MiniMAX Stems 01.13.105LRecalls & reported problems
Medacta International SA · miniMAX Stem - Size 5 Left · Class 2 · Implant
All models of MiniMAX Stems (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07630030803239), received between Jan 3, 2018 and Jun 26, 2025 — classified as 8 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Nov 15, 2017Class IITerminated
Complaint monitoring noted a trend of packaging being breached for the MiniMAX Stem.
Z-0196-2019 · Root cause: Package design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Fracture2
- Device Dislodged or Dislocated1
- Inadequacy of Device Shape and/or Size1
- Migration1
- Unsealed Device Packaging1
Patient problems
- Bone Fracture(s)4
- No Known Impact Or Consequence To Patient2
- Joint Disorder1
- Pain1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJun 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING THE PRIMARY TOTAL HIP ARTHROPLASTY (THA) IN 2021, THE PATIENT SUSTAINED A FRACTURE THAT WAS TREATED WITH CERCLAGE. FROM THE OUTSET, SHE EXPERIENCED LIMPING AND PAIN, WHICH EVENTUALLY LED TO A REVISION OF THE IMPLANT AT 3 YEARS AND 11 MONTHS POST PRIMARY. DURING THE REMOVAL, THE MINIMAX STEM, DESPITE THE ABSENCE OF A LARGE PORTION OF THE GREATER TROCHANTER AND THE PROXIMAL PERIPROSTHETIC BONE, WAS FOUND TO BE VERY STABLE. IT WAS REPLACED WITH A MODULAR M-VIZION STEM. THE ACETABULAR COMPONENT WAS LEFT UNTOUCHED, AS IT WAS FOUND TO BE INTACT AND STABLE.
Patient problems: Pain
MAUDE report 22334724 ↗InjurySep 19, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT HAD A PRIMARY HIP SURGERY ON (B)(6) 2018. ON (B)(6) 2020, THE PATIENT CAME IN REPORTING PAIN DUE TO A SUBSIDED STEM WHERE DURING THE REVISION SURGERY AN OSTEOTOMY OF THE FEMUR TO REMOVE THE STEM WAS NECESSARY. THE SURGEON USED CABLES AND REVISED THE HEAD AND STEM. THE SURGERY WAS COMPLETED SUCCESSFULLY. ON (B)(6) 2022, THE PATIENT CAME IN REPORTING PAIN DUE TO A FRACTURED FEMUR. THE SURGEON REVISED THE STEM WITH A COMPETITOR STEM AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Bone Fracture(s)
MAUDE report 15444795 ↗InjuryDec 9, 2020Migration; Migration
4 WEEKS AFTER PRIMARY THR. PATIENT CAME TO THE CLINIC COMPLAINING ON PAIN. X-RAYS SHOW SUBSIDENCE OF THE FEMORAL STEM AND FRACTURE OF THE CALCAR. REVISION OF THE STEM PERFORMED WITH DIFFERENT MANUFACTURER STEM. IT SHOULD BE NOTED THAT THE PATIENT HAS SEVERE OSTEOPOROSIS.
Patient problems: Bone Fracture(s)
MAUDE report 10975810 ↗InjuryFeb 21, 2020Device Dislodged or Dislocated
REVISION SURGERY PERFORMED 3 DAYS AFTER PRIMARY DUE TO HIP DISLOCATION (HEAD FROM LINER, CERAMIC ON CERAMIC) PROBABLY DUE TO STEM SINKING. THE MINIMAX STEM WAS REMOVED AND A CEMENTED X-ACTA SHORT WAS IMPLANTED. THE PATIENT WAS OPERATED IN PRIMARY DUE TO BONE FEMUR FRACTURE.
Patient problems: Joint Disorder
MAUDE report 9738554 ↗InjuryJul 3, 2019Fracture
REVISION SURGERY PERFORMED 14 DAYS AFTER THE PRIMARY DUE TO PAIN CAUSED BY A FRACTURED FEMUR AND SUBSIDED STEM (THE CAUSE OF THE FRACTURE IS UNKNOWN). THE SURGEON REVISED THE STEM AND THE HEAD AND CABLED THE FRACTURE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Bone Fracture(s)
MAUDE report 8757404 ↗InjuryJan 11, 2019Inadequacy of Device Shape and/or Size
REVISION SURGERY 1 MONTH AFTER PRIMARY SURGERY. THE STEM WAS TOO DEEP INSIDE THE FEMORAL BONE. STEM AND HEAD WERE REVISED. THE STEM LOOKS SLIGHTLY UNDERSIZED BASED ON RADIOGRAPHIC ANALYSIS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8239869 ↗InjuryJul 20, 2018Fracture
POST-OP THE SURGEON NOTICED THAT THE PATIENT HAD A FRACTURE ON THE FEMUR. THE SURGEON WENT BACK IN ON THE SAME DAY TO REVISE THE PATIENT. STEM, HEAD AND BIPOLAR HEAD WERE REPLACED.
Patient problems: Bone Fracture(s)
MAUDE report 7706398 ↗InjuryJan 3, 2018Unsealed Device Packaging
DURING THE IMPLANT OPENING THE PRIMARY PACKAGING BLISTER RESULTED SCRATCHED AND BROKEN. A BACK-UP STEM WAS OPENED. THE SECOND STEM HAD THE PRIMARY PACKAGING BLISTER SCRATCHED BUT NOT BROKEN AND THE SURGEON CONSIDERED IT SAFE FOR THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7160287 ↗
Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.