Skip to content

M.U.S.T. Pedicle Screw System 03.61.324Recalls & reported problems

Medacta International SA · Fenestrated polyax. cann. ped. scr. 6x45,dual-diam · Class 2 · Implant

All models of M.U.S.T. Pedicle Screw System (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630030862120), received between Jun 25, 2025 and Nov 13, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • No Apparent Adverse Event1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 13, 2025No Apparent Adverse Event; No Apparent Adverse Event

    DURING A POSTERIOR SPINAL SURGERY, THE FIRST STEP REDUCER FAILED TO PROPERLY ATTACH TO THE SCREW HEAD. INSPECTION REVEALED THAT A METAL PLATE WAS MISSING FROM THE DISTAL PART OF THE INSTRUMENT. INTRAOPERATIVE FLUOROSCOPY CONFIRMED THAT THE PIECE WAS NOT RETAINED INSIDE THE PATIENT; IT MAY HAVE ALREADY BEEN MISSING PRIOR TO SURGERY. ON THE RIGHT L5 SCREW, THE SECOND STEP REDUCER CAUSED THE SET SCREW TO REPEATEDLY CROSS-THREAD, DAMAGING THE SCREW HEAD THREADS. THIS REDUCER ALSO LOST THE SAME SMALL METAL COMPONENT. AS A RESULT, THE SURGEON REPLACED THE RIGHT L5 SCREW AND BENT THE ROD INTO GREATER LORDOSIS TO PLACE THE SET SCREW WITHOUT A PERSUADER, AS NO FUNCTIONING REDUCTION INSTRUMENT WAS AVAILABLE. SURGERY COMPLETED WITH 30/40 MIN DELAY (TOTAL TIME SURGERY 3 HOURS 10 MINUTES). NO PIECES LEFT IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23543692 ↗
  • InjuryJun 25, 2025Break; Break

    MODIFIED EVENT DESCRIPTION: REVISION SURGERY PERFORMED ABOUT 4 MONTHS AFTER THE PRIMARY SURGERY DUE TO S1 SCREWS BREAKAGE (BELOW THE HEAD). PATIENT WITH GOOD QUALITY BONE, HEAVY PATIENT (AROUND 90KG). NO TRAUMA REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 22327510 ↗

Other Bone Cement, Posterior Screw Augmentation devices

All Bone Cement, Posterior Screw Augmentation devices →

Sources and updates