Skip to content

Neo Medical Pedicle screw kit RD-00-40-PBRecalls & reported problems

Neo Medical SA · Rod kit pre-bent 40 mm (2x) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07640177820181), received between Apr 2, 2025 and Apr 2, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 2, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    EVENT DESCRIPTION: SET SCREW LOOSENING AND ROD SLIPPAGE (DISCONNECTION BETWEEN THE TULIP AND THE ROD) OCCURRED A FEW WEEKS POST OP (ORIGINAL POST OP ABOVE AND 3 WEEK POST OP BELOW) ON ONE SIDE OF A L2/L3 LLIF FUSION WITH POSTERIOR FIXATION. AFTER REVIEW, IT IS EXPECTED THAT THE ROD USED FOR THE SEGMENT WAS 5MM TOO SHORT. COMPLAINT INVESTIGATION: NEO MEDICAL GATHERED THE FOLLOWING INFORMATION BY INTERVIEWING SURGEONS AND SALES REP: "A TOO SHORT ROD WAS USED". THIS TYPE OF ISSUE IS KNOWN AND DISPLAYED IN THE SURGICAL TECHNIQE WITH A FEW MM OVERHANG IS DESCRIBED. BY ANALYSING THE MRI AND X-RAYS IMAGES WE CAN CONCLUDE THAT FROM MEASURING THE OVERHANG A 0.5MM DIFFERENCE ON ONE SIDE OF THE ROD CAUSED ONE SIDE TO SLIP (1.5MM OVERHANG) WHILE THE OTHER SIDE DID NOT (2.0 MM OVERHANG). NEO MEDICAL ALSO LOOKED AT THE BATCH PRODUCTION RECORDS FOR THE AFFECTED LOTS, THE REPORT CONCLUDED NO RELEVANT DEVIATIONS RELATED TO PRODUCT QUALITY. THE MOST PROBABLE ROOT CAUSE OF THE FAILURE WAS THAT THE ROD USED WAS TOO SHORT RESULTING DUE TO A LACK OF CONTACT POINT BETWEEN ROD AND SCREW HEAD ("THE TULIP") RESULTING IN A DISCONNECTION OF THE CONSTRUCT. THIS IS THE FIRST TIME THAT A TOO SHORT ROD HAVE BEEN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21758861 ↗

Other Thoracolumbosacral Pedicle Screw System devices

All Thoracolumbosacral Pedicle Screw System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page