Skip to content

M.U.S.T. Pedicle Screw System 03.68.921Recalls & reported problems

Medacta International SA · Pre-Contoured Ø5.5 Ti - G · Class 2 · Implant

All models of M.U.S.T. Pedicle Screw System (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630542719363), received between Aug 28, 2024 and Aug 28, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 28, 2024Migration; Migration

    PRIMARY SURGERY WAS PERFORMED AT TH9-S2AI. XLIF SURGERY WAS PERFORMED AT L1-L5, TLIF SURGERY WAS PERFORMED AT L5-S. REVISION SURGERY 6 DAYS AFTER THE PRIMARY SURGERY WAS PERFORMED DUE TO THE RODS COMING OUT FROM THE MUST S2AI PEDICLE SCREW HEAD. THE PEDICLE SCREWS AT BOTH SIDES OF SA1 WERE REPLACED WITH ANOTHER COMPETITOR'S SCREW. THE ROD AT THE LEFT SIDE WAS ALSO REPLACED BECAUSE THE LENGTH OF THE ROD WAS SHORT. IN THE X-RAYS THAT WERE TAKEN DURING THE PRIMARY SURGERY, THE ROD ON THE RIGHT SIDE SEEMED TO BE SHORT, BUT BY CHANGING THE ANGLE OF FLUOROSCOPY, IT WAS CONFIRMED THAT THE ROD WAS CORRECTLY PROTRUDED ABOUT 5 MM, AND THE SURGERY WAS FINISHED. THE PATIENT AND THE CONSTRUCT IS STABLE AFTER THE REVISION PROCEDURE.

    Patient problems: Insufficient Information

    MAUDE report 20085911 ↗

Other Thoracolumbosacral Pedicle Screw System devices

All Thoracolumbosacral Pedicle Screw System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page