RHA® Redensity RHA® Redensity US 1ml Schott/2 aig. Terumo TW30G1/2 (REF 940056)Recalls & reported problems
Teoxane SA · "One box of RHA® Redensity contains two blisters and one instruction for use. Each blister contains a 1mL prefilled syringe and two needles (30G 1/2). Box Packaging (ID 07640173230083) is defined as unit of sales and identified with a serial number." · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07640173231653), received between Jul 10, 2023 and Aug 9, 2024 — classified as 5 injuries.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method4
- Off-Label Use3
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Foreign Body Embolism4
- Dyspnea1
- Embolism/Embolus1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryAug 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PRIMEVIGILANCE NOTIFIED TEOXANE OF AN ADVERSE EVENT ON 10-JUL-2024. THE PATIENT WAS INJECTED ON (B)(6) 2024 WITH: 1 ML OF RHA REDENSITY IN THE PERIORAL LINES. THE INJECTION WAS DONE WITH A 31G NEEDLE (CURRENT COMPLAINT). 1.2 ML OF RHA 4 IN THE CHEEKBONE. THE INJECTION WAS DONE WITH A 27G NEEDLE (LINKED COMPLAINT: (B)(4). APPROXIMATELY THREE WEEKS AFTER THE INJECTION, ON (B)(6) 2024, THE PATIENT RECEIVED MMR VACCINE (VACCINE AGAINST MEASLES, MUMPS, AND RUBELLA). 2 WEEKS LATER, ON (B)(6) 2024, THE PATIENT NOTED SWELLING. AT THIS TIME HCP INITIATED PATIENT ON PREDNISONE 40 MG, COLCHICINE, AND KEFLEX DAILY. THE DAY AFTER, ON (B)(6) 2024 SHE VISITED THE EMERGENCY ROOM DUE TO SWELLING THAT CAUSED DIFFICULTY BREATHING AND RECEIVED IV STEROID. THE PATIENT NOTICED SOME IMPROVEMENT OVER THE WEEK AFTER THE TREATMENT BUT ON (B)(6)2024 THE PATIENT REPORTED HARD NODULES AND SWELLING. THE SWELLING WAS MORE CONCENTRATED IN AREAS WHERE THE FILLER WAS PLACED (UNKNOWN WHETHER RHA REDENSITY OR RHA 4), AND ACCORDING TO THE ( HCP), SOME GENERALIZED SWELLING WAS PRESENT AS WELL. BASED ON THE INFORMATION THE RELATEDNESS OF THE DIFFICULTY TO BREATH WITH THE INJECTION OF TEOXANE PRODUCTS COULD NOT BE EXCLUD
Patient problems: Dyspnea
MAUDE report 19947220 ↗InjuryAug 1, 2024Off-Label Use; Improper or Incorrect Procedure or Method; Off-Label Use; Improper or Incorrect Procedure or Method
PRIMEVIGILANCE NOTIFIED TEOXANE OF AN ADVERSE EVENT ON 03-JUL-2024. A PATIENT WAS INJECTED WITH RHA REDENSITY INTO THE FOREHEAD ON (B)(6) 2024. THE INJECTION WAS ADMINISTERED VERY SUPERFICIALLY USING AN UNSPECIFIED NEEDLE. TWO DAYS AFTER THE INJECTION, ON (B)(6) 2024, THE PATIENT DEVELOPED REDNESS, WITH BLISTERS AND WHITEHEADS ON THE FOREHEAD. THE FOLLOWING DAY, ON (B)(6) 2024, THE REDNESS APPEARED TO DARKEN. THE INJECTOR SUGGESTED THE POSSIBILITY OF VASCULAR OCCLUSION AND INJECTED HYALURONIDASE (HYLENEX) AS A TREATMENT. DESPITE REMINDERS, NO UPDATES WERE PROVIDED ABOUT THE PATIENT, THUS THE CASE WAS CLOSED AS IS.
