Skip to content

M.U.S.T. Reduction Pedicle Screws 03.50.732Recalls & reported problems

Medacta International SA · Reduction Pedicle Screw 7x50 mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630030846977), received between Jun 1, 2021 and Jun 1, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Integrity Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Integrity Problem1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 1, 2021Material Integrity Problem; Material Integrity Problem

    DURING THE PRIMARY SPINE CASE WHEN THE SURGEON INSERTED THE MUST PEDICLE SCREWS FROM T08 TO S02 USING THE MYSPINE GUIDES, THE ROD TO CONNECT ALL THE SCREWS, AND THE SET SCREW INTO EACH SCREW, DURING FINAL TIGHTENING, ONE OF THE SCREW FLANGE (CONTAINING THE ROD AND SET SCREW) DEFORMED AND A FRAGMENT OF THE TULIP POPPED OFF THE SCREW SHAFT. THE SURGEON HAD TO REMOVE THE SET SCREWS, REMOVED THE ROD, REMOVED THE AFFECTED SCREW, REPLACED THE BROKEN SCREW. THERE WAS A 1 HOUR AND 30-MINUTE DELAY AND ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: Failure of Implant

    MAUDE report 11912724 ↗

Other Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease devices

All Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page