MiniMAX Stems 01.13.103RRecalls & reported problems
Medacta International SA · miniMAX Stem - Size 3 Right · Class 2 · Implant
All models of MiniMAX Stems (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07630030803208), received between Mar 16, 2018 and Aug 22, 2024 — classified as 5 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Nov 15, 2017Terminated
Complaint monitoring noted a trend of packaging being breached for the MiniMAX Stem.
Z-0196-2019 · Root cause: Package design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available3
- Fracture2
- Osseointegration Problem1
Patient problems
- Bone Fracture(s)3
- Inadequate Osseointegration1
- Unspecified Infection1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryAug 22, 2024Osseointegration Problem; Osseointegration Problem
THE PATIENT HAS A PRIMARY HIP SURGERY ON (B)(6) 2016. SUBSEQUENTLY, THE PATIENT CAME IN REPORTING PAIN DUE TO ASEPTIC LOOSENING OF THE FEMORAL STEM. ON (B)(6) 2021, THE SURGEON REVISED THE STEM, HEAD AND LINER. (MDR 2021-00641) (B)(6) 2024: THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM. THE SURGEON REVISED THE MEDACTA STEM WITH A COMPETITOR STEM, AND REVISED THE MEDACTA HEAD AND LINER WITH ANOTHER MEDACTA HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration
MAUDE report 20033012 ↗InjuryJun 18, 2019Fracture; Appropriate Device Problem Term/Code Not Available
THE PATIENT HAD A FEMUR SPIRAL FRACTURE 3 DAYS AFTER SURGERY. THE REASON OF FRACTURE IS UNKNOWN. THE SURGEON REVISED SUCCESSFULLY THE PATIENT SWAPPING THE STEM AND THE CERAMIC HEAD.
Patient problems: Bone Fracture(s)
MAUDE report 8707192 ↗InjuryJun 11, 2019Appropriate Device Problem Term/Code Not Available
REVISION SURGERY PERFORMED 2 WEEKS AFTER THE PRIMARY DUE TO BONE FRACTURE AND SUBSIDED STEM. THE PATIENT WAS WALKING TO THE BATHROOM AND FELT A CRACK IN HER HIP, COULD THEN NOT WEIGHT BARE. THE MINI MAX STEM HAS BEEN REMOVED, IMPLANTED A XACTA 37.5MM AND A NEW HEAD, PLUS 3 CABLES.
Patient problems: Bone Fracture(s)
MAUDE report 8685954 ↗InjuryNov 20, 2018Fracture
9 DAYS POST PRIMARY, AFTER A SESSION OF PHYSIOTHERAPY, THE PATIENT HAS FELT SEVERE PAIN AND COULD NOT STAND. X-RAY REVEALED A PERIPROSTHETIC FRACTURE, SO ON THE NEXT DAY THE PATIENT UNDERWENT TO A REVISION SURGERY: THE FEMUR WAS CABLED WITH 2 CERCLAGE WIRES AND THE STEM WAS CONVERTED TO QUADRA STEM. THE SURGERY WAS COMPLETED SUCCESSFULLY. IT WAS NOT BELIEVED THAT THE FRACTURE HAD BEEN DUE TO AGGRESSIVE PHYSIOTHERAPY, BUT THEY BELIEVED THAT IT COULD HAVE BEEN DUE TO RAPID RECOVERY OF THE PATIENT COUPLED WITH AN INITIAL VALGUS POSITION AND UNDERSIZED STEM - THAT MIGHT HAVE SUBSIDED DUE TO VERY EARLY MOBILIZATION INTO NEUTRAL/VARUS POSITION (AND THEN BECAME "OVERSIZED" HAVE CONTRIBUTED TO THE FRACTURE).
Patient problems: Bone Fracture(s)
MAUDE report 8089816 ↗InjuryMar 16, 2018Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN DUE TO SIGNS OF INFECTION. THE PATHOGEN IS UNKNOWN. THE SURGEON WASHED OUT THE HIP AND REVISED THE STEM, HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 7344708 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page