MODULITH® SLX-F2 (REF 29000)Recalls & reported problems
Storz Medical AG · Equipment for the extracorporeal induced shock wave therapy for stationary and mobile use. · Class 2
All models of MODULITH (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 16
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 07630039100919), received between Oct 20, 2021 and Feb 19, 2026 — classified as 16 injuries.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem16
Patient problems
- Hematoma16
- Unspecified Kidney or Urinary Problem2
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
InjuryFeb 19, 2026Adverse Event Without Identified Device or Use Problem
InjuryMay 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROVIDER REPORTED A HEMATOMA OCCURRED AROUND A YEAR AGO POST ESWL. THE PATIENT INVOLVED THEN CONSULTED ANOTHER MEDICAL DOCTOR WHO ADVISED THAT THE KIDNEY SHOULD BE REMOVED. MANUFACTURER IS UNABLE TO CONFIRM THAT THE PATIENT HAS RECEIVED A NEPHRECTOMY. REPEATED ATTEMPTS TO COLLECT MORE INFORMATION HAVE BEEN UNSUCCESSFUL. THIS EVENT WAS REPORTED TO THE MANUFACTURER ON 4/25/2025.
Patient problems: Hematoma
MAUDE report 22071945 ↗InjuryMay 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A RENAL HEMATOMA, MEETING THE CRITERIA FOR BEING SERIOUS, OCCURRED AFTER ESWL TREATMENT ON (B)(6) 2025. THE PATIENT WAS SYMPTOMATIC AND ATTENDED A&E THE DAY AFTER ESWL WITH FLANK PAIN. CT WAS PERFORMED WHICH SHOWED THE HEMATOMA. PATIENT WAS ADMITTED TO THE HOSPITAL FOR 3 DAYS AND WAS GIVEN IV FLUIDS AND IV ANTIBIOTICS. PATIENT WAS MONITORED BUT REMAINED CLINICALLY STABLE. AT DISCHARGE THE PATIENT WAS GIVEN ANTIBIOTICS AND WAS ASKED TO RETURN 2 TIMES POST TREATMENT FOR BLOOD TESTS BUT WAS NOT READMITTED. PATIENT HAS BEEN ASKED TO FOLLOW UP THE END OF (B)(6).
Patient problems: Hematoma
MAUDE report 22064438 ↗InjuryJun 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SUBCAPSULAR HEMATOMA 43MM WITHOUT BLEEDING POST EXTRA CORPOREAL SHOCK WAVE TREATMENT.
Patient problems: Hematoma
MAUDE report 19574332 ↗InjuryJun 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SUBCAPSULAR HEMATOMA 35MM WITHOUT ACTIVE BLEEDING POST EXTRACORPOREAL SHOCK WAVE TREATMENT.
Patient problems: Hematoma
MAUDE report 19574252 ↗InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OEM STORZ MEDICAL WAS INFORMED BY MEDICAL PERSONNEL FROM TREATING HOSPITAL THAT A PATIENT WAS DIAGNOSED WITH HEMATOMA POST ESWL TREATMENT AND A NEPHRECTOMY WAS PERFORMED. NO INFORMATION CONCERNING THE PATIENT OR THEIR PRE AND POST ESWL CONDITION HAS BEEN FORTHCOMING FROM THE TREATING PHYSICIAN OR FACILITY. THE DEVICE HAS BEEN EVALUATED AND NO ISSUES WERE DETECTED. 11/28/22 ADDITIONAL INFORMATION RECEIVED: ESWL TREATMENT ON (B)(6) 2022, LEFT VESICOURETERIC JUNCTION, 1 STONE 5MM, AT ENERGY LEVELS 1.0 - 5.0 FOR A TOTAL OF 3000 SHOCKS. PATIENT ADMITTED (B)(6) 2022 FOR COMPLAINT OF LOIN PAIN. CT PERFORMED (B)(6) 2022 SHOWS LRFT PERINEPHRIC HEMATOMA.
Patient problems: Hematoma; Unspecified Kidney or Urinary Problem
MAUDE report 16857340 ↗InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OEM STORZ MEDICAL AG WAS INFORMED BY MEDICAL PERSONNEL FROM TREATING HOSPITAL THAT A PATIENT WAS DIAGNOSED WITH HEMATOMA POST ESWL TREATMENT. PATIENT ADMITTED FOR LEFT RENAL STONE TREATMENT WAS COMPLICATED BY LEFT ACUTE SUBCAPSULAR PERI-RENAL HEMATOMA WITH MASS EFFECT (IN POST OPERATIVE PERIOD).
Patient problems: Hematoma
MAUDE report 16857064 ↗InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OEM STORZ MEDICAL AG WAS INFORMED BY MEDICAL PERSONNEL FROM (B)(6) HOSPITAL THAT A PATIENT WAS DIAGNOSED WITH HEMATOMA POST ESWL TREATMENT. THE FACILITY AND PHYSICIAN HAS PROVIDED LIMITED INFORMATION CONCERNING THE PATIENT PRE AND POST ESWL TREATMENT. HEMATOMA SIZE WAS BETWEEN 2.2 AND 3CM. PATIENT DID NOT REQUIRE TRANSFUSION BUT WAS ADMITTED FOR STRICT MONITORING. THE DEVICE HAS BEEN EVALUATED AND NO ISSUES WERE DETECTED.
Patient problems: Hematoma
MAUDE report 16856573 ↗InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OEM STORZ MEDICAL WAS INFORMED BY MEDICAL PERSONNEL FROM TREATING HOSPITAL THAT A PATIENT WAS DIAGNOSED WITH HEMATOMA POST ESWL TREATMENT. NO INFORMATION CONCERNING THE PATIENT OR THEIR PRE AND POST ESWL CONDITION HAS BEEN FORTHCOMING FROM THE TREATING PHYSICIAN OR FACILITY. THE DEVICE HAS BEEN EVALUATED AND NO ISSUES WERE DETECTED. (B)(6) 2022 ADDITIONAL INFORMATION RECEIVED: ESWL TREATMENT (B)(6) 2022, RIGHT KIDNEY STONE 5MM AT ENERGY LEVEL 1.0 - 5.0 FOR TOTAL OF 3000 SHOCKS. PATIENT TREATED FOR SEVERE RIGHT LOIN PAIN, ADMITTED TO WARD AND CT SHOWS RIGHT PERINEPHRIC HEMATOMA WITH ACTIVE CONTRAST EXTRAVASATION UP TO 4.8CM. BLOOD PRESSURE BORDERLINE, HB DROPPED FROM 13.5 TO 9.3.
Patient problems: Hematoma
MAUDE report 16856558 ↗InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OEM STORZ MEDICAL AG WAS INFORMED BY MEDICAL PERSONNEL FROM TREATING HOSPITAL THAT A PATIENT WAS DIAGNOSED WITH HEMATOMA POST ESWL TREATMENT. THE FACILITY AND PHYSICIAN HAS PROVIDED LIMITED INFORMATION CONCERNING THE PATIENT PRE AND POST ESWL TREATMENT. HEMATOMA SIZE WAS BETWEEN 2.2 AND 3CM. PATIENT DID NOT REQUIRE TRANSFUSION BUT WAS ADMITTED FOR STRICT MONITORING. THE DEVICE HAS BEEN EVALUATED AND NO ISSUES WERE DETECTED.
Patient problems: Hematoma
MAUDE report 16855163 ↗InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OEM STORZ MEDICAL AG WAS INFORMED BY MEDICAL PERSONNEL FROM TREATING HOSPITAL THAT A PATIENT WAS DIAGNOSED WITH HEMATOMA POST ESWL TREATMENT. THE FACILITY AND PHYSICIAN HAS PROVIDED LIMITED INFORMATION CONCERNING THE PATIENT PRE AND POST ESWL TREATMENT. HEMATOMA SIZE WAS BETWEEN 2.2 AND 3CM. PATIENT DID NOT REQUIRE TRANSFUSION BUT WAS ADMITTED FOR STRICT MONITORING. THE DEVICE HAS BEEN EVALUATED AND NO ISSUES WERE DETECTED.
Patient problems: Hematoma
MAUDE report 16855162 ↗InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OEM STORZ MEDICAL AG WAS INFORMED BY MEDICAL PERSONNEL FROM TREATING HOSPITAL THAT A PATIENT WAS DIAGNOSED WITH HEMATOMA POST ESWL TREATMENT. THE FACILITY AND PHYSICIAN HAS PROVIDED LIMITED INFORMATION CONCERNING THE PATIENT PRE AND POST ESWL TREATMENT. PATIENT WAS ADMITTED FOR TREATMENT ON (B)(6) 2023 FOR LEFT RENAL STONE WHICH WAS COMPLICATED BY 7X5X9CM PERINEPHRIC HEMATOMA WITH HAEMORRHAGIC SHOCK.
Patient problems: Hematoma
MAUDE report 16855161 ↗InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OEM STORZ MEDICAL AG WAS INFORMED BY MEDICAL PERSONNEL FROM TREATING HOSPITAL THAT A PATIENT WAS DIAGNOSED WITH HEMATOMA POST ESWL TREATMENT. THE FACILITY AND PHYSICIAN HAS PROVIDED LIMITED INFORMATION CONCERNING THE PATIENT PRE AND POST ESWL TREATMENT. HEMATOMA SIZE WAS BETWEEN 2.2 AND 3CM. PATIENT DID NOT REQUIRE TRANSFUSION BUT WAS ADMITTED FOR STRICT MONITORING. THE DEVICE HAS BEEN EVALUATED AND NO ISSUES WERE DETECTED.
Patient problems: Hematoma
MAUDE report 16855160 ↗InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OEM STORZ MEDICAL WAS INFORMED BY MEDICAL PERSONNEL FROM TREATING HOSPITAL THAT A PATIENT WAS DIAGNOSED WITH HEMATOMA POST ESWL TREATMENT. NO INFORMATION CONCERNING THE PATIENT OR THEIR PRE AND POST ESWL CONDITION HAS BEEN FORTHCOMING FROM THE TREATING PHYSICIAN OR FACILITY. THE DEVICE HAS BEEN EVALUATED AND NO ISSUES WERE DETECTED. 11/28/22 ADDITIONAL INFORMATION RECEIVED: ESWL TREATMENT ON (B)(6) 2022 LEFT MIDDLE CALYX STONE 13MM X 3MM AT ENERGY LEVEL 1.0 - 6.0 FOR TOTAL OF 3000 SHOCKS. PATIENT UNDERWENT CT ON (B)(6) 2022 AND DIAGNOSED WITH LEFT PERINEPHRIC HEMATOMA (18CM X 10CM) WITH ACTIVE BLEEDING AT LOWER POLE OF KIDNEY. PERSISTENT TACHYCARDIA AND PAIN WITH HB DROP.
Patient problems: Hematoma
MAUDE report 16855154 ↗InjuryMay 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OEM STORZ MEDICAL WAS INFORMED BY MEDICAL PERSONNEL FROM TREATING HOSPITAL THAT A PATIENT WAS DIAGNOSED WITH HEMATOMA POST ESWL TREATMENT AND A NEPHRECTOMY WAS PERFORMED. NO INFORMATION CONCERNING THE PATIENT, OR THEIR PRE AND POST ESWL CONDITION HAS BEEN FORTHCOMING FROM THE TREATING PHYSICIAN OR FACILITY. THE DEVICE HAS BEEN EVALUATED AND NO ISSUES WERE DETECTED. ON (B)(6) 2022 ADDITIONAL INFORMATION RECEIVED: ESWL TREATMENT ON (B)(6) 2022 TO 14MM RIGHT RENAL STONE AT ENERGY LEVEL 1.0 - 6.0 FOR A TOTAL OF 3000 SHOCKS. PATIENT COMPLAINED OF LOIN PAIN AND HYPOTENSION AND HB DROPPED. CT OF ABDOMEN AND PELVIS WITH CONTRAST DONE ON (B)(6) 2022 SHOWS RIGHT ACUTE PERINEPHRIC HEMATOMA WITH ACUTE BLOOD SEEN EXTENDING TO RIGHT PARACOLIC GUTTER AND DOWN TO PELVIS. A 7MM ARTERIAL ENHANCING FOCUS IS NOTED AT POSTERIOR ASPECT OF UPPER POLE OF RIGHT KIDNEY, WHICH PERSISTS ON PORTOVENOUS PHASE AND SHOWED SIGNIFICANT CHANGE OF SIZE OR SHAPE. A FEW TINY ARTERIALLY ENHANCING FOCI AT POSTERIOR ASPECT WITHOUT DEFINITE POOLING SEEN ON PORTOVENOUS PHASES. ACTIVE CONTRAST ESTRAVASATION CANNOT BE EXCLUDED. RIGHT RENAL EMBOLIZATION PERFORMED ON (B)(6) 20/22.
Patient problems: Hematoma; Unspecified Kidney or Urinary Problem
MAUDE report 16855150 ↗InjuryOct 20, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED THAT A PATIENT PRESENTED WITH A HEMATOMA FOLLOWING ESWL TREATMENT ON (B)(6) 2021. THE PATIENT WAS ADMITTED TO THE HOSPITAL ON (B)(6) 2021 AND WAS DISCHARGED IN STABLE CONDITION ON (B)(6) 2021.
Patient problems: Hematoma
MAUDE report 12668553 ↗
FDA clearance and classification
Cleared through 510(k) K201001 on Mar 30, 2021 (decision: Substantially Equivalent). Submitted by Karl Storz Endoscopy America as “Modulith SLX-F2”. The FDA review took 348 days. FDA record ↗
Cleared through Special 510(k) K070579 on Aug 30, 2007 (decision: Substantially Equivalent). Submitted by KARL STORZ Endoscopy-America, Inc. as “Storz Slx-F2 Storm Touch”. The FDA review took 182 days. FDA record ↗
The FDA regulates product code LNS (lithotriptor, extracorporeal shock-wave,urological) under 21 CFR 876.5990 as a Class II device. 21 CFR 876.5990 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the MODULITH® SLX-F2
Has the MODULITH® SLX-F2 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07630039100919), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the MODULITH® SLX-F2?
The FDA's MAUDE database lists 16 adverse event reports for this device, received between Oct 20, 2021 and Feb 19, 2026. In those reports, the most common patient problems are “Hematoma” (16) and “Unspecified Kidney or Urinary Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the MODULITH® SLX-F2?
None of the reports are classified as deaths, 16 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the MODULITH® SLX-F2 FDA approved?
It was cleared through Special 510(k) K070579 on Aug 30, 2007; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the MODULITH® SLX-F2, and is it still sold?
The FDA's device database lists Storz Medical AG as the company responsible for it (brand MODULITH®, model SLX-F2). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Lithotriptor, Extracorporeal Shock-Wave,Urological devices
- Dornier MedTech Delta IIIDornier MedTech GmbH · Lithotriptor, Extracorporeal Shock-Wave,Urological27 reports
- Dornier MedTech Compact SigmaDornier MedTech GmbH · Lithotriptor, Extracorporeal Shock-Wave,Urological17 reports
- MODULITH® SLX-F2Storz Medical AG · Lithotriptor, Extracorporeal Shock-Wave,Urological2 deaths15 reports
- PIEZOLITH 3000 TRIPLE-FOCUS LITHOTRIPSY SYSTEM COMPLETE P3000002United Service Solutions, LLC · Lithotriptor, Extracorporeal Shock-Wave, Urological1 death1 report
- Modularis Variostar 01157200Siemens Healthcare GmbH · Lithotriptor, extracorporeal shock-wave,urological1 report
- Dornier Delta III K1045938Dornier MedTech GmbH · Lithotriptor, Extracorporeal Shock-Wave, Urological1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page