Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08033390167277), received between Oct 3, 2024 and Feb 18, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Material Twisted/Bent1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2026Break

    DURING A SHOULDER SURGERY PERFORMED ON (B)(6) 2026, WHILE REAMING FOR A 19-DEGREE AUGMENT, THE SMR SHAFT ANGLED GLEN REAMER (PRODUCT CODE 9013.75.355, LOT NO. 21BH06Q) ENGAGED WITH A RETRACTOR, RESULTING IN FRACTURE OF THE DISTAL TIP OF THE REAMER SHAFT. THE DETACHED PART FELL INTO PATIENT AND THEN REMOVED. AS THIS WAS A REGIONAL CASE, A DROP SET WAS AVAILABLE ON SITE. THE INSTRUMENT WAS REPLACED, AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER COMPLICATIONS. THE SURGERY WAS DELAYED BY LESS THAN 15 MINUTES. THE PATIENT IS A 63-YEAR-OLD MALE. EVENT HAPPENED IN AUSTRALIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24377891 ↗
  • MalfunctionOct 3, 2024Mechanical Problem; Material Twisted/Bent; Mechanical Problem; Material Twisted/Bent

    BEFORE A SURGERY WITH THE AUGMENTED BASEPLATE STARTED ON (B)(6) 2024, IT WAS NOTICED THAT THE SMR SHAFT ANGLED GLENOID REAMER (PRODUCT CODE 9013.75.355, LOT #22BH0EN) WAS BENT. AS A CONSEQUENCE, AUGMENTED REAMERS COULDN'T BE USED, AND A STANDARD REAMER FOR A STANDARD BASEPLATE WAS USED INSTEAD. THE BEND WAS NOTICED BEFORE SURGERY STARTED, NO SIGNIFICANT EXTENSION OF THE SURGICAL TIME. IT WAS REPORTED THAT THE SURGEON WAS SATISFIED WITH THE STABILITY OF THE FINAL IMPLANTATION. NUMBER OF USES OF THE INSTRUMENT IS UNKNOWN. EVENT HAPPENED IN AUSTRALIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20358700 ↗