MySpine 7.0725Recalls & reported problems
Medacta International SA · MySpine-MC drill guide L05 · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630040717557), received between Dec 10, 2019 and Apr 1, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Material Protrusion/Extrusion” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available2
- Material Protrusion/Extrusion1
Patient problems
- Bone Fracture(s)1
- Pain1
- Perforation1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 1, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
TLIF SURGERY WAS SUCCESSFULLY PERFORMED AT L5-S1 WITH MYSPINE MC GUIDE. A FEW DAYS LATER, CT SCAN REVEALED THAT UPPER-LEVEL L5 SCREW HAD PERFORATED THE CRANIAL INTERVERTEBRAL DISC. THE SURGEON POINTED OUT THAT THE LOWER PART OF THE GUIDE SLEEVE HAD BEEN SHORTENED AND THERE WAS AN EXCESSIVE SPACE BETWEEN THE GUIDE AND THE BONE MODEL.
Patient problems: Perforation
MAUDE report 21737328 ↗InjuryMay 28, 2020Appropriate Device Problem Term/Code Not Available
DURING THE PRIMARY SPINE SURGERY, WHILE THE SURGEON WAS IMPLANTING THE LEFT SCREW AT L5, A LATERAL BREACH OF THE LEFT L5 VERTEBRAE OCCURRED. THERE WAS A 15-MINUTE DELAY AND THE SURGEON FREE-HANDED THE SCREW. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE SURGEON FOLLOWED THE PROCEDURE AND DID NOT EXPERIENCE ANY ISSUES WITH ANY OF THE OTHER SCREWS IN THIS CASE OR HIS PREVIOUS CASE. THERE WAS NO BREACH UNTIL THE SURGEON FULLY IMPLANTED THE SCREW. NO FITTING ISSUE HAS BEEN REPORTED BETWEEN BONE MODEL/PATIENT'S BONE AND GUIDES.
Patient problems: Bone Fracture(s)
MAUDE report 10095418 ↗InjuryDec 10, 2019Material Protrusion/Extrusion
REVISION SURGERY PERFORMED DUE TO BACK PAIN. THE TRAJECTORY OF THE PEDICLE SCREW IN L5 RIGHT WAS NOT CORRECT AND THE TIP PROTRUDED OUT OF THE PEDICLE, IT WAS REPOSITIONED THE SAME SCREW. CASE CODE: (B)(6).
Patient problems: Pain
MAUDE report 9443539 ↗
Other Pedicle Screw Placement Guide devices
- MySpine 7.0726Medacta International SA · Pedicle Screw Placement Guide3 reports
- MySpine 7.0724Medacta International SA · Pedicle Screw Placement Guide2 reports
- MySpine 7.0704Medacta International SA · Pedicle Screw Placement Guide2 reports
- MySpine 7.0723Medacta International SA · Pedicle Screw Placement Guide2 reports
- My Spine Low Profile Placement Guides 7.0412Medacta International SA · Pedicle screw placement guide1 report
- My Spine Low Profile Placement Guides 7.0524Medacta International SA · Pedicle screw placement guide1 report
- My Spine S2-Alar/Alar-Iliac Pedicle Screw Placement Guides 7.0727Medacta International SA · Pedicle Screw Placement Guide1 report
- MySpine Unilateral Guide 7.1027Medacta International SA · Pedicle Screw Placement Guide1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.