SMR Shoulder 1360.50.820Recalls & reported problems
Limacorporate SPA · SMR Shoulder - Reverse liners 36mm (UHMWPE) +6mm · Class 2 · Implant
All models of SMR Shoulder (67)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08033390080637), received between May 30, 2025 and Feb 17, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K110598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Patient Device Interaction Problem2
- Device Damaged by Another Device1
Patient problems
- Joint Laxity2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryFeb 17, 2026Device Damaged by Another Device
REVISION SURGERY WAS PERFORMED ON (B)(6) 2025 DUE TO GLENOID COMPONENTS FAILURE. PREVIOUS SURGERY WAS PERFORMED ON (B)(6) 2023 WHEN THE ORIGINAL ASSEMBLY, CONSISTING OF SMR HUMERAL COMPONENTS COUPLED WITH ZIMMERBIOMET VRS CUSTOM GLENOID, WAS IMPLANTED. DURING REVISION THE SURGEON REMOVED THE FAILED GLENOID COMPONENTS, AND REPLACED THE SMR REVERSE HUMERAL BODY (PN 1352.15.010, LOT 2222575, STERILIZATION, (B)(4) AND SMR REVERSE LINER +6 MM (PN 1360.50.820, LOT 19AT06N, STERILIZATION (B)(4)) WITH ADAPTOR TAPER NEUTRAL (PN 1330.15.270), HUMERAL BODY TRAUMA LONG (PN 1350.15.020) AND CTA HEAD DIA54MM (PN 1323.09.540) TO CONVERT THE REVERSE PROSTHESIS TO AN ANATOMIC CONFIGURATION. SURGERY WAS COMPLETED AS INTENDED. PATIENT IS MALE, DATE OF BIRTH (B)(6) 1954. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Joint Laxity
MAUDE report 24362075 ↗InjuryJul 18, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
REVISION SURGERY WAS PERFORMED ON (B)(6) 2025, DUE TO LOOSENING OF THE GLENOID COMPONENT. A COMPLAINT HAS BEEN FILED REGARDING THE SMR REVERSE LINER (HUMERAL COMPONENT), AS INFORMATION ABOUT THE SUSPECTED DEVICES RELATED TO THE GLENOID PART WAS NOT PROVIDED. THE SMR REVERSE LINER INVOLVED IN THIS EVENT IS THE FOLLOWING: SMR REVERSE LINER +6 MM (PART CODE 1360.50.820, LOT NUMBER 23AT0VA, STERILIZATION UNKNOWN) THE DATE OF THE INITIAL SURGERY IS UNKNOWN. THE PATIENT IS A FEMALE. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: Insufficient Information
MAUDE report 22540944 ↗InjuryMay 30, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
REVISION SURGERY PERFORMED ON (B)(6) 2025, DUE TO LOOSENING. THE GLENOID IMPLANT AND BASEPLATE WERE PLACED TOO SUPERIOR IN THE GLENOID LEADING TO LOOSENING. AT THE TIME OF THE REVISION, THE FOLLOWING COMPONENTS WERE IMPLANTED ON THE GLENOID SIDE: SMR GLENOSPHERE Ø 36MM (PART CODE 1374.09.111, LOT NUMBER 2213083, STERILIZATION (B)(4), SMR REVERSE LINER +6 MM (PART CODE 1360.50.820, LOT NUMBER 20AT5KQ, STERILIZATION (B)(4), SMR SMALL-R CONNECTOR +4 (PART CODE 1374.15.314, LOT NUMBER 2212074, STERILIZATION (B)(4), SMR GLENOID PEG TT SMALL-R #L (PART CODE 1375.14.653, LOT NUMBER 2007768, STERILIZATION (B)(4), TT AUGM.360 BASEPLATE #S-R 7° (PART CODE 1375.15.507, LOT NUMBER 2213740, STERILIZATION (B)(4). DURING THE REVISION SURGERY, THE SMR REVERSE LINER PREVIOUSLY IMPLANTED WAS REPLACED BY THE FOLLOWING NEW COMPONENT: SMR RETENTIVE LINER STD (PART CODE 1361.50.810, LOT NUMBER 22AT144, STERILIZATION (B)(4). MOREOVER, IN REVISION THE SURGEON IMPLANTED A SMR REVERSE HUMERAL BODY 140° (PART CODE 1352.15.011), AND AN EXTENSION FOR HUMERAL REVERSE BODY (PART CODE 1352.15.001) AS WELL. PREVIOUS SURGERY WAS PERFORMED IN (B)(6) 2023. THE PATIENT IS A MALE, DATE OF BIRTH (B)(6) 1963. THE EVENT H
Patient problems: Joint Laxity
MAUDE report 22113297 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K110598. One submission often covers many models, so check the product and lot details in each record.
- Z-1711-2016 · Apr 8, 2016 · TerminatedSMR glenosphere impactor/extractor; Used to implant the glenosphere in the SMR Reverse Shoulder replacement.
Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page