PRIMA TT Glenoid 1975.14.615Recalls & reported problems
Limacorporate SPA · PRIMA TT Glenoid - Modular Reverse TT Baseplate Dia. 25 mm; Full Wedge 15° · Class 2 · Implant
All models of PRIMA TT Glenoid (16)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033390244848), received between Aug 1, 2025 and Aug 1, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 29, 2023Class IIOpen, Classified
Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.
Z-0726-2024 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Patient Device Interaction Problem1
Patient problems
- Implant Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 1, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
SHOULDER REVISION SURGERY DUE TO LOOSENING OF THE GLENOID COMPONENT PERFORMED ON (B)(6) 2025. THE ORIGINAL SURGERY TOOK PLACE ON (B)(6), FEW WEEKS LATER, THE PATIENT SHOWED UP IN CLINIC COMPLAINING OF PAIN. SHE DIDN'T DESCRIBE AN INCIDENT. THE X-RAYS REVEALED LOOSE GLENOID COMPONENT. ON (B)(6), THE SURGEON REVISED THE PATIENT WITH BONE GRAFTING AND REPLACING THE GLENOID. SIGNIFICANT BONE LOSS WAS NOTED. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION. THE BASEPLATE INVOLVED IN THE COMPLAINT IS THE FOLLOWING: BASEPLATE D.25MM FULL WEDGE15°, PRODUCT CODE 1975.14.615, LOT N. 2435904, STER. (B)(4). THE PATIENT IS FEMALE, BORN IN 1945. THE EVENT HAPPENED IN THE UNITED STATES.
Patient problems: Implant Pain
MAUDE report 22687144 ↗
FDA clearance and classification
The FDA regulates product code PHX (shoulder prosthesis, reverse configuration) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: Intended to be used to relieve pain and to replace the glenoid and humeral side of the shoulder joint in order to restore mobility in a grossly rotator cuff deficient joint with severe arthropathy or a previous joint replacement with a grossly rotator cuff deficient joint.
Common questions about the PRIMA TT Glenoid 1975.14.615
Has the PRIMA TT Glenoid 1975.14.615 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0726-2024 (Class II), began on Aug 29, 2023: Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the PRIMA TT Glenoid 1975.14.615?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Aug 1, 2025. That report lists device problem “Patient Device Interaction Problem” (1); patient problem “Implant Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PRIMA TT Glenoid 1975.14.615?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PRIMA TT Glenoid 1975.14.615 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the PRIMA TT Glenoid 1975.14.615, and is it still sold?
The FDA's device database lists Limacorporate SPA as the company responsible for it (brand PRIMA TT Glenoid, model 1975.14.615). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Shoulder Prosthesis, Reverse Configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration513 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration435 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration427 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall258 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall202 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration239 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration221 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration124 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page