Skip to content

SportMeds Instruments 05.05.10.0120Recalls & reported problems

Medacta International SA · Cannulated Screwdriver Shaft T25 · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07630971281790), received between Sep 2, 2024 and Jun 18, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Joint Laxity1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 18, 2025Fracture; Fracture

    THE TIP OF THE CANNULATED SCREWDRIVER SHAFT T25 BROKE DURING SURGERY. NO PIECES LEFT IN THE PATIENT. BACK UP USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22284233 ↗
  • MalfunctionNov 15, 2024Fracture; Fracture

    THE TIP OF THE CANNULATED SCREWDRIVER SHAFT T25 BROKE BECAUSE THE NITINOL WAS BENT. NO PIECES LEFT IN THE PATIENT. THE SCREWDRIVER HAS BEEN USED IN COMBINATION WITH A COMPETITOR'S NITINOL WIRE.

    Patient problems: Joint Laxity

    MAUDE report 20693613 ↗
  • MalfunctionOct 24, 2024Fracture; Fracture

    THE TIP OF THE CANNULATED SCREWDRIVER SHAFT T25 BROKE DURING ACL SURGERY DURING MECTASCREW C INTRODUCTION INTO THE FEMORAL TUNNEL. A FRAGMENT REMAINED IN THE PATIENT. THE NITINOL GUIDE WAS DEFORMED. MECTASCREW C HAD NO DAMAGE AND HAS BEEN IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20522209 ↗
  • MalfunctionSep 2, 2024Break; Break

    THE TIP OF THE CANNULATED SCREWDRIVER SHAFT T25 AND THE MECTASCREW C BROKE, PROBABLY DUE TO MALPOSITIONING OF THE NITINOL WIRE. THE GRAFT WAS UNDAMAGED AND THE SURGEON USED A NEW SCREW AND ANOTHER SCREWDRIVER TO SUCCESSFULLY COMPLETE THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20118027 ↗

Other Accessories, Arthroscopic devices

All Accessories, Arthroscopic devices →

Sources and updates