MODULITH® SLX-F2 (UDI 07630039100230)Recalls & reported problems

Storz Medical AG · Equipment for the extracorporeal induced shock wave therapy for stationary and mobile use. · Class 2

All models of MODULITH (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
2
Injuries reported
13
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 07630039100230), received between Apr 15, 2019 and Nov 6, 2025 — classified as 2 deaths, 13 injuries.

The most frequently reported specific device problem is “Misfocusing” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem13
  • Misfocusing2
  • Use of Device Problem2

Patient problems

  • Hematoma14
  • Death1
  • Insufficient Information1
  • Rupture1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 6, 2025Misfocusing; Use of Device Problem; Misfocusing; Use of Device Problem

    POST ESWL TREATMENT, CT SCAN CONFIRMED LARGE SUBCAPSULAR HEMATOMA, NO ACTIVE BLEED. ADMITTED WITH PAIN, BLOOD PRESSURE DROP, NO ACTIVE BLEED, NO TRANSFUSION.

    Patient problems: Hematoma

    MAUDE report 23486491 ↗
  • InjuryNov 6, 2025Misfocusing; Use of Device Problem; Misfocusing; Use of Device Problem

    ON (B)(6) 2025 CTU POST ESWL TREATMENT CONFIRMED LARGE SUBCAPSULAR HEMATOMA. ADMITTED ON (B)(6) 2025 FOR OBSERVATION, NO TRANSFUSION OR OTHER INTERVENTION, DISCHARGED ON (B)(6) 2025.

    Patient problems: Hematoma

    MAUDE report 23486490 ↗
  • InjuryJul 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CONFIRMED HEMATOMA POST ESWL TREATMENT. TREATING PHYSICIAN INITIALLY REPORTED PATIENT HAD MULTIPLE EMBOLIZATIONS AND EVENTUALLY WENT ON TO REQUIRE A NEPHRECTOMY. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED, INCLUDING DATE OF EVENT, ADDITIONAL MEDICAL TREATMENT, PROGNOSIS. HEMATOMA ARE A KNOWN SIDE EFFECT OF ESWL TREATMENT.

    Patient problems: Hematoma

    MAUDE report 22545247 ↗
  • InjuryJul 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CONFIRMED HEMATOMA POST ESWL TREATMENT, SEVERITY UNKNOWN. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED, INCLUDING DATE OF EVENT, MEDICAL TREATMENT OR PROGNOSIS. HEMATOMA ARE A KNOWN SIDE EFFECT OF ESWL TREATMENT.

    Patient problems: Hematoma

    MAUDE report 22545236 ↗
  • InjuryJul 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CONFIRMED HEMATOMA POST ESWL TREATMENT, SEVERITY UNKNOWN. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED, INCLUDING DATE OF THE EVENT, MEDICAL TREATMENT OR PROGNOSIS. HEMATOMA ARE A KNOWN SIDE EFFECT OF ESWL TREATMENT.

    Patient problems: Hematoma

    MAUDE report 22545230 ↗
  • InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED THAT A PATIENT PRESENTED WITH A HEMATOMA FOLLOWING ESWL TREATMENT ON (B)(6) 2022. PATIENT WAS ADMITTED TO THE HOSPITAL BUT ADDITIONAL INFORMATION CONCERNING ADMISSION AND DISCHARGE HAVE NOT BEEN PROVIDED BY THE PHYSICIAN.

    Patient problems: Hematoma

    MAUDE report 16855149 ↗
  • InjuryApr 16, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED THAT A PATIENT WAS DIAGNOSED WITH A RUPTURED KIDNEY FOLLOWING ESWL TREATMENT ON (B)(6) 2021 AND WAS ADMITTED TO THE HOSPITAL LATER THAT DAY. PATIENT WAS GIVEN 1 UNIT OF BLOOD IN THE ER AND HAS SUBSEQUENTLY BEEN GIVEN 2 MORE IN ICU. HEMOGLOBIN IS NOW STABLE AT 7 FOR OVER 24 HRS AND HAS BEEN DISCHARGED FROM ICU, BUT REMAINS HOSPITALIZED FOR FURTHER EVALUATION. PATIENT IS EXPECTED TO FULLY RECOVER.

    Patient problems: Hematoma; Rupture

    MAUDE report 11682500 ↗
  • InjuryMar 24, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED THAT A PATIENT PRESENTED WITH A HEMATOMA FOLLOWING ESWL TREATMENT ON (B)(6) 2021. THE PATIENT WAS ADMITTED TO THE HOSPITAL ON (B)(6) 2021 AND WAS DISCHARGED IN STABLE CONDITION ON (B)(6) 2021.

    Patient problems: Hematoma

    MAUDE report 11565737 ↗
  • DeathJan 25, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    WE LEARNED THAT A PATIENT PASSED AWAY DURING AN EXTRACORPOREAL SHOCK WAVE LITHOTRIPSY PROCEDURE AT (B)(6) HOSPITAL ADA, IN ADA, (B)(6) ON (B)(6) 2021.

