SMR Shoulder Cemented Glenoid (REF 1379.50.020)Recalls & reported problems
Limacorporate SPA · SMR 3-Pegs Small · Class 2 · Implant
All models of SMR Shoulder (67)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08033390096362), received between Jan 2, 2026 and Jan 9, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Patient Device Interaction Problem1
Patient problems
- Joint Laxity1
- Loose body in joint1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 9, 2026Detachment of Device or Device Component
REVISION SURGERY WAS PERFORMED ON (B)(6) 2025 DUE TO LOOSENING/DISASSEMBLY OF GLENOID COMPONENTS. PATIENT UNDERWENT PRIMARY SURGERY IN 2014 AND RECEIVED AN ANATOMIC SHOULDER REPLACEMENT UTILIZING SMR HUMERAL COMPONENTS INCLUDING: - SMR HUMERAL HEAD DIA 42MM (PART NUMBER 1322.09.420), - SMR ECC. ADAPTOR TAPER STANDARD (PART NUMBER 1330.15.272), - SMR TRAUMA HUMERAL BODY #MEDIUM (PART NUMBER 1350.15.010), AND SMALL 3-PEGGED CEMENTED GLENOID (PART NUMBER 1379.50.020, LOT UNKNOWN). DUE TO GLENOID IMPLANT LOOSENING OVER TIME, OSTEOLYSIS CAUSED SIGNIFICANT AMOUNT OF GLENOID BONE LOSS. ABOVE-MENTIONED COMPONENTS WERE REMOVED AND DEMINERALIZED BONE MATRICE AND CANCELLOUS BONE CHIPS WERE USED TO GRAFT IN THE GLENOID. NEW HUMERAL BODY, ADAPTER, AND HUMERAL HEAD WERE IMPLANTED AND PATIENT CURRENTLY HAS A HEMIARTHROPLASTY, UNTIL A SECOND PROCEDURE CAN BE UNDERGONE IN THE FUTURE TO IMPLANT A BASEPLATE. PATIENT IS A FEMALE, DATE OF BIRTH (B)(6) 1948. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Loose body in joint
MAUDE report 24018190 ↗InjuryJan 2, 2026Patient Device Interaction Problem
REVISION SURGERY WAS PERFORMED ON (B)(6) 2025 DUE TO LOOSENING OF THE CEMENTED GLENOID. PATIENT HAD A PRIOR ANATOMIC SHOULDER WITH CEMENTED 3-PEGGED POLY (PART NUMBER 1379.50.020, LOT UNKNOWN). AFTER LOOSENING OF THE POLY GLENOID WAS DETECTED IN A CLINIC VISIT, A TWO-STAGED REVISION WAS PLANNED. DURING REVISION SURGERY, 3-PEGGED CEMENTED POLY WAS REMOVED AND THE VOID IN GLENOID WAS FILLED WITH BONE GRAFT. A NEW HUMERAL HEAD (PART NUMBER 1324.09.461) AND ADAPTER (PART NUMBER 1330.15.270) OF THE SAME SIZE WERE IMPLANTED FOR RENEWAL REASONS, BUT NO DEFECT WERE REPORTED ON ORIGINAL HUMERAL COMPONENTS. HUMERAL BODY WAS INDEED RETAINED. PREVIOUS SURGERY DATE IS UNKNOWN. PATIENT IS FEMALE, DATE OF BIRTH (B)(6) 1962. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Joint Laxity
MAUDE report 23943074 ↗
Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.