Neo Medical Pedicle Screw System SC-60-30-PERecalls & reported problems
Neo Medical SA · Pedicle screw kit 2x(Ø6.0 x 30mm) · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07640177821027), received between Jul 30, 2025 and Jul 30, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 30, 2025Break; Break
DURING A MINIMALLY INVASIVE REVISION SURGERY DUE TO AN ADJACENT LEVEL FRACTURE, THE INITIAL CONSTRUCT WAS TO BE REMOVED AND REPLACED. WHILE ATTEMPTING TO REMOVE THE LOCKING SCREWS FROM THE ORIGINALLY PLACED 10-SCREW CONSTRUCT, ONE OF THE SCREWS PROVED DIFFICULT TO EXTRACT. IT WAS EVENTUALLY DISCOVERED THAT THE LOCKING SCREW CONTAINED A BROKEN SET SCREWDRIVER TIP. DESPITE THIS COMPLICATION, THE ENTIRE CONSTRUCT WAS SUCCESSFULLY REMOVED WITH CONSIDERABLE EFFORT. THE INITIAL SURGERY, WHICH WAS ALSO PERFORMED MINIMALLY INVASIVELY, WAS SUPPORTED ON-SITE BY A COMPANY REPRESENTATIVE. NO IRREGULARITIES WERE NOTED DURING THE FINAL FIXATION OF THE SET SCREW AT THAT TIME. IT WAS REPORTED THIS EVENT DIDN'T IMPACT THE PATIENT OUTCOME. ALSO, NO PATIENT PAIN RELATED TO IMPLANT WAS REPORTED IN BETWEEN THE INITIAL SURGERY AND THE REVISION SURGERY. HOWEVER, WE LEARNED THAT THE REVISION SURGERY TOOK LONGER THAN EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22666371 ↗
Other Thoracolumbosacral Pedicle Screw System devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page