Skip to content

PRIMA TT Glenoid recallZ-0726-2024 · Limacorporate S.p.A · initiated Aug 29, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Limacorporate S.p.A or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.

FDA root cause category: Device Design

What the company is doing

On 8/29/23, the firm phoned all agents and surgeons to immediately stop surgeries with affected devices. On 10/10/23 the firm emailed recall notices to customers asking them to take the following actions: 1) Surgeons are suggested to perform a short-term patient s monitoring (radiographs at 6 weeks, 3-6 months and 1 year might be considered) and acting accordingly, based on a case-by-case evaluation. 2) Locate, quarantine, and return affected devices. Complete and return the response form to [email protected] 3) Customers with questions can email the firm at [email protected]

Recall details

Recall number
Z-0726-2024
Status
Open, Classified
Initiated by the company
Aug 29, 2023
Posted by the FDA
Jan 12, 2024
Recalling firm
Limacorporate S.p.A
Quantity
179
Product code
MBF
FDA submissions
K222427
Distribution
US: IL, MO, CA, IN, MI, OK

Product description

PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870

Code information (models, lots, serial numbers)

Product Code/UDI-DI: 1975.14.500/08033390244824, 1975.14.800/08033390244893, 1975.14.510/08033390231299, 1975.14.810/08033390244909, 1975.14.615/08033390244848, 1975.14.815/08033390244916, 1975.14.865/08033390244930, 1975.14.620/08033390231312, 1975.14.820/08033390244923, 1975.14.870/08033390244947

Devices named in this recall

10 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page