XTRA PROCEDURE SET BX/225Recalls & reported problems
Sorin Group Italia Srl · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08033178110808), received between Jul 6, 2016 and Jul 4, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Operates Differently Than Expected1
- Disconnection1
- Fracture1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 4, 2024Fracture; Fracture
SORIN GROUP ITALY RECEIVED A REPORT THAT THE UPPER PORTION OF AN XTRA BOWL COMING SEPARATED FROM THE BOTTOM PORTION THAT SITS IN THE CENTRIFUGE WELL AND AS A CONSEQUENCE THERE WAS A SIGNIFICANT AMOUNT OF FLUID LOSS. BLOOD LEAKED OUT THE BOWL AND CENTRIFUGE AND THE USER WAS CONTAMINATED. THE ISSUE OCCURRED DURING PROCEDURE. THE BOWL PLATE WAS LEFT IN THE CENTRIFUGE WELL AND THE CUSTOMER HAD TO USE A SCREWDRIVER TO PRY THE DISPOSABLE BASE FROM THE CENTRIFUGE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19674951 ↗MalfunctionApr 5, 2017Disconnection; Disconnection
SORIN GROUP (B)(6) RECEIVED A REPORT THAT THE BLUE TUBING FROM THE XTRA PROCEDURE SET DISCONNECTED FROM THE RESERVOIR DURING A CASE AND BLOOD SPILLED ON FLOOR AND DOCTOR. THE TUBING WAS RECONNECTED AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER ISSUE. THERE WAS NO PATIENT INJURY, AND NO INJURY WAS REPORTED FOR THE DOCTOR.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6459594 ↗Jul 6, 2016Device Operates Differently Than Expected; Device Operates Differently Than Expected
SORIN GROUP (B)(4) RECEIVED A USER MEDWATCH REPORT (MW5062325) ON JUNE 13, 2016 STATING THAT THE XTRA BOWL FOR THE XTRA AUTOTRANSFUSION SYSTEM DID NOT EMPTY PROPERLY DURING A PROCEDURE. THE BOWL WOULD PARTIALLY EMPTY AND THEN STOP. THE BOWL WAS CHANGED TO ANOTHER FROM THE SAME LOT AND THE SAME ISSUE OCCURRED. A THIRD DIFFERENTLY SIZED BOWL FROM A DIFFERENT LOT WAS SET UP AND WORKED PROPERLY. THE CUSTOMER REPORTED THAT 2000ML OF SALVAGED BLOOD WAS DISCARDED DUE TO THE ISSUE. THERE WAS NO REPORT OF PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5774159 ↗
Other APPARATUS, AUTOTRANSFUSION devices
- OASIS 3650-100Atrium Medical Corporation · APPARATUS, AUTOTRANSFUSION2 recalls22 reports
- EXPRESS 4050-100NAtrium Medical Corporation · APPARATUS, AUTOTRANSFUSION2 recalls2 reports
- NA ATL2001Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 death79 reports
- autoLog IQ™ ATLGIQMedtronic, Inc. · APPARATUS, AUTOTRANSFUSION58 reports
- NA BTC93Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 recall2 reports
- OCEAN 2050-000Atrium Medical Corporation · Apparatus, Autotransfusion1 recall2 reports
- NA BTC98Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 recall0 reports
- HAEMONETICS CELL SAVER ELITE SYSTEM CSE-P-125Haemonetics Corporation · Apparatus, Autotransfusion47 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.