Skip to content

PERLA ® TL TLF-DS 45 25-SRecalls & reported problems

SpineArt SA · 25D SCREW Ø4.5 L25 · Class 2 · Implant

All models of PERLA TL (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 15, 2024Class IIOpen, Classified

    Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew

    Z-1758-2024 · Root cause: Process design · Recall details · FDA record ↗

Other Appliance, Fixation, Spinal Interlaminal devices

All Appliance, Fixation, Spinal Interlaminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page