Guidewire 0.018 4/320cm angled (Straub Medical®) Version 1.0Recalls & reported problems
Straub Medical AG · The Rotarex™ S and Aspirex™ S Guidewire has a LubriSkin PTFE-coated nitinol core wire with a gold-plated Tungsten tip. The guidewire is used to initially cross through the area to be treated. To be used with Straub Medical catheters. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07640142810261), received between Nov 5, 2020 and Aug 26, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K172315, K211738); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Break1
- Component Missing1
- Detachment of Device or Device Component1
- Overheating of Device1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Stenosis1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 26, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A THROMBECTOMY AND ATHERECTOMY PROCEDURE, THE GUIDEWIRE WAS ALLEGEDLY BROKEN AND SEPARATED INTO TWO PARTS. IT WAS FURTHER REPORTED THAT THE SEPARATED PART OF THE GUIDEWIRE WAS REMOVED USING SNARE DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20061962 ↗MalfunctionNov 5, 2020Fracture; Fracture; Overheating of Device; Component Missing
LEFT GROIN TREATING RIGHT LEG. NO ACCESS ISSUES, NO CONCERN ABOUT CASE. CTO DISTAL SFA, DISEASE IN COMMON FEMORAL. CROSSED CTO EASILY. USED IVIS. USED ROTAREX IN COMMON FEM, MOVED TO DISTAL SFA. 2/3 THE WAY THROUGH HIT "ROCK", 80% STENOSED. BOGGED DOWN. CLUTCHED OUT TWICE, THEN PUSHED PAST THE STENOSIS TO TREAT FARTHER DOWN. BACK AND FORTH MOTION PECK MOTION. MOVED AWAY AND DOWN TPT DRUNK. NOTICED SLOWER ASPIRATION BUT WORKING NORMAL. REMOVED DEVICE, SET ASIDE, DIDN'T NOTICE TIP WAS MISSING FROM DEVICE. ROTAREX CATHETER FRACTURE WHEN CROSSING THE BIFURCATION. TROUBLE PUTTING BALLOON INTO SHEATH. LOOKED AT FLOURO - SAW HELIX INSIDE OF SHEATH, STILL ON WIRE BUT DISTAL TO SHEATH. USED SNARE TO REMOVE DEVICE. FRAGMENT WAS 1/3 OF HELIX LENGTH. ALL GOT PUSHED OUT DISTAL TO SHEATH WHEN BALLOON WAS PUT IN. ADDED ~3MIN TO CASE TO RETREIVE DEVICE. 7FR COOK ANSEL SHEATH, NOT A HIGH FLEX. WHEN WIRE PINNED, WHEN FINISHING UP AND GETTING READY TO PULL DEVICE OUT (FINISHING DISTAL TPT TRUNK), THE WIRE STARTED TO FEEL WARM. DEVICE RUNNING WHEN REMOVED THROUGH CTO.
Patient problems: No Clinical Signs, Symptoms or Conditions; Stenosis
MAUDE report 10791987 ↗MalfunctionNov 5, 2020Fracture; Physical Resistance/Sticking
STRAIGHT FORWARD CASE. 6F 45 CM FLEX ANSEL UP AND OVER SHEATH WAS USED. A .018 STAUB WIRE ADVANCED ONCE ROTAREX USAGE DETERMINED. ROTAREX 6F 135 REF #(B)(4) LOT# 200541 PRESENTED FOR ISR AND IN BETWEEN THE 2 SFA/POP STENTS. DOC AWARE OF OFF LABEL INDICATION. DR. CARTER TREATED THE 2 STENTS AND THE GAP IN BETWEEN STENTS. DURING PROCEDURE THE ROTAREX CLUTCHED OUT ONE TIME, DOC BACKED UP AND REENGAGED LESION. TERRIFIC RESULTS. REMOVED ROTAREX AND INSPECTED. INTACT. UPON INTRODUCING AN ULTRAVERSE 5X150 PTA, PTA WOULDN'T ADVANCE. NOTICED THE PROXIMAL TIP OF WIRE HAD FRACTURED. THE FRACTURED PORTION OF THE PROXIMAL 30CM PORTION OF WIRE WAS IN THE PROXIMAL SFA. PHYSICIAN SNARED WIRE PORTION WITH NO COMPLICATION. CASE WAS COMPLETED WITH .035 WIRE, PTA AND STENTING. EXCELLENT RESULT. PHYSICIAN SAID ABSOLUTELY NO ISSUE WITH DEVICE. COMLAINED IS GUIDEWIRE WHICH IS PART OF SET ROTAREXS 6F 135CM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10791442 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K172315, K211738. One submission often covers many models, so check the product and lot details in each record.
- Z-1244-2025 · Feb 5, 2025 · Open, ClassifiedRotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239
Other Wire, Guide, Catheter devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page