The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 08022057013132), received between Oct 19, 2016 and Sep 6, 2024 — classified as 1 death, 12 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information10
  • Adverse Event Without Identified Device or Use Problem7
  • Patient-Device Incompatibility1

Patient problems

  • No Information9
  • No Known Impact Or Consequence To Patient4
  • Mitral Regurgitation2
  • Heart Failure/Congestive Heart Failure1
  • Insufficient Information1
  • Myocardial Infarction1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjurySep 6, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MANUFACTURER WAS INFORMED OF THE FOLLOWING EVENT THROUGH COMPETITOR'S PATIENT TRACKING. BASED ON THE INFORMATION RECEIVED, THE MEMO 3D RING ICV0970 WAS IMPLANTED IN A PATIENT ON (B)(6) 2023. REPORTEDLY, THE VALVE WAS EXPLANTED ON (B)(6) 2023 AND REPLACED WITH PHYSIO II 30MM. MANUFACTURER WAS INFORMED THAT NO FURTHER INFORMATION WILL BE PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 20161219 ↗
  • DeathApr 21, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    THE MANUFACTURER WAS INFORMED THROUGH MANTRA STUDY THAT A PATIENT IMPLANTED WITH A MEMO 3D SIZE 32 MITRAL ANNULOPLASTY RING PASSED AWAY ON (B)(6) 2022. THE EVENT WAS A SUDDEN DEATH OCCURRING DURING THE NIGHT WHILE THE PATIENT WAS SLEEPING. THE RING HAD BEEN IMPLANTED ON (B)(6) 2021 THROUGH MEDIAN STERNOTOMY AND CONCOMITANT CABG HAD BEEN PERFORMED IN THE SAME SURGERY PRIOR TO RING IMPLANTATION. THE PATIENT WAS IN PARTICULARLY COMPLEX HEALTH CONDITIONS AND SUFFERED FROM BRONCHOPATHY WITH SLEEP APNEA. HE HAD A PREVIOUS HISTORY OF CORONARY ARTERY DISEASE, MYOCARDIAL INFARCTION (<= 90 DAYS AT THE TIME OF SURGERY), ISCHEMIC HEART FAILURE, DIABETES MELLITUS TYPE II, DYSLIPIDEMIA. HE ALSO HAD A TOBACCO HISTORY AND WAS NYHA CLASS III.

    Patient problems: Heart Failure/Congestive Heart Failure; Myocardial Infarction

    MAUDE report 16782945 ↗
  • InjuryNov 4, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    THE MANUFACTURER WAS INFORMED ON THIS EVENT THROUGH THE DEVICE TRACKING DEPARTMENT. BASED ON THE INFORMATION REPORTED ON THE PATIENT IMPLANT FORM, A MEMO 3D SMD32 WAS IMPLANTED AND EXPLANTED ON (B)(6) 2020. NO FURTHER INFORMATION IS PRESENTLY AVAILABLE. NO INDICATION OF A DEVICE MALFUNCTION NOR SERIOUS INJURY WAS RECEIVED FROM THE SITE REGARDING THIS EVENT.

    Patient problems: No Information

    MAUDE report 10787766 ↗
  • InjuryNov 4, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    THE MANUFACTURER WAS INFORMED ON THIS EVENT THROUGH THE DEVICE TRACKING DEPARTMENT. BASED ON THE INFORMATION REPORTED ON THE PATIENT IMPLANT FORM, A MEMO 3D SMD32 WAS IMPLANTED AND EXPLANTED ON (B)(6) 2020. NO FURTHER INFORMATION IS PRESENTLY AVAILABLE. NO INDICATION OF A DEVICE MALFUNCTION NOR SERIOUS INJURY WAS RECEIVED FROM THE SITE REGARDING THIS EVENT.

    Patient problems: No Information

    MAUDE report 10787723 ↗
  • InjuryApr 9, 2020Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    THE MANUFACTURER WAS INFORMED ON THIS EVENT THROUGH THE DEVICE TRACKING DEPARTMENT. ON (B)(6) 2020, A MEMO 3D #32 WAS IMPLANTED AND EXPLANTED INTRAOPERATIVELY. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. NO COMPLAINT RECEIVED FROM THE SITE REGARDING THIS EVENT.

    Patient problems: No Information; No Known Impact Or Consequence To Patient

    MAUDE report 9944205 ↗
  • InjuryApr 9, 2020Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    THE MANUFACTURER WAS INFORMED ON THIS EVENT THROUGH THE DEVICE TRACKING DEPARTMENT. ON 07 FEB 2020, A MEMO 3D #32 WAS IMPLANTED AND EXPLANTED INTRAOPERATIVELY. NO COMPLAINT RECEIVED FROM THE SITE REGARDING THIS EVENT. PER ADDITIONAL INFORMATION RECEIVED, THE RING WAS EXPLANTED INTRAOPERATIVELY DUE TO A FAILED REPAIR, ULTIMATELY CONVERTED TO A FULL VALVE REPLACEMENT. THE RING WAS DISCARDED. THE PRE-OPERATIVE DIAGNOSIS WAS OF SEVERE, SYMPTOMATIC MITRAL INSUFFICIENCY. THE OP NOTE REPORTS THAT THERE WAS A RUPTURED P2 CHORD, AND A REPAIR OF THE VALVE WAS INITIALLY ATTEMPTED TO AMELIORATE THE INSUFFICIENCY. AFTER THE IMPLANT OF THE 32 MEMO 3D RING, "SALINE TESTING AT THIS POINT SHOWED THAT THERE WAS AN ASYMMETRIC BILLOWING OF A1 AND A2 THAT HAD NOT BEEN APPRECIATED AND THERE WAS A SIGNIFICANT LEAK. THE RING WAS REMOVED AND THE ANTERIOR LEAFLET EXCISED. A 31 MM MEDTRONIC MOSAIC PORCINE TISSUE VALVE WAS IMPLANTED". THE PATIENT WAS NOT ADVERSELY IMPACTED BY THE EVENT BASED ON THE REPORT RECEIVED.