    Patient problems: Insufficient Information

    MAUDE report 11227277 ↗
  • DeathNov 19, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ALLEGEDLY, FOLLOWING AN EXTRACORPOREAL SHOCK WAVE LITHOTRIPSY PROCEDURE AT (B)(6) SVC. IN (B)(6) ON (B)(6) 2019, THE PATIENT REPORTED TO THE HOSPITAL THAT EVENING. THE PATIENT WAS HEMODYNAMICALLY UNSTABLE. THE PATIENT WAS TRANSFUSED WITH 7 UNITS OF BLOOD AND GIVEN VASOPRESSORS. ANOTHER 8 UNITS OF BLOOD WERE TRANSFUSED. CAT SCAN REVEALED VERY LARGE INTRA- AND RETROPERITONEAL HEMATOMA OF THE RIGHT KIDNEY. WITHIN AN UNDISCLOSED TIME FRAME THE PATIENT WENT INTO RENAL FAILURE AND WAS PLACED ON DIALYSIS. SOON TO FOLLOW THE PATIENT DEVELOPED DIFFERENTIAL INTRAVASCULAR COAGULOPATHY AND THE FAMILY MADE THE DECISION TO TERMINATE HIS VITAL FUNCTIONS ON SUNDAY, (B)(6) 2019. THE PATIENT HAD A MEDICAL HISTORY THAT CONSISTED OF MORBID OBESITY (5'8" TALL (B)(6) LBS), HYPERTENSION, CIRRHOSIS, A PLATELET COUNT OF 58,000, AND A HEART VALVE REPLACEMENT.

    Patient problems: Death; Hematoma

    MAUDE report 9345162 ↗
  • InjuryOct 7, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ALLEGEDLY, THE PATIENT PRESENTED WITH A RENAL HEMATOMA FOLLOWING AN EXTRACORPOREAL SHOCK WAVE LITHOTRIPSY PROCEDURE AND WAS HOSPITALIZED. WE HAVE NOT BEEN ABLE TO OBTAIN ANY FURTHER DETAILS ABOUT THIS EVENT OR PATIENT CONDITION.

    Patient problems: Hematoma

    MAUDE report 9163309 ↗
  • InjurySep 4, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ALLEGEDLY, THE PATIENT PRESENTED WITH A HEMATOMA FOLLOWING AN EXTRACORPOREAL SHOCK WAVE LITHOTRIPSY PROCEDURE ON (B)(6) 2019. THE PATIENT WAS ADMITTED THAT DAY AND WAS TRANSFUSED. THE PATIENT WAS DISCHARGED ON (B)(6) 2019.

    Patient problems: Hematoma

    MAUDE report 8962893 ↗
  • InjurySep 4, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ALLEGEDLY, THE PATIENT PRESENTED WITH A HEMATOMA FOLLOWING AN EXTRACORPOREAL SHOCK WAVE LITHOTRIPSY PROCEDURE ON (B)(6) 2019. THE PATIENT WAS ADMITTED THAT DAY AND WAS TRANSFUSED. THE PATIENT WAS DISCHARGED ON (B)(6) 2019.

    Patient problems: Hematoma

    MAUDE report 8962891 ↗
  • InjuryMay 15, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ALLEGEDLY, THE PATIENT PRESENTED WITH A HEMATOMA FOLLOWING AN EXTRACORPOREAL SHOCK WAVE LITHOTRIPSY PROCEDURE. THE ONLY PATIENT INFORMATION PROVIDED WAS THAT PATIENT WAS ELDERLY AND FRAIL AND THE POST-OP SYMPTOMS WERE NOT SEVERE. WE HAVE NOT BEEN ABLE TO OBTAIN ANY FURTHER DETAILS ABOUT THIS EVENT OR PATIENT CONDITION.

    Patient problems: Hematoma

    MAUDE report 8612131 ↗
  • InjuryApr 15, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ALLEGEDLY, THE PATIENT PRESENTED WITH A RECTO-PERITONEAL BLEED THE DAY AFTER A LEFT EXTRACORPOREAL SHOCK WAVE LITHOTRIPSY PROCEDURE. THE PATIENT WAS ADMITTED AND WAS HOSPITALIZED FOR 8 DAYS. WE HAVE NOT BEEN ABLE TO OBTAIN ANY FURTHER DETAILS ABOUT THIS EVENT OR PATIENT CONDITION.

    Patient problems: Hematoma

    MAUDE report 8515172 ↗

FDA clearance and classification

Cleared through 510(k) K072788 on Dec 26, 2007 (decision: Substantially Equivalent). Submitted by KARL STORZ Endoscopy-America, Inc. as “Storz Modulith Lithotripter Slx-F2-F180”. The FDA review took 86 days. FDA record ↗

Cleared through Special 510(k) K070579 on Aug 30, 2007 (decision: Substantially Equivalent). Submitted by KARL STORZ Endoscopy-America, Inc. as “Storz Slx-F2 Storm Touch”. The FDA review took 182 days. FDA record ↗

Cleared through 510(k) K004037 on Feb 2, 2001 (decision: Substantially Equivalent). Submitted by Karl Storz Endoscopy as “X-Ray System Xr-Mx”. The FDA review took 36 days. FDA record ↗

The FDA regulates product code LNS (lithotriptor, extracorporeal shock-wave,urological) under 21 CFR 876.5990 as a Class II device. 21 CFR 876.5990 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the MODULITH® SLX-F2

Has the MODULITH® SLX-F2 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07630039100230), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the MODULITH® SLX-F2?

The FDA's MAUDE database lists 15 adverse event reports for this device, received between Apr 15, 2019 and Nov 6, 2025. In those reports, the most common specific device problems are “Misfocusing” (2) and “Use of Device Problem” (2); the most common patient problems are “Hematoma” (14), “Death” (1) and “Rupture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the MODULITH® SLX-F2?

2 reports are classified as deaths, 13 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the MODULITH® SLX-F2 FDA approved?

It was cleared through 510(k) K004037 on Feb 2, 2001; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the MODULITH® SLX-F2, and is it still sold?

The FDA's device database lists Storz Medical AG as the company responsible for it (brand MODULITH®, model SLX-F2). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Lithotriptor, Extracorporeal Shock-Wave,Urological devices

All Lithotriptor, Extracorporeal Shock-Wave,Urological devices →