SMR Shoulder Liner (REF 1365.50.820)Recalls & reported problems
Limacorporate SPA · Reverse Liner 40mm +6mm · Class 2 · Implant
All models of SMR Shoulder (65)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08033390102940), received between Aug 6, 2024 and Jul 2, 2026 — classified as 7 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated3
- Detachment of Device or Device Component2
- Adverse Event Without Identified Device or Use Problem1
- Insufficient Device Problem Information1
- Unstable1
Patient problems
- Joint Dislocation3
- Insufficient Information2
- Joint Laxity2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryJul 2, 2026Insufficient Device Problem Information
SHOULDER REVISION SURGERY HAS BEEN PERFORMED ON (B)(6), 2026, DUE TO FAILURE OF GLENOID COMPONENTS. PREVIOUS SURGERY OCCURRED ON (B)(6), 2024, WHEN A REVERSE CONFIGURATION CONSISTING OF BOTH LIMACORPORATE AND ANOTHER MANUFACTURER COMPONENTS WERE IMPLANTED. DURING THE REVISION SURGERY HEREBY REPORTED, THE PRE-EXISTING REVERSE SHOULDER CONFIGURATION WAS CONVERTED TO A HEMIARTHROPLASTY CONFIGURATION, DUE TO FAILURE OF THE GLENOID COMPONENTS PRODUCED BY THE OTHER MANUFACTURER. THE PREVIOUSLY IMPLANTED COMPONENTS, INCLUDING THE SMR REVERSE LINER +6 MM D.40 MM (PN 1365.50.820, LOT. 23AT2PJ, STER. 2400073), WERE EXPLANTED AND FOLLOWING COMPONENTS WERE SUBSEQUENTLY IMPLANTED: · SMR FINNED STEM, DIAMETER 19 MM, LENGTH 80 MM (PN 1304.15.190); · SMR 140° REVERSE HUMERAL BODY (PN 1352.15.011); · SMR ADAPTER 36 MM FOR REVERSE BODY (PN 1352.15.200); · CTA HEAD, DIAMETER 54 MM (PN 1323.09.540). THE PROCEDURE WAS COMPLETED AS INTENDED. THE PATIENT IS FEMALE, DATE OF BIRTH (B)(6) 1963. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Insufficient Information
MAUDE report 25687434 ↗InjuryMay 18, 2026Device Dislodged or Dislocated
REVISION SURGERY WAS PERFORMED ON (B)(6) 2026 DUE TO SHOULDER DISLOCATION. PREVIOUS SURGERY OCCURRED ON (B)(6) 2023 FOR A REVERSE TOTAL SHOULDER ARTHROPLASTY, WITH FOLLOWING COMPONENTS: - SMR TT BASEPLATE SMALL-R (PN 1375.15.605), - SMR GLENOID PEG TT SMALL-R #MEDIUM (PN 1375.14.652), - CONNECTOR LAR +2MM SMALL-R (PN 1374.15.312), - GLENOSPHERE DIA40MM (PN 1374.09.121, LOT 2213444, STERILIZATION (B)(4)), - FINNED STEM SHORT DIA25MM (PN 1304.15.025), - 140° REVERSE HUMERAL BODY (PN 1352.15.011). - REVERSE LINER +6MM DIA40MM (PN 1365.50.820, LOT 22AT0WC, STERILIZATION (B)(4)). AROUND 3 YEARS AFTER PREVIOUS SURGERY, PATIENT DISLOCATED. PRE-EXISTING CONNECTOR, GLENOSPHERE, REVERSE HUMERAL BODY AND LINER WERE REMOVED AND REPLACED WITH A MORE LATERALIZED AND THICK ASSEMBLY CONSISTING OF: - CONNECTOR LAT +4MM (PN 1374.15.314), - GLENOSPHERE ECCENTRICAL DIA 40MM (PN 1376.09.041), - 140° REVERSE HUMERAL BODY (PN 1352.15.011), - EXTENSION FOR HUMERAL REVERSE BODY (PN 1352.15.001), - REVERSE LINER +6MM DIA40MM (PN 1365.50.820). SURGERY WAS COMPLETED AS INTENDED. PATIENT IS MALE, DATE OF BIRTH (B)(6) 1953. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Joint Dislocation
