Skip to content

TEMA Elbow recallZ-1261-2023 · Limacorporate S.p.A · initiated Jan 25, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Limacorporate S.p.A or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to the incorrect component, that is not cleared to be marketed in the U.S., being included in its elbow prosthesis.

FDA root cause category: Employee error

What the company is doing

On or about 01/25/2023, the firm sent a "VOLUNTARY CORRECTION" Letter via email to customers informing them that the firm has become aware that it incorrectly included low constraint Tema axle small components (not cleared for marketing in U.S.) with its high constraint humeral bodies system. According to the firm's records, 3 low constraint components were implanted between 12/10/2021 to 07/05/2022. Customers are being instructed to: LimaCorporate suggests to Surgeons and Hospitals where an affected device listed in Table 1 was implanted in combination with high constraint humeral bodies components belonging to the product family 1550.15.0XX to yearly follow-up patients with radiographs and varus/valgus range-of-motion evaluation to assess for poly bushing wear. If the follow-up monitoring after 24 months from the implant date shows uneventful and no abnormality, surgeons are invited to do patient follow-up according to the standard protocols, except in case of pain subsequently experienced by the patient. In addition, customers will be instructed by the firm's U.S Regulatory Affairs Associate to post the letter in or near implanted patients' records to continuously remind surgeons and hospitals to monitor their implant patients regarding this Voluntary Correction. For questions or further assistance, contact the firm via email at [email protected].

Recall details

Recall number
Z-1261-2023
Status
Open, Classified
Initiated by the company
Jan 25, 2023
Posted by the FDA
Mar 16, 2023
Recalling firm
Limacorporate S.p.A
Quantity
5 systems
Product code
JDC
FDA submissions
K181362
Distribution
U.S. Distribution to: TX

Product description

TEMA Axle Small, REF 1590.15.010 (additional information is pending receipt of the firm's product label/labeling)

Code information (models, lots, serial numbers)

Product Code: 1590.15.010 UDI-DI Code: 08033390128865 Lot Number: 2115522 Sterilization Lot Number: 2100241

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page