SMR System 1377.50.020Recalls & reported problems
Limacorporate SPA · SMR System - Metal Back Liner Small · Class 2 · Implant
All models of SMR System (47)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08033390272704), received between Jan 30, 2026 and May 4, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
Patient problems
- Loose body in joint1
- Metal Related Pathology1
- Muscle/Tendon Damage1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 4, 2026Detachment of Device or Device Component
REVISION SURGERY WAS PERFORMED ON (B)(6) 2026 DUE TO DISASSEMBLY OF COMPONENTS AND METALLOSIS. PREVIOUS SURGERY IS UNKNOWN, AS WELL AS COMPLETE PRODUCT INFORMATION OF ORIGINAL IMPLANT. DURING REVISION, THE PRE-EXISTING LINER FOR METAL BACK GLENOID SMALL (PN 1377.50.020) WAS FOUND LOOSE FROM THE SMR METAL-BACK GLENOID SMALL (PN 1375.21.020) AND FLOATING IN THE JOINT, CAUSING METAL ON METAL FRICTION AND THEREFORE SEVERE METALLOSIS. SURGEON REMOVED ALL THE GLENOID COMPONENTS, AS WELL AS THE PRE-EXISTING SMR HUMERAL HEAD Ø50 MM (PN 1322.09.500) AND ADAPTOR TAPER (PN 1330.15.274) AND RE-IMPLANTED A NEW TAPER AND HEAD ON THE HUMERAL SIDE TO RESTORE SHOULDER FUNCTIONALITY WITH A HEMI ARTHROPLASTY. SURGERY WAS COMPLETED AS INTENDED. PATIENT IS MALE, DATE OF BIRTH (B)(6)1960. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Loose body in joint; Metal Related Pathology
MAUDE report 25067102 ↗InjuryJan 30, 2026Adverse Event Without Identified Device or Use Problem
A SHOULDER REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 DUE TO CUFF FAILURE, COMPONENT WEAR, LOSS OF RANGE OF MOTION, AND PAIN. THE COMPLAINT SOURCE ALSO REPORTED THAT CUFF FAILURE CAUSED SUPERIOR MIGRATION OF THE HUMERAL HEAD, WITH SIGNIFICANT WEAR THROUGH THE LINER. THE LINER WAS STILL ENGAGED IN THE METAL-BACK COMPONENT. AS A RESULT, THE SURGEON DECIDED TO REMOVE THE ANATOMIC COMPONENTS AND CONVERT THE IMPLANT TO A REVERSE CONFIGURATION. THE FOLLOWING DEVICES WERE EXPLANTED: · SMR HUMERAL HEAD Ø50 MM (PRODUCT CODE: 1322.09.500, LOT NUMBER: UNKNOWN). · NEUTRAL ADAPTOR TAPER STANDARD (PRODUCT CODE: 1330.15.270, LOT NUMBER: UNKNOWN). · SMR FINNED HUMERAL BODY (PRODUCT CODE: 1350.15.110, LOT NUMBER: UNKNOWN). · LINER F. MET.BACK GLEN. SMALL (PRODUCT CODE: 1377.50.020, LOT NUMBER: UNKNOWN). · METAL-BACK GLENOID SMALL (PRODUCT CODE: 1375.20.020, LOT NUMBER: UNKNOWN). THE INITIAL SHOULDER IMPLANTATION SURGERY WAS PERFORMED ON (B)(6) 2009. THE PATIENT WAS FEMALE AND WAS 88 YEARS OLD AT THE TIME