Skip to content

MOTO Partial Knee System - Medial 02.18.TF1.RMRecalls & reported problems

Medacta International SA · Tibial Tray Fix Cemented S1 RM · Class 2 · Implant

All models of MOTO Partial Knee System - Medial (46)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630030896705), received between Jun 24, 2021 and May 31, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Migration1

Patient problems

  • Inadequate Osseointegration1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 31, 2022Migration; Migration

    REVISION SURGERY AT 7 MONTHS FROM PRIMARY FOR TIBIAL IMPLANT SUBSIDENCE. THE SUBSIDENCE IS LIKELY DUE TO PATIENT WEIGHT AND INSUFFICIENT BONE STABILITY. ALL IMPLANTS WERE REVISED AND GMK SPHERE WAS IMPLANTED.

    Patient problems: Insufficient Information

    MAUDE report 14545852 ↗
  • InjuryJun 24, 2021Appropriate Device Problem Term/Code Not Available; Loosening of Implant Not Related to Bone-Ingrowth

    REVISION DUE TO ASEPTIC LOOSENING OF THE TIBIAL COMPONENT 1 YEAR AND 4 MONTHS AFTER PRIMARY. ALL DEVICES WERE EXPLANTED AND GMK SPHERE SYSTEM WAS IMPLANTED. IT MAY BE THAT THE ANTERIOR PART OF THE TIBIAL IMPLANT WAS NOT PERFECTLY SEATED ON THE CORTICAL BONE. THE PATIENT HAS A RATHER HIGH BMI.

    Patient problems: Inadequate Osseointegration

    MAUDE report 12058170 ↗

Other Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page