Shoulder System 04.01.0183Recalls & reported problems
Medacta International SA · Short Humeral diaphysis - cementless - 10 · Class 2 · Implant
All models of Shoulder System (63)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07630040715911), received between Sep 4, 2020 and Apr 2, 2024 — classified as 5 injuries.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Device Appears to Trigger Rejection1
- Device Dislodged or Dislocated1
- Loose or Intermittent Connection1
- Osseointegration Problem1
Patient problems
- Bone Fracture(s)2
- Inadequate Osseointegration1
- Joint Disorder1
- Medical device site infection1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryApr 2, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
REVISION SURGERY WAS PERFORMED AT 2 YEARS AND 2 MONTHS FROM THE PRIMARY SURGERY DUE TO DISLOCATION OF THE HUMERAL HEAD FROM THE GLENOID. INTRAOPERATIVELY, IT WAS DETERMINED THAT THE PATIENT'S HUMERUS WAS FRACTURED AND THE SURGEON OPTED TO USE A FRACTURE STEM FROM EXACTECH TO COMPLETE THE CASE. THE SURGEON CONVERTED THE PATIENT FROM AN ANATOMIC SHOULDER TO A REVERSE SHOULDER AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Bone Fracture(s)
MAUDE report 19018506 ↗InjuryJul 27, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
AT ABOUT 1 YEAR AND 2 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN AND RESULTED POSITIVE FOR CURTIBACTERIUM ACNES. THE SURGEON PERFORMED AN ANALGESIC TREATMENT AND REMOVED ALL THE IMPLANTS. THE PATIENT IS UNDER MONITORING.
Patient problems: Medical device site infection
MAUDE report 17404478 ↗InjuryNov 25, 2022Osseointegration Problem; Osseointegration Problem
THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE SHORT STEM AND THE CAUSE IS UNKNOWN. AT ABOUT 11 MONTHS FROM PRIMARY THE SURGEON REVISED THE METAPHYSIS, DIAPHYSIS, AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration
MAUDE report 15864356 ↗InjuryOct 6, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
DIRECTLY AFTER THE PRIMARY SHOULDER SURGERY, WHEN THE PATIENT WAS IN POST-OP RECOVERY, IT WAS OBSERVED THAT THE TIP OF THE SHORT STEM BROKE THROUGH THE LATERAL CORTEX. IT HAS BEEN REPORTED THAT THE PATIENT HAD INCREDIBLY HARD BONE AND IT WAS DIFFICULT TO INSERT THE STEM. THE REVISION SURGERY WAS COMPLETED SUCCESSFULLY EXPLANTING THE POLY, STEM, AND METAPHYSIS WHILE A STANDARD LENGTH STEM, +9 METAPHYSIS AND A PLUS 3 POLY WERE IMPLANTED.
Patient problems: Bone Fracture(s)
MAUDE report 15550761 ↗InjurySep 4, 2020Loose or Intermittent Connection
REVISION SURGERY PERFORMED DUE TO DISLOCATION CAUSED BY STEM SUBSIDENCE ONE MONTH AFTER THE PRIMARY SURGERY. PROBABLY THE STEM WAS INSERTED A BIT TOO DEEPLY DURING THE PRIMARY SURGERY AND ALSO SUBSIDED AFTER THE PRIMARY SURGERY, WHAT CAUSED THAT THE STEM WAS FINALLY 3MM BELOW RESPECT TO THE PLANNED POSITION. THE LINER +0MM WAS REVISED WITH +3MM.
Patient problems: Joint Disorder
MAUDE report 10493615 ↗
Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page