Skip to content

M.U.S.T. Pedicle Screw System 03.56.503Recalls & reported problems

Medacta International SA · MUST MC Cross-Connector 19-29mm adjustable · Class 2 · Implant

All models of M.U.S.T. Pedicle Screw System (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630040728799), received between Feb 9, 2023 and Feb 9, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Material Deformation1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 9, 2023Material Deformation; Loosening of Implant Not Related to Bone-Ingrowth; Material Deformation; Loosening of Implant Not Related to Bone-Ingrowth

    PRIMARY SURGERY PERFORMED 6 MONTHS AGO (JULY 2022), L4-L5-S1 TLIF. SIX MONTHS LATER, L4-L5 PRESENTED BONY FUSION, HOWEVER, L5-S1 RESULTED IN PSEUDOARTHROSIS WITH S1 SCREW LOOSENING. THE CROSS-CONNECTOR WAS LOCATED AT L4-L5. AT THE REVISION SURGERY, THE CROSS-CONNECTOR WAS NOT STABLE ALREADY BEFORE REMOVING THE SET SCREWS. L4 AND L5 PEDICLE SCREWS WERE OK WITHOUT ANY SIGNS OF LOOSENING AND L4-L5 WAS COMPLETELY FUSED, BUT THE CROSS-CONNECTOR AT L4-L5 SHOWED LOOSENING; WHILE IN L5-S1 WAS NON-UNION AND S1 SCREWS WERE LOOSE.

    Patient problems: Insufficient Information

    MAUDE report 16332042 ↗

Other Thoracolumbosacral Pedicle Screw System devices

All Thoracolumbosacral Pedicle Screw System devices →