Patient problems: Embolism/Embolus; Foreign Body Embolism
MAUDE report 19879957 ↗InjuryMay 16, 2024Off-Label Use; Improper or Incorrect Procedure or Method; Off-Label Use; Improper or Incorrect Procedure or Method
PRIMEVIGILANCE NOTIFIED TEOXANE OF AN ADVERSE EVENT ON (B)(6) 2024. THE PATIENT WAS INJECTED ON (B)(6) 2024 WITH A TOTAL OF 2 ML OF RHA REDENSITY IN THE MARIONETTE LINES (0.25 ML ON EACH SIDE), CHIN (0.5 ML), AND TEAR TROUGH (1 ML). THE INJECTIONS WERE DONE WITH A 25G CANNULA. DURING THE INJECTION, THE SKIN AROUND THE RIGHT MARIONETTE BLANCHED. THE AREA WAS MASSAGED, AND THE CAPILLARY REFILL WAS CHECKED AND WAS SLOW (TAKING 2-3 SECONDS). THE PATIENT DID NOT REPORT PAIN IN THE AREA BUT WAS UNDER CONSTANT MEDICAL OBSERVATION UNTIL THE CAPILLARY REFILL RETURNED TO NORMAL. AFTER THE PATIENT LEFT THE CLINIC, THE INJECTOR CONTINUED TO CLOSELY FOLLOW-UP WITH THE PATIENT WITH PHOTOS AND VIDEOS. IN THE EVENING, THE PATIENT PRESENTED WITH MOTTLING AND DELAYED CAPILLARY REFILL AND IMMEDIATELY WENT BACK TO THE CLINIC TO BE TREATED WITH 900 I.U. OF HYALURONIDASE (HYLENEX). THE EVENT WAS DIAGNOSED AS A VASCULAR OCCLUSION OF MODERATE INTENSITY. ON (B)(6) 2024, THE CAPILLARY REFILL RETURNED TO NORMAL LIMITS, BUT THE AFFECTED AREA WAS STILL LOOKING MOTTLED. THUS, UNTIL (B)(6) 2024, THE INJECTOR CONTINUED TO TREAT THE PATIENT WITH HYALURONIDASE (A TOTAL OF 4700 UNITS IN THREE DAYS). IN ADDITION, T
Patient problems: Foreign Body Embolism
MAUDE report 19330071 ↗InjuryApr 26, 2024Off-Label Use; Improper or Incorrect Procedure or Method; Off-Label Use; Improper or Incorrect Procedure or Method
PRIMEVIGILANCE NOTIFIED TEOXANE OF AN ADVERSE EVENT ON (B)(6) 2024. THE PATIENT WAS INJECTED ON (B)(6) 2024 WITH 1 ML OF RHA REDENSITY IN THE LIPS. THE INJECTION WAS DONE WITH A CANNULA USING THE RETROTRACING TECHNIQUE. BEFORE THE PROCEDURE, THE PATIENT RECEIVED LIDOCAINE AND AN EPINEPHRINE INJECTION NEAR TO THE ORAL COMMISSURES. THE MEDICAL HISTORY OF THE PATIENT MENTIONED HYPOTHYROIDISM AND SEVERAL MEDICATIONS: LEVOTHYROXINE (SYNTHROID) AND PLEXUS IN ADDITION TO MULTIVITAMINS, VITAMIN C, AND VITAMIN D FIVE MINUTES AFTER THE INJECTION, ON (B)(6) 2024, THE INJECTOR NOTICED THAT LIPS STARTED TURNING PALE AND CAPILLARY REFILL SLOWED. THE PATIENT FELT ''NUMB''. A VASCULAR OCCLUSION WAS DIAGNOSED. THE PATIENT WAS UNDER OBSERVATION FOR 1.5 HOURS. DURING THIS TIME, AS A TREATMENT, THE INJECTOR USED 1 ML OF HYALURONIDASE (HYLENEX) ON THE LEFT SIDE OF THE PERIORAL AREA VIA CANNULA AND GAVE A WARM COMPRESS. AFTER THAT, THE AFFECTED AREA AND CAPILLARY REFILL IMPROVED. THE INJECTOR MASSAGED THE AFFECTED AREA, AND THEN THE AREA BECAME WHITE AGAIN, BUT THE CAPILLARY REFILL WAS GOOD. AN ADDITIONAL 0.5 ML OF HYALURONIDASE (HYLENEX) WAS USED, AND THE LIPS RETURNED TO THEIR PINK COLOR. AROUND 1.5
Patient problems: Foreign Body Embolism
MAUDE report 19189303 ↗InjuryJul 10, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
THIS OCCURRED OUTSIDE OF THE USA, IN GERMANY. A NATUROPATH INJECTOR NOTIFIED TEOXANE OF AN ADVERSE EVENT ON (B)(6) 2023. A FEMALE PATIENT WAS INJECTED ON (B)(6) 2023 AT ABOUT 18:30 WITH 0.3 ML OF RHA REDENSITY, SUPERFICIALLY INTO ACNE SCARS ON THE LEFT SIDE OF THE CHIN. IMMEDIATELY AFTER THE INJECTION, THE PATIENT PRESENTED A CHANGE IN SKIN COLOUR DESCRIBED AS "BLUE" AND A "TYPICAL LIVID DISCOLORATION". THE INJECTOR RECOGNIZED AN EMBOLISM AND IMMEDIATELY TREATED THE AREA WITH HYALURONIDASE (1500 IU). THE NEXT MORNING, ON (B)(6) 2023, THE REACTION PROGRESSED TO "CLEAR REDDISH-BLUE PAINFUL MUCOSAL LESION" AND BLISTERS THAT LOOK LIKE "THREE MINI PIMPLES" ON THE INSIDE OF THE LIP MUCOSA. THE INJECTOR TREATED WITH ANOTHER ROUND OF HYALURONIDASE (1500 IU). AS OF (B)(6) 2023, THE ADVERSE REACTION WAS PARTIALLY RESOLVED. DESPITE REMINDERS, NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Foreign Body Embolism
MAUDE report 17287560 ↗
Other Implant, Dermal, For Aesthetic Use devices
- JUVEDERM VOLUMA XC 94640Allergan, Inc. · Implant, Dermal, For Aesthetic Use672 reports
- JUVEDERM ULTRA XC 94154Allergan, Inc. · Implant, Dermal, For Aesthetic Use541 reports
- JUVEDERM VOLLURE XC 95661Allergan, Inc. · Implant, Dermal, For Aesthetic Use380 reports
- JUVEDERM ULTRA PLUS XC 94155Allergan, Inc. · Implant, Dermal, For Aesthetic Use319 reports
- JUVÉDERM VOLBELLA XC 95834Allergan, Inc. · Implant, Dermal, For Aesthetic Use187 reports
- Skinvive 98307Allergan, Inc. · Implant, Dermal, For Aesthetic Use183 reports
- Revanesse Lips+ with Lidocaine 1.2 mL 40149Prollenium Medical · Implant, Dermal, For Aesthetic Use72 reports
- Radiesse (+) Lidocaine 8063M0K2Merz North America, Inc. · Implant, Dermal, For Aesthetic Use61 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page