M.U.S.T. Midline Cortical (MC) Screw System 03.58.234Recalls & reported problems
Medacta International SA · MUST MC Screw Ø5.5x40 cannulated · Class 2 · Implant
All models of M.U.S.T. Midline Cortical (MC) Screw System (17)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07630971293465), received between Apr 25, 2025 and Jul 16, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Malposition of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Malposition of Device1
Patient problems
- Bone Fracture(s)1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2025Malposition of Device; Adverse Event Without Identified Device or Use Problem; Malposition of Device; Adverse Event Without Identified Device or Use Problem
PLIF SURGERY WAS PERFORMED AT L3-L5 WITH THE MYSPINE MC GUIDE ON (B)(6) 2025. THE DAY AFTER THE SURGERY, CT SCAN WAS TAKEN AND IT APPEARED THAT THE ENTIRE GUIDE HAD BEEN PLACED TILTED TO THE LEFT (MEANING THAT THE TRAJECTORY OF THE SCREWS ARE VERY DIFFERENT FROM THE PLANNING). NO FITTING ISSUE WITH THE GUIDES WAS EXPERIENCED DURING SURGERY. THE PATIENT IS CURRENTLY ASYMPTOMATIC, AND THE SURGEON IS CURRENTLY WAITING FOR THE BONE FUSION. WEAK BONE QUALITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22520976 ↗InjuryApr 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SPINAL FIXATION WAS PERFORMED AT L4-L5 WITH MYSPINE MC GUIDE. WHEN THE SURGEON TRIED TO INSERT THE PEDICLE SCREW (5.5MM X 40MM) AT RIGHT SIDE OF L5, THE PATIENT BONE BROKE. PATIENT'S BONE QUALITY WEAK. HOLE PREPARATION: 5MM UNDER TAP 6MM SAME TAP. THE SURGEON CHANGED THE TRAJECTORY AND INSERTED THE PEDICLE SCREW (5.5MM X 35MM). THE REASON FOR CHANGING THE SCREW LENGTH WAS THAT THE TRAJECTORY WAS CHANGED, RESULTING IN A SHORTER TRAJECTORY.
Patient problems: Bone Fracture(s)
MAUDE report 21908585 ↗
Other Thoracolumbosacral Pedicle Screw System devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page