Skip to content

M.U.S.T. MINI Polyaxial Screw 03.75.001Recalls & reported problems

Medacta International SA · Polyaxial Screw 3.5 x 12 Full thread + nut · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630030863158), received between Apr 6, 2018 and Apr 2, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Migration1

Patient problems

  • Bone Fracture(s)1
  • Insufficient Information1
  • No Code Available1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 2, 2026Adverse Event Without Identified Device or Use Problem

    BONE FRACTURE DURING C2D T3 FIXATION AT THE BASE OF THE RIGHT C4/C6 SCREWS. PEDICLE PREPARATION WAS CARRIED OUT USING THE MYSPINE GUIDE, FOLLOWED BY DECOMPRESSION AND SCREW INSERTION. AFTER THE INSERTION OF THE ROD INTO THE PEDICLE SCREW TULIP, A FRACTURE OCCURRED AND THE SCREWS WERE REMOVED. FIXATION AT THE AFFECTED LEVEL WAS SKIPPED, AND THE PROCEDURE WAS CONCLUDED. NO CRITICAL DELAY.

    Patient problems: Bone Fracture(s)

    MAUDE report 24762670 ↗
  • InjuryMar 8, 2022Migration; Migration

    DURING A FOLLOW-UP VISIT, 1 MONTH AFTER THE PRIMARY SURGERY, IT WAS DISCOVERED THAT THE IMPLANTED RIGHT ROD WAS DISPLACED TO THE CRANIO-CAUDAL SIDE. DURING THE REVISION SURGERY, IT WAS ALSO FOUND THAT THE SET SCREWS WERE LOOSE ON THE LEFT SIDE AS WELL, SO, ALL THE SCREWS AND RODS WERE REPLACED.

    Patient problems: Insufficient Information

    MAUDE report 13693886 ↗
  • InjuryApr 6, 2018Break; Break

    THE SURGEON HAD DIFFICULTY TO TIGHTEN THREE SET SCREWS INTO THE PEDICLE SCREW TULIPS. AFTER MANY ATTEMPTS, THE THREAD OF THE THREE SET SCREWS BROKE (LOTS THAT BROKE OFF ARE UNKNOWN, 18 SCREWS WERE USED). THREE NEW SET SCREWS WERE USED IN ORDER TO COMPLETE THE SURGERY.

    Patient problems: No Code Available

    MAUDE report 7402533 ↗

Other Appliance, Fixation, Spinal Interlaminal devices

All Appliance, Fixation, Spinal Interlaminal devices →