Skip to content

MOTO Partial Knee System - Medial 02.18.IF4.08.RMRecalls & reported problems

Medacta International SA · Tibial insert fix S4 RM - 8mm · Class 2 · Implant

All models of MOTO Partial Knee System - Medial (46)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07630030896101), received between Sep 19, 2018 and Aug 12, 2025 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Microbial Contamination of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available3
  • Adverse Event Without Identified Device or Use Problem2
  • Microbial Contamination of Device1
  • Positioning Problem1

Patient problems

  • Unspecified Infection4
  • Hematoma1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2025Positioning Problem; Positioning Problem

    TIBIAL INSERT 8MM COULD NOT BE COUPLED WITH THE TIBIAL TRAY; 9MM INSERT WAS USED AND THE SURGERY WAS COMPLETED SUCCESSFULLY. NO TIBIA RECUT NEEDED. THE SURGERY WAS PROLONGED BY ABOUT 20 MINUTES. DEVICES COUPLED IN BACKTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22781806 ↗
  • InjuryJun 13, 2022Appropriate Device Problem Term/Code Not Available

    THE PATIENT HAD A PRIMARY KNEE SURGERY ON (B)(6) 2022. ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN WAS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY. PRESENTLY, ON (B)(6) 2022, THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN WAS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE POLY AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 14675383 ↗
  • InjuryJun 10, 2022Appropriate Device Problem Term/Code Not Available

    THE SURGEON PLANS TO EXCHANGE THE EXISTING 10MM TIBIAL INSERT FOR A THICKER ONE DUE TO SOME VARUS/VALGUS INSTABILITY. REVISION SURGERY IS PLANNED FOR THE (B)(6) 2022.

    Patient problems: Insufficient Information

    MAUDE report 14652251 ↗
  • InjuryJan 25, 2022Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN 19 DAYS AFTER THE PRIMARY SURGERY DUE TO SIGNS OF INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE POLY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 13340928 ↗
  • InjuryJan 4, 2021Adverse Event Without Identified Device or Use Problem

    THE PATIENT CAME IN 1 MONTH AFTER THE PRIMARY SURGERY FOR A POST-OP APPOINTMENT AND IT WAS OBSERVED THAT A HEMATOMA HAD DEVELOPED. THE CAUSE OF THE HEMATOMA IS UNKNOWN. THE SURGEON PERFORMED AN WASHOUT AND REVISED THE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Hematoma

    MAUDE report 11106976 ↗
  • InjuryApr 23, 2020Microbial Contamination of Device

    THE PATIENT CAME IN, 3 MONTHS AFTER PRIMARY SURGERY, DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. THE SURGEON PERFORMED A WASHOUT AND REVISED THE POLY, AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 9989993 ↗
  • InjurySep 19, 2018Adverse Event Without Identified Device or Use Problem

    REVISION PERFORMED DUE TO INFECTION AFTER THREE MONTHS FROM PRIMARY SURGERY. THE SURGEON REVISED THE INSERT.

    Patient problems: Unspecified Infection

    MAUDE report 7888903 ↗

Other Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page