Skip to content

MiniMAX Stems recallZ-0196-2019 · Medacta Usa Inc · initiated Nov 15, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medacta Usa Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Complaint monitoring noted a trend of packaging being breached for the MiniMAX Stem.

FDA root cause category: Package design/selection

What the company is doing

The firm initiated the recall by telephone to its distributors. The firm requested return of all recalled units.

Recall details

Recall number
Z-0196-2019
Status
Terminated
Initiated by the company
Nov 15, 2017
Recalling firm
Medacta Usa Inc
Quantity
837 units
Product code
LZO
FDA submissions
K170845
Distribution
NC, ID, Belgium, Switzerland, Spain, Germany, Australia, France, Italy

Product description

Medacta International MiniMAX CEMENTLESS ANATOMICAL STEM, 12/14: for cementless use in total or partial hip arthroplasty in primary or revision surgery.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

a) REF 01.13.100L, GTIN 07630030803130 b) REF 01.13.101L, GTIN 07630030803154 c) REF 01.13.102L, GTIN 07630030803178 d) REF 01.13.103L, GTIN 07630030803192 e) REF 01.13.104L, GTIN 07630030803215 f) REF 01.13.105L, GTIN 07630030803239 g) REF 01.13.106L, GTIN 07630030803253 h) REF 01.13.107L, GTIN 07630030803277 I) REF 01.13.108L, GTIN 07630030803291 j) REF 01.13.100R, GTIN 07630030803147 k) REF 01.13.101R, GTIN 07630030803161 l) REF 01.13.102R, GTIN 07630030803185 m) REF 01.13.103R, GTIN 07630030803208 n) REF 01.13.104R, GTIN 07630030803222 o) REF 01.13.105R, GTIN 07630030803246 p) REF 01.13.106R, GTIN 07630030803260 q) REF 01.13.107R, GTIN 07630030803284 r) REF 01.13.108R, GTIN 07630030803307 Lots produced with the old version of the packaging, including lot numbers less than 169041.

Devices named in this recall

18 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates