icotec Pedicle System, Nut Screw, Ti 16-57-41000Recalls & reported problems

icotec AG · icotec Pedicle System, Nut Screw, Ti · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07640140431994), received between Sep 29, 2023 and May 31, 2026 — classified as 6 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Mechanical Problem2
  • Defective Device1

Patient problems

  • Failure of Implant6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 31, 2026Mechanical Problem

    MFR REPORT #: 3006493760-2026-00003 A PATIENT NOTICED INCREASING PAIN WHICH LED TO A FOLLOW-UP IMAGING (DATE NOT REPORTED) WHERE IT WAS NOTICED THAT NUT SCREWS LOOSENED. A REVISION SURGERY WAS PERFORMED ((B)(6) 2026). DURING THE REVISION IT WAS DETECTED THAT TWO PEDICLE SCREWS WERE BROKEN. THE SURGEON DECIDED TO REPLACE THE PEDICLE SYSTEM. EARLIER ((B)(6) 2023), THE PATIENT HAD IMPLANTED AN ICOTEC PEDICLE SYSTEM WITH A REVISION ON (B)(6) 2024 (MDR NO. (B)(4) WHERE THE SURGEON DECIDED TO IMPLANT A 5.5 MM DIAMETER PEDICLE ROD FROM A THIRD PARTY MANUFACTURER (OFF-LABEL). DURING THIS REVISION THE ROD WAS SECURED IN THE PEDICLE SCREWS USING NEW ICOTEC NUT SCREWS.

    Patient problems: Failure of Implant

    MAUDE report 25345301 ↗
  • InjuryMay 31, 2026Mechanical Problem

    MFR REPORT #: 3006493760-2026-00001. A PATIENT WAS IMPLANTED WITH AN ICOTEC PEDICLE SCREW SYSTEM ((B)(6) 2025). POST-SURGERY THREE NUT SCREWS (SET SCREWS) LOOSENED (EXACT DATE WERE NOT REPORTED). X-RAY IMAGING SHOWS THAT THE PEDICLE SYSTEM WAS USED TOGETHER WITH TWO ILIAC SCREWS (FROM AN UNKNOWN MANUFACTURER). A REVISION SURGERY WAS PERFORMED AND IMPLANTS WERE REPLACED ((B)(6) 2026).

    Patient problems: Failure of Implant

    MAUDE report 25344876 ↗
  • InjuryApr 10, 2026Detachment of Device or Device Component

    A PATIENT NOTICED INCREASING PAIN WHICH LED TO A FOLLOW-UP IMAGING (DATE NOT REPORTED) WHERE IT WAS NOTICED THAT NUT SCREWS LOOSENED. A REVISION SURGERY WAS PERFORMED (B)(6) 2026). DURING THE REVISION IT WAS DETECTED THAT TWO PEDICLE SCREWS WERE BROKEN. THE SURGEON DECIDED TO REPLACE THE PEDICLE SYSTEM. EARLIER (B)(6) 2023), THE PATIENT HAD IMPLANTED AN ICOTEC PEDICLE SYSTEM WITH A REVISION ON (B)(6) 2024 (MDR NO. ON . WHERE THE SURGEON DECIDED TO IMPLANT A 5.5 MM DIAMETER PEDICLE ROD FROM A THIRD-PARTY MANUFACTURER (OFF-LABEL). DURING THIS REVISION THE ROD WAS SECURED IN THE PEDICLE SCREWS USING NEW ICOTEC NUT SCREWS.

    Patient problems: Failure of Implant

    MAUDE report 24843209 ↗
  • InjuryMar 6, 2026Detachment of Device or Device Component

    A PATIENT WAS IMPLANTED WITH AN ICOTEC PEDICLE SCREW SYSTEM ((B)(6) 2025). POST-SURGERY THREE NUT SCREWS (SET SCREWS) LOOSENED (EXACT DATE WERE NOT REPORTED). X-RAY IMAGING SHOWS THAT THE PEDICLE SYSTEM WAS USED TOGETHER WITH TWO ILIAC SCREWS (FROM AN UNKNOWN MANUFACTURER). A REVISION SURGERY WAS PERFORMED AND IMPLANTS WERE REPLACED ((B)(6) 2026).

    Patient problems: Failure of Implant

    MAUDE report 24532096 ↗
  • InjuryDec 5, 2023Defective Device; Defective Device

    A PATIENT WAS TREATED WITH THE ICOTEC PEDICLE SYSTEM IN A SURGERY. DURING A POST-OP CHECK 7-10 DAYS AFTER, IT WAS DETECTED THAT SIX OUT OF EIGHT NUT SCREWS HAD LOOSENED. A REVISION SURGERY HAD TO BE CARRIED OUT. THE LOOSENED NUT SCREWS AND THE PEDICLE SCREWS WERE REMOVED AND REVISED.

    Patient problems: Failure of Implant

    MAUDE report 18266154 ↗
  • InjurySep 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    A PATIENT WAS TREATED WITH THE ICOTEC PEDICLE SYSTEM IN A SURGERY ((B)(6) 2022). DURING A POST-OP CHECK, IT WAS DETECTED THAT FOUR NUT SCREWS HAD LOOSENED. A REVISION SURGERY HAD TO BE CARRIED OUT ((B)(6) 2023). DURING THE REVISION SURGERY, IT WAS REALIZED THAT ONE PEDICLE SCREW HAD BROKEN. THE BROKEN PEDICLE SCREW WAS NOT REMOVED. THE LOOSENED NUT SCREWS WERE REVISED AND THE BROKEN SCREW REMAINED IN THE PATIENT. NO ADDITIONAL LEVEL WAS TREATED.

    Patient problems: Failure of Implant

    MAUDE report 17842817 ↗