Thoratec Switzerland GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07640135140115), received between Sep 15, 2022 and Mar 9, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Infusion or Flow Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Infusion or Flow Problem3
  • Mechanical Problem1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 9, 2026Mechanical Problem

    IT WAS REPORTED THAT THE POWER BUTTON ON THE BACK OF THE CENTRIMAG CONSOLE WAS LOOSE. THIS WAS DISCOVERED DURING A REGULAR EQUIPMENT CHECK, AND THE DEVICE WAS TAKEN OUT OF SERVICE FOR EVALUATION. THE FLOW PROBE WAS RETURNED AND DETERMINED TO HAVE BEEN DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24552679 ↗
  • MalfunctionMar 15, 2024Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT THE PATIENT, WHO WAS ON A PEDIMAG, HAD SUCTION EVENTS ON THE EVENING ON (B)(6) 2023. NO ALARMS WERE TRIGGERED. LOG FILES WERE SUBMITTED FOR REVIEW AND CAPTURED A LOW FLOW EVENT SHOWING A DROP TO 0 LITERS PER MINUTE (LPM), ASSOCIATED WITH A LOSS OF FLOW SIGNAL. THE EVENT WAS MUTED OR CLEARED WITHIN 5 SECONDS. THE FLOW DROPPED TO APPROXIMATELY 1.0 LPM (LOWEST VALUE RECORDED WAS 0.9 LPM.) THIS DID NOT RESULT IN AN ALARM BECAUSE THE ALARM THRESHOLD WAS SET TO 0.5 LPM AT THE TIME. THERE WAS ANOTHER LOW FLOW SHOWING 0 LPM, AGAIN ASSOCIATED WITH A LOSS OF FLOW SIGNAL. THIS ALARM WAS ALSO MUTED/CLEARED. AGAIN, THERE WAS ANOTHER LOW FLOW SHOWING 0 LPM, AGAIN ASSOCIATED WITH A LOSS OF FLOW SIGNAL, WHICH WAS ALSO MUTED/CLEARED. RELATED MANUFACTURER REFERENCE NUMBER: 3003306248-2024-00442, 3003306248-2024-00439, 3003306248-2024-00441

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18915429 ↗
  • InjuryFeb 16, 2024Infusion or Flow Problem; No Apparent Adverse Event; Infusion or Flow Problem; No Apparent Adverse Event

    IT WAS REPORTED THAT THE SCREEN SHOWED NO FLOW. THE PATIENT WENT INTO ARREST AND ADVANCED RESUSCITATION MANEUVERS WERE INITIATED, AS WELL AS HIGH INOTROPIC SUPPORT. THEY PERFORMED AN EMERGENCY CHANGE; THEY WENT TO THE BACKUP MOTOR AND BACKUP CONSOLE. FLOW WAS REESTABLISHED AND THE PATIENT CAME OUT OF ARREST. IT WAS NOTED THAT THE PATIENT HAD BEEN ON EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) SUPPORT SINCE (B)(6) 2023 FOR TRACHEOESOPHAGEAL FISTULA/ACUTE RESPIRATORY DISTRESS SYNDROME (ARDS). THE PATIENT WAS HEMODYNAMICALLY STABLE. RELATED MANUFACTURER REFERENCE NUMBER: 2916596-2024-00789 (CENTRIMAG CONSOLE). RELATED MANUFACTURER REFERENCE NUMBER: 3003306248-2024-00407 (CENTRIMAG MOTOR). RELATED MANUFACTURER REFERENCE NUMBER: 3003306248-2024-00408 (CENTRIMAG BLOOD PUMP).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18725743 ↗
  • MalfunctionSep 15, 2022Infusion or Flow Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT THE PATIENT WAS IN THE INTENSIVE CARE UNIT (ICU) WITH EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO). THE PATIENT PRESENTED HEMODYNAMIC DECOMPENSATION THAT REQUIRED INOTROPIC SUPPORT, THE PATIENT RECOVERED WITHOUT PROBLEM. THE PATIENT PRESENTED LOSS OF FLOWS ON THREE OCCASIONS. THE CONSOLE DID NOT REGISTER THE FLOW. THE DECISION WAS MADE TO CHANGE THE CONSOLE AND BACKUP FLOW PROBE AND THE FLOW WAS NORMALIZED. THE ENGINE WAS NOT CHANGED. THERE WERE NO CONSEQUENCES FOR THE PATIENT. THE PATIENT WAS STABLE. MANUFACTURER REPORT NUMBER OF CONSOLE: 3003306248-2022-12509.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15434276 ↗

Other Control, Pump Speed, Cardiopulmonary Bypass devices