SMR Shoulder Liner (REF 1360.50.810)Recalls & reported problems
Limacorporate SPA · Reverse liners 36mm (UHMWPE) Standard · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08033390080613), received between Jul 18, 2024 and Jan 27, 2026 — classified as 4 injuries.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K110598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Patient Device Interaction Problem2
- Device Dislodged or Dislocated1
Patient problems
- Medical device site infection2
- Insufficient Information1
- Joint Dislocation1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJan 27, 2026Patient Device Interaction Problem
SHOULDER REVISION SURGERY PERFORMED ON (B)(6) 2025, DUE TO GLENOID LOOSENING. THE PATIENT IS A FEMALE, DATE OF BIRTH (B)(6) 1956. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: Insufficient Information
MAUDE report 24188853 ↗InjuryJun 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION SURGERY PERFORMED ON (B)(6) 2025, DUE TO INFECTION. THERE WAS AN INFECTION FROM THE SKIN INCISION, THE SURGEON WANTED TO DO A POLY SWAP. THE FOLLOWING LINER WAS THE ONLY COMPONENT REMOVED AND REPLACED BY A SIMILAR DEVICE: SMR REVERSE LINER STANDARD (PART CODE: 1360.50.810, LOT NUMBER: 22AT1GH, STERILIZATION 2200243). THE OTHER COMPONENTS WERE LEFT IN SITU: SMR METAL-BACK GLENOID SMALL R (PART CODE: 1375.21.005, LOT NUMBER: 2433154, STERILIZATION 2500010) . SMR SMALL-R CONNECTOR +2 (PART CODE: 1374.15.312, LOT NUMBER: 2500708, STERILIZATION 2500041). SMR GLENOSPHERE DIA. 36MM (PART CODE: 1374.09.111, LOT NUMBER: 2436656, STERILIZATION 2500015). FINNED REVERSE HUM. BODY 140° (PART CODE: 1352.15.051, LOT NUMBER: 2416428, STERILIZATION 2400165). SMR CEMENTLESS FINNED STEM (PART CODE: 1304.15.140, LOT NUMBER: 2312939, STERILIZATION 2300232). PREVIOUS SURGERY TOOK PLACE ON (B)(6) 2025. THE PATIENT IS A FEMALE, DATE OF BIRTH ON (B)(6) 1969. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: Medical device site infection
MAUDE report 22288364 ↗InjuryFeb 12, 2025Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem
REVISION SURGERY DUE TO SEPTIC LOOSENING PERFORMED ON (B)(6) 2025. PATIENT COMPLAINED OF PAIN AND X-RAYS WERE TAKEN. IT WAS DETERMINED THAT THE STEM APPEARED LOOSE A DUE TO A POSSIBLE INFECTION. A 1ST STAGE INFECTION SURGERY WAS SCHEDULED. THE FOLLOWING COMPONENTS WERE REMOVED: - SMR CEMENTLESS FINNED STEM (PART NUMBER 1304.15.180, LOT NUMBER 1808774, STERILIZATION (B)(4)) - SMR REVERSE HUMERAL BODY (PART NUMBER 1352.15.010, LOT NUMBER 1809657, STERILIZATION (B)(4)) - SMR REVERSE LINER STANDARD (PART NUMBER 1360.50.810, LOT NUMBER 18AT0GK, STERILIZATION (B)(4)) - SMR GLENOSPHERE Ø36 MM (PART NUMBER 1376.09.030, LOT NUMBER 1809534, STERILIZATION (B)(4)). THE LIMA LONG REVISION STEM WAS TRIALED ALONG WITH MULTIPLE EXTENSIONS AND LINERS. IT WAS DETERMINED THAT THIS COMBINATION WOULD NOT REDUCE PROPERLY. A TORNIER REVISION STEM WAS TRIALED ALONG WITH VARIOUS TRAYS AND LINERS. IT WAS DETERMINED THAT THIS WAS A GOOD FIT AND THE DEFINITIVE IMPLANTS WERE IMPLANTED ALONG WITH A LIMA 36 ECCENTRIC GLENOSPHERE. PREVIOUS SURGERY TOOK PLACE ON (B)(6) 2018. THE PATIENT IS A FEMALE, DATE OF BIRTH (B)(6) 1964. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: Medical device site infection
MAUDE report 21362127 ↗InjuryJul 18, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
SHOULDER REVISION SURGERY PERFORMED ON (B)(6) 2024, DUE TO DISLOCATION. DURING THE REVISION SURGERY, THE FOLLOWING COMPONENTS WERE REMOVED: SMR REVERSE LINER STANDARD (PART CODE 1360.50.810, LOT NUMBER 23AT0V8, STERILIZATION NUMBER (B)(4). SMR GLENOSPHERE Ø 36MM (PART CODE 1374.09.111, LOT NUMBER 2325319, STERILIZATION NUMBER (B)(4). SMR CONNECTOR SMALL R (PART CODE 1374.15.305, LOT NUMBER 2327395, STERILIZATION NUMBER (B)(4). THE COMPONENTS ABOVE MENTIONED WERE REPLACED BY THE FOLLOWING NEW COMPONENTS: SMR REVERSE LINER RETENTIVE (PART CODE 1361.50.815, LOT NUMBER 21AT0F5, STERILIZATION NUMBER (B)(4). SMR GLENOSPHERE Ø 36MM (PART CODE 1374.09.111, LOT NUMBER 2218959, STERILIZATION NUMBER (B)(4). SMR CONNECTOR+SCREW SMALL R (PART CODE 1374.15.305, LOT NUMBER 2411586, STERILIZATION NUMBER (B)(4). THE PREVIOUS SURGERY TOOK PLACE ON (B)(6)2024. DURING THIS SURGERY, THE SMR REVERSE HUMERAL BODY SHORT (PART CODE 1352.15.005, LOT NUMBER 2202133, STERILIZATION NUMBER (B)(4). WAS ALSO IMPLANTED (AND NOT REMOVED DURING THE REVISION SURGERY). THE PATIENT IS A FEMALE, DATE OF BIRTH (B)(6)1956. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: Joint Dislocation
MAUDE report 19772700 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K110598. One submission often covers many models, so check the product and lot details in each record.
- Z-1711-2016 · Apr 8, 2016 · TerminatedSMR glenosphere impactor/extractor; Used to implant the glenosphere in the SMR Reverse Shoulder replacement.
Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page