M.U.S.T. Midline Cortical (MC) Screw System 03.58.244Recalls & reported problems
Medacta International SA · MUST MC Screw Ø6x40 cannulated · Class 2 · Implant
All models of M.U.S.T. Midline Cortical (MC) Screw System (17)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07630971293564), received between Apr 5, 2024 and Nov 18, 2025 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
Patient problems
- Implant Pain2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryNov 18, 2025Fracture; Fracture
PATIENT UNDERWENT SURGERY FOR VERTEBRAL FUSION L3-L5. 6 MONTHS AFTER PRIMARY THE PATIENT REPORTED PAIN. SURGEON PROCEEDED WITH REVISION AND FOUND 3 BROKEN SCREWS (L3-L4 RIGHT SIDE AND L5 ON THE LEFT SIDE). BREAKAGE OCCURRED AT THE BASE OF THE TULIP IN ALL SCREWS. IT WAS REPORTED THAT PATIENT'S BONE WAS ARTHRITIC AND THAT VERTEBRAL FUSION FAILED. CONSTRUCT WAS REVISED, SURGEON REPLACED THE SCREW IN L3-L5 ON THE RIGHT SIDE, REMOVED THE SCREW FROM LEFT L5 WHICH WAS NOT REPLACES, AND ADDED A SCREW IN LEFT S1.
Patient problems: Implant Pain
MAUDE report 23583105 ↗InjuryMay 23, 2025Fracture; Fracture
PATIENT UNDERWENT PRIMARY SURGERY AT L4-L5 (PLIF, CBT TRAJECTORY). MYSPINE CASE CODE: H-SUG-OPS-RS-24061941. REVISION SURGERY PERFORMED AT ABOUT 2 MONTHS FROM PRIMARY DUE TO PAIN, CAUSE OF PAIN WAS IDENTIFIED AS A COLLAPSED L5 VERTEBRA. DURING SURGERY THE LEFT L5 PEDICLE SCREW WAS FOUND BROKEN AROUND THE SCREW HEAD. THE RIGHT L5 PEDICLE SCREW WAS NOT BROKEN. THE FIXATION RANGE WAS EXTENDED FROM L4 TO S2 WITH THE ADDITION OF A SCREW IN L4. NO PATIENT FALLS WERE REPORTED AFTER THE PRIMARY SURGERY.
Patient problems: Implant Pain
MAUDE report 22076622 ↗MalfunctionApr 5, 2024Fracture; Fracture
THE SET SCREW OF MUST MC SCREW Ø6X40 CANNULATED CRACKED DURING FINAL TIGHTENING. THE SET SCREW COULD NOT BE DISENGAGED, SO THE SURGEON CUT AND REMOVED THE ROD, REVISED THE PEDICLE SCREW AND ENGAGED A NEW ROD. THERE WAS 20 MIN OF DELAY (TOTAL SURGERY TIME 3 HOURS).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 19049140 ↗
Other Thoracolumbosacral Pedicle Screw System devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page