    Patient problems: No Information; No Known Impact Or Consequence To Patient

    MAUDE report 9944196 ↗
  • InjuryApr 1, 2020Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    THE MANUFACTURER WAS INFORMED ON THIS EVENT THROUGH THE DEVICE TRACKING TEAM. PER THE INFORMATION REPORTED IN THE PATIENT IMPLANT FORM, ON (B)(6) 2020 A PATIENT RECEIVED A MEMO 3D #32 IN MITRAL POSITION. THE DEVICE WAS EXPLANTED INTRAOPERATIVELY. NO OTHER INFORMATION IS PRESENTLY AVAILABLE. NO COMPLAINT HAS BEEN RECEIVED FROM THE SITE REGARDING THIS EVENT.

    Patient problems: No Information; No Known Impact Or Consequence To Patient

    MAUDE report 9912435 ↗
  • InjuryApr 1, 2020Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    THE MANUFACTURER WAS INFORMED ON THIS EVENT THROUGH THE DEVICE TRACKING TEAM. PER THE INFORMATION REPORTED IN THE PATIENT IMPLANT FORM, ON (B)(6) 2020 A PATIENT RECEIVED A MEMO 3D #32 IN MITRAL POSITION. THE DEVICE WAS EXPLANTED INTRAOPERATIVELY. NO OTHER INFORMATION IS PRESENTLY AVAILABLE. NO COMPLAINT HAS BEEN RECEIVED FROM THE SITE REGARDING THIS EVENT.

    Patient problems: No Information; No Known Impact Or Consequence To Patient

    MAUDE report 9912377 ↗
  • InjuryAug 1, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MEMO 3D, SIZE (B)(6) WAS IMPLANTED AND EXPLANTED ON (B)(6) 2018. THE SURGEON IMPLANTED THE DEVICE, REALIZED IT WAS TOO SMALL, AND REPLACED IT WITH A SIZE (B)(6). IT WAS REPORTED THAT THERE WAS NOTHING WRONG WITH THE DEVICE, OR ANY HARM TO THE PATIENT.

    Patient problems: No Information

    MAUDE report 7740052 ↗
  • InjuryMay 22, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2008 A MEMO 3D SIZE 32 MECHANICAL MITRAL REPAIR RING WAS IMPLANTED. ON (B)(6) 2018 A CARBOMEDICS M7-027 STANDARD MITRAL VALVE WAS IMPLANTED. NO FURTHER EVENT INFORMATION WAS RECEIVED.

    Patient problems: No Information

    MAUDE report 7535329 ↗
  • InjuryNov 28, 2017Patient-Device Incompatibility; Adverse Event Without Identified Device or Use Problem; Patient-Device Incompatibility; Adverse Event Without Identified Device or Use Problem

    A MEMO3D RECHORD WAS IMPLANTED AND EXPLANTED ON (B)(6) 2017. THE SURGEON INDICATED THAT NOTHING WAS WRONG WITH THE RING. RATHER, HE CHOSE TO REPLACE THE MITRAL VALVE AFTER FIRST ATTEMPTING TO REPAIR THE VALVE WITH THE MEMO RING.

    Patient problems: No Information

    MAUDE report 7065644 ↗
  • InjuryApr 27, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED FROM PATIENT REGISTRATION FORM THAT THE PATIENT HAD A MITRAL REPAIR SURGERY ON (B)(6) 2017 AND ON (B)(6) 2017 UNDERWENT A SECOND MITRAL REPAIR SURGERY WITH A SMALLER SIZE RING. IT WAS STATED THAT THE INCORRECT SIZE REPAIR (TOO LARGE) WAS IMPLANTED WHICH LED TO MODERATE TO SEVERE MITRAL REGURGITATION. THE RING WAS REPLACED WITH A SMALLER RING WHICH RESOLVED THE ISSUE. THE RING WAS DISCARDED.

    Patient problems: Mitral Regurgitation

    MAUDE report 6528147 ↗
  • InjuryOct 19, 2016Adverse Event Without Identified Device or Use Problem

    PT (B)(6) RECEIVED A PRF FOR A MITRAL RING SMD32 SN # (B)(4) WHICH WAS REMOVED ON (B)(6) 2016 AT (B)(6) MEDICAL CENTER. PATIENT (B)(6) IS MALE, DOB (B)(6) 1947. UPDATE FROM SALES 29 SEPT 2016: DR (B)(6) WAS OPERATING SURGEON. HE PLACED THE 32 RING INTO THE PATIENT. WHEN COMING OFF BYPASS AND POST TEE PERFORMED, SEVERE MR WAS NOTED. WENT BACK ONTO BYPASS, REMOVED THE RING AND REPLACED WITH AN EDWARD'S TISSUE VALVE. PATIENT HAD AN EXTENDED POST OP RECOVERY INCLUDING LONG INTUBATION TIME, POST OP ACUTE RENAL FAILURE, ELEVATED LIVER FUNCTIONS AND A STROKE. PATIENT WAS DISCHARGED AFTER 14 DAYS TO A REHAB FACILITY.

    Patient problems: Mitral Regurgitation

    MAUDE report 6041602 ↗

Other RING, ANNULOPLASTY devices

All RING, ANNULOPLASTY devices →