Skip to content

MOTO Partial Knee System - Medial 02.18.IF2.08.LMRecalls & reported problems

Medacta International SA · Tibial insert fix S2 LM - 8mm · Class 2 · Implant

All models of MOTO Partial Knee System - Medial (46)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630030895852), received between Oct 24, 2022 and Jul 10, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Appears to Trigger Rejection1
  • Difficult to Insert1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Joint Laxity1
  • Medical device site infection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 10, 2025Difficult to Insert; Difficult to Insert

    IT WAS NOT POSSIBLE TO COUPLE THE PE 8 MM INSERT WITH THE FIX CEMENTED TIBIAL TRAY. A NEW 9 MM INSERT WAS USED INSTEAD. SURGERY COMPLETED WITH 20 MIN DELAY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 22467161 ↗
  • InjuryMay 16, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AT 1 YEAR AND 10 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 19328372 ↗
  • InjuryOct 24, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2019. ON (B)(6) 2020, THE PATIENT CAME IN REPORTING TIGHTNESS OF THE KNEE. THE SURGEON REVISED THE 10MM POLY WITH AN 8MM POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) 2022, THE PATIENT CAME IN FOR A POST-OP APPOINTMENT AND THE SURGEON OBSERVED THAT THE PATIENT WAS VERY VALGUS AND THE KNEE WAS VERY LAX. THE SURGEON REVISED ALL COMPONENTS AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Laxity

    MAUDE report 15657370 ↗

Other Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page