MAUDE report 25200752 ↗InjuryMay 11, 2026Detachment of Device or Device Component; Device Dislodged or Dislocated
REVISION SURGERY HAS BEEN PERFORMED ON (B)(6) 2026, DUE TO LOOSENING OF JOINT AND DISLOCATION. DURING THE REVISION SURGERY PRE-EXISTING SMR REVERSE LINER +6MM DIA.40MM (PN 1365.50.820, LOT. 24AT3VE) HAS BEEN EXPLANTED AND REPLACED WITH SMR REVERSE LINER RETENTIVE +6MM DIA. 40MM (PN 1365.50.821). SURGERY HAS BEEN COMPLETED AS INTENDED. PREVIOUS SURGERY OCCURRED ON (B)(6) 2026. PATIENT DATE OF BIRTH IS (B)(6) 1953. EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Joint Dislocation
MAUDE report 25134380 ↗InjuryMar 6, 2026Detachment of Device or Device Component
REVISION SURGERY WAS PERFORMED ON (B)(6) 2026 DUE TO DISASSEMBLY OF REVERSE LINER. PREVIOUS SURGERY OCCURRED ON DECEMBER 30TH 2025 WHEN THE FOLLOWING SMR REVERSE CONFIGURATION WAS IMPLANTED: - SMR CONNECTOR SMALL STD (PART NUMBER 1374.15.310, LOT 2500908, STERILIZATION (B)(4)), - SMR GLENOSPHERE Ø 40MM (PART NUMBER 1374.09.121, LOT 2501638, STERILIZATION (B)(4)), - SMR REVERSE HUMERAL BODY SHORT (PART NUMBER 1352.15.005, LOT 2507908, STERILIZATION (B)(4)), - SMR REVERSE LINER +6MM D.40MM (PART NUMBER 1365.50.820, LOT 24AT3VE, STERILIZATION (B)(4)). DURING REVISION THE ABOVE-MENTIONED LINER WAS REMOVED AND REPLACED WITH A NEW THINNER REVERSE LINER +3MM DIA40MM (PN 1365.50.815). SURGERY WAS COMPLETED AS INTENDED. THE PATIENT IS MALE, DATE OF BIRTH (B)(6) 1960. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Insufficient Information
MAUDE report 24531027 ↗InjuryFeb 12, 2026Device Dislodged or Dislocated
REVISION SURGERY WAS PERFORMED ON (B)(6) 2025 BECAUSE PATIENT DISLOCATED. PREVIOUS SURGERY IS UNKNOWN, AS WELL AS COMPLETE INFORMATION OF ORIGINAL ASSEMBLY. DURING REVISION THE SURGEON REMOVED THE PRE-EXISTING SMR REVERSE LINER +6MM D.40MM (PN 1365.50.820) AND IMPLANTED AN EXTENSION FOR HUMERAL REVERSE BODY (PN 1352.15.001) AND A REVERSE LINER +6MM DIA40MM (PN 1365.54.820). SURGERY WAS COMPLETED AS INTENDED TO RESTORE SHOULDER FUNCTIONALITY. PATIENT IS MALE, DATE OF BIRTH (B)(6) 1964. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Joint Dislocation
MAUDE report 24322377 ↗InjuryJan 29, 2026Adverse Event Without Identified Device or Use Problem
REVISION SURGERY PERFORMED ON (B)(6) 2025 DUE TO GLENOSPHERE FAILURE. REPORTEDLY, ORIGINAL IMPLANT WAS A REVERSE CONFIGURATION COMBINING ZIMMERBIOMET VRS ON THE GLENOID SIDE AND SMR COMPONENTS ON THE HUMERAL SIDE. PREVIOUS SURGERY IS UNKNOWN, AS WELL AS COMPLETE PRODUCT INFORMATION OR ORIGINAL COMPONENTS. DURING REVISION THE SURGEON REMOVED THE PRE-EXISTING REVERSE LINER +6MM DIA40MM (PART NUMBER 1365.50.820) AND IMPLANTED AN EXTENSION FOR REVERSE HUMERAL BODY (PN 1352.15.001) AND A REVERSE LINER +3MM (PN 1361.50.815). SURGERY WAS COMPLETED AS INTENDED. PATIENT IS MALE, DATE OF BIRTH (B)(6)1956. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Joint Laxity
MAUDE report 24210807 ↗InjuryAug 6, 2024Unstable; Unstable
