Dexter Monopolar Hook Dexter L6Recalls & reported problems

Distalmotion SA · Dexter L6 Monopolar Hook · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07640184840622), received between Nov 21, 2025 and Nov 27, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Therapy Delivered to Incorrect Body Area” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Apparent Adverse Event1
  • Therapy Delivered to Incorrect Body Area1

Patient problems

  • Burn(s)1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 27, 2025Therapy Delivered to Incorrect Body Area; Therapy Delivered to Incorrect Body Area

    DURING A CHOLECYSTECTOMY, A SMALL HOLE WAS OBSERVED IN THE INSULATION OF THE TIP SLEEVE OF THE MONOPOLAR HOOK INSTRUMENT. WHILE APPLYING ENERGY TO DISSECT THE GALLBLADDER, SPARKING WAS NOTED FROM THE HOLE, WHICH RESULTED IN ONE SMALL, LINEAR, CAUTERIZATION MARK TO THE LIVER CAPSULE. THE SAME INSTRUMENT WAS BEING USED FOR APPROXIMATELY 19 MINUTES WITHOUT ISSUES AND WAS EXTRACTED AND RE-INSERTED JUST BEFORE THE EVENT. NO FRICTION WAS REPORTED WHEN PASSING THE MONOPOLAR HOOK THROUGH THE TROCAR. SINCE THIS FINAL SURGICAL TASK REQUIRED LESS THAN 5 MINUTES, THE SURGEON DECIDED TO COMPLETE THE PROCEDURE WITHOUT REPLACING THE INSTRUMENT. THIS RESULTED IN A FEW SIMILAR, SMALL, CAUTERIZATION MARKS ON THE LIVER CAPSULE, CREATED UNDER FULL VISUAL CONTROL AND WITH DELIBERATE, CONTROLLED INSTRUMENT MOVEMENTS. PATIENT HARM WAS LIMITED TO MINOR LINEAR CAUTERIZATION OF THE LIVER CAPSULE, VISIBLE ON VIDEO. NO LACERATION OR HAEMORRHAGE WERE SEEN; NO HEPATIC STRUCTURES WERE INJURED. NO ADDITIONAL INTERVENTIONS WERE PERFORMED OR PLANNED. ENERGY SOURCE USED WAS ETHICON MEGADYNE, SET AT 30.

    Patient problems: Burn(s)

    MAUDE report 23666057 ↗
  • MalfunctionNov 21, 2025No Apparent Adverse Event; No Apparent Adverse Event

    DURING TISSUE COAGULATION IN A REFLUXSTOP¿ DEVICE IMPLANTATION PROCEDURE, IT WAS REPORTED THAT AFTER APPROXIMATELY 40 MINUTES OF USING THE MONOPOLAR HOOK, A SPARK TRAIL INDICATED A DEFECT IN THE TIP SLEEVE. THERE WAS NO HARM TO THE PATIENT. THE INSTRUMENT WAS IMMEDIATELY REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND WITHOUT INCIDENT. THERE WERE NO REPORTS OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23616367 ↗

Other Electromechanical Surgical System Wherein The Surgeon Annd The User Interface Can Be In The Sterile Field devices

All Electromechanical Surgical System Wherein The Surgeon Annd The User Interface Can Be In The Sterile Field devices →