Skip to content

M.U.S.T. Pedicle Screw System 03.61.338Recalls & reported problems

Medacta International SA · Fenestrated polyax. cann. ped. scr. 7x45,dual-diam · Class 2 · Implant

All models of M.U.S.T. Pedicle Screw System (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630030862168), received between Nov 8, 2024 and Nov 8, 2024 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MONOSEGMENTAL LUMBAR SPINE SURGERY (L04, L05) WITH NEXTAR SYSTEM (VERSION 1.2.2). THE SURGEON USED THE DRILL, PLACED THE K-WIRES WITH THE NEXTAR DRILL GUIDE (3.2 MM, 40 MM DEPTH), FIRST ON THE RIGHT SIDE, THEN ON THE LEFT. AFTER THE PLACEMENT OF THE SCREWS, THE INTRA-OPERATIVE 3D SCAN SHOWS A LATERAL MISPLACEMENT OF BOTH RIGHT SCREWS. THE SURGEON REPLACED THE SCREWS ON THE RIGHT SIDE WITHOUT NAVIGATION, USING X-RAYS AND VISUALIZATION OF THE ANATOMICAL LANDMARKS. THE INSTRUMENTS WERE REMOVED AND THE SURGICAL WOUND WAS SUTURED. AFTER THE SURGERY, THE PATIENT WAS MOVED FROM THE OPERATION TABLE IN THE BED AND HAD A CIRCULATION COLLAPSE, THE AORTA AND VENA CAVA WERE REPAIRED IN AN EMERGENCY. THE BLEEDING WAS STOPPED SUCCESSFULLY AND THE PATIENT WAS FINE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20639941 ↗

Other Bone Cement, Posterior Screw Augmentation devices

All Bone Cement, Posterior Screw Augmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page