SHOULDER REVISION SURGERY PERFORMED ON (B)(6) 2024 DUE TO INSTABILITY. THE PRIMARY SURGERY TOOK PLACE ON (B)(6) 2024, WHEN THE FOLLOWING COMPONENTS WERE IMPLANTED: SMR TT AUGMENTED 360 BASEPLATE (PART CODE 1375.15.507, LOT NUMBER 2311893, STERILIZATION (B)(4). SMR GLENOSPHERE ECCENTRICAL DIA. 40 MM (PART CODE 1376.09.041, LOT NUMBER 2226888, STERILIZATION (B)(4). SMR CONNECTOR + SCREW SMALL-R (PART CODE 1374.15.305, LOT NUMBER 2329787, STERILIZATION (B)(4). SMR GLENOID PEG TT SMALL-R (PART CODE 1375.14.652, LOT NUMBER 2313251, STERILIZATION (B)(4). SMR REVERSE LINER +6MM DIA. 40 MM (PART CODE 1365.50.820, LOT NUMBER 19AT2BA, STERILIZATION (B)(4). SMR HUMERAL BODY FINNED REVERSE (PART CODE 1352.15.050, LOT NUMBER 2403061, STERILIZATION (B)(4) SMR FINNED STEM (PART CODE 1304.15.180, LOT NUMBER 2303828, STERILIZATION (B)(4) FOUR CORTICAL BONE SCREWS. DURING THE REVISION SURGERY, THE GLENOSPHERE, THE CONNECTOR + SCREW, AND THE REVERSE LINER PREVIOUSLY IMPLANTED WERE REMOVED AND REPLACED BY THE FOLLOWING NEW COMPONENTS: SMR GLENOSPHERE DIA. 40 MM (PART CODE 1374.09.121, LOT NUMBER 2331854, STERILIZATION (B)(4). SMR CONNECTOR LAT. + 2MM + SCREW (PART CODE 1374.15.312, LOT NUMBER 2405388,
Patient problems: Joint Laxity
MAUDE report 19915913 ↗
FDA clearance and classification
Cleared through 510(k) K142139 on Nov 7, 2014 (decision: Substantially Equivalent). Submitted by Lima Corporate S.P.A. as “Smr 40mm Glenosphere”. The FDA review took 94 days. FDA record ↗
The FDA regulates product code KWS (prosthesis, shoulder, semi-constrained, metal/polymer cemented) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the SMR Shoulder Liner
Has the SMR Shoulder Liner been recalled?
No FDA recall record in our data names this exact device identifier (DI 08033390102940), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SMR Shoulder Liner?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Aug 6, 2024 and Jul 2, 2026. In those reports, the most common specific device problems are “Device Dislodged or Dislocated” (3), “Detachment of Device or Device Component” (2) and “Unstable” (1); the most common patient problems are “Joint Dislocation” (3) and “Joint Laxity” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SMR Shoulder Liner?
None of the reports are classified as deaths, 7 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SMR Shoulder Liner FDA approved?
It was cleared through 510(k) K142139 on Nov 7, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the SMR Shoulder Liner MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the SMR Shoulder Liner, and is it still sold?
The FDA's device database lists Limacorporate SPA as the company responsible for it (brand SMR Shoulder, model Liner). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall547 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall135 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED158 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented125 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page