX Stems (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630030803185), received between Jul 15, 2019 and Mar 13, 2025 — classified as 4 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (2 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Nov 15, 2017Terminated
Complaint monitoring noted a trend of packaging being breached for the MiniMAX Stem.
Z-0196-2019 · Root cause: Package design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Osseointegration Problem2
- Adverse Event Without Identified Device or Use Problem1
- Device Dislodged or Dislocated1
Patient problems
- Inadequate Osseointegration2
- Joint Disorder2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMar 13, 2025Osseointegration Problem; Osseointegration Problem
THE PATIENT HAD A PRIMARY HIP SURGERY ON (B)(6) 2023. SUBSEQUENTLY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM. ON (B)(6) 2024, THE SURGEON REVISED THE MEDACTA STEM, HEAD AND LINER WITH ANOTHER MEDACTA STEM, HEAD AND LINER. PRESENTLY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM. THE SURGEON REVISED THE MEDACTA STEM AND HEAD WITH A COMPETITOR'S COMPONENTS AND REVISED THE MEDACTA LINER WITH ANOTHER MEDACTA LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration
MAUDE report 21591415 ↗InjuryJul 11, 2024Osseointegration Problem; Osseointegration Problem
MINIMAX STEM MOBILIZED AND ROTATED IN THE PATIENT'S FEMUR. STEM AND HEAD REVISED AT ABOUT 3 MONTHS FROM THE PREVIOUS REVISION SURGERY.
Patient problems: Inadequate Osseointegration
MAUDE report 19719882 ↗InjuryJul 3, 2020Device Dislodged or Dislocated
THE PATIENT CAME IN REPORTING A LEG LENGTH DISCREPANCY THAT WAS CAUSED BY A SUBSIDED STEM. THE SURGEON REVISED THE STEM, HEAD AND LINER 7 MONTHS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Disorder
MAUDE report 10229609 ↗InjuryJul 15, 2019Adverse Event Without Identified Device or Use Problem
ON (B)(4) 2019 WE WERE ALERTED OF A REVISION PERFORMED ON (B)(6) 2019. REVISION SURGERY HAS BEEN PERFORMED 6 MONTHS AFTER PRIMARY DUE TO LEG LENGTH DISCREPANCY, 15 MM (APPARENTLY DUE TO SEVERE SCOLIOSIS OF THE SPINE). THROUGH A POSTERIOR APPROACH THE SURGEON REMOVED THE STEM, LINER AND HEAD. HE REPLACED THESE EXPLANTED ITEMS WITH AN MPACT HOODED LINER AND A COMPETITOR REVISION STEM AND HEAD.
Patient problems: Joint Disorder
MAUDE report 8789471 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page