Anatomic Shoulder Prosthesis 04.01.0003Recalls & reported problems
Medacta International SA · Standard Humeral Diaphysis – Cementless, Size 8 · Class 2 · Implant
All models of Anatomic Shoulder Prosthesis (29)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630040706025), received between Mar 7, 2019 and Dec 9, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
1 other recall cover devices cleared under the same FDA submission (K170910); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated1
- Device Slipped1
- Osseointegration Problem1
- Positioning Problem1
Patient problems
- Aseptic loosening1
- Failure of Implant1
- Inadequate Osseointegration1
- Joint Dislocation1
- Joint Disorder1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 9, 2025Device Dislodged or Dislocated; Osseointegration Problem; Device Dislodged or Dislocated; Osseointegration Problem
REVISION SURGERY DUE TO JOINT SUBLUXATION AT ABOUT 5 YEARS 10 MONTHS FROM PRIMARY. NO PAIN REPORTED. IT WAS THEREFORE DECIDED TO PERFORM SURGERY TO INCREASE THE HEIGHT OF THE INLAY IN ORDER TO RESTORE SUFFICIENT STABILITY. DURING THE REVISION SURGERY IT WAS FOUND THAT THE STEM WAS ALSO LOOSE. THE SURGEON DECIDED TO REVISE THE STEM (SAME SIZE, CEMENTLESS TO CEMENTED), METAPHYSIS (+0 TO +9 MM) AND LINER (+3MM TO +6 MM E-CROSS CONSTRAINED) AT 155 DEGREES. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Aseptic loosening; Inadequate Osseointegration; Joint Dislocation
MAUDE report 23758551 ↗InjuryAug 5, 2022Positioning Problem
REVISION SURGERY PERFORMED 4 YEARS AFTER THE PRIMARY SURGERY DUE TO GLENOSPHERE POSITION TOO SUPERIOR (IMPINGEMENT) AND STEM WAS NOT STABLE. ALL IMPLANTS REVISED. THE PATIENT ALREADY HAD PREVIOUS REVISION SURGERY DUE TO INFECTION.
Patient problems: Failure of Implant
MAUDE report 15174131 ↗InjuryMar 7, 2019Device Slipped
THE PATIENT CAME IN COMPLAINING OF INSTABILITY PROBABLY BECAUSE OF LOOSE SOFT TISSUE TENSION 2 WEEKS AFTER A FIRST REVISION OF DJO PRODUCTS WITH MEDACTA HUMERUS DEVICES. THE SURGEON REVISED THE MEDACTA LINER, METAPHYSIS AND DIAPHYSIS AND THE SURGERY WAS COMPLETED SUCCESSFULLY. INITIAL SURGERY WAS A PRIMARY REVERSE WITH DJO. THE PATIENT'S DJO PROSTHESIS BASEPLATE CENTRAL SCREW BROKE. THE PATIENT HAS BEEN REVISED TO (2ND SURGERY) MEDACTA HUMERUS AND TORNIER BASEPLATE AND GLENOSPHERE. THIRD PROCEDURE (THIS CASE IN QUESTION) THE BASEPLATE AND GLENOSPHERE IN WERE TORNIER, THEN WAS REVISED TO TORNIER. MEDACTA HUMERUS WAS REMOVED AND PLACED IN A MORE RETROVERTED POSITION AND UPSIZED WITH ALL NEW COMPONENTS.
Patient problems: Joint Disorder
MAUDE report 8399585 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K170910. One submission often covers many models, so check the product and lot details in each record.
- Z-0775-2020 · Nov 4, 2019 · TerminatedMedacta Shoulder SYSTEM HUMERAL ANATOMICAL METAPHYSIS, REF 04.01.0023 through 04.01.0055, all sizes - Product Usage: The Medacta Anatomic Shoulder Prosthesis is indicated for treat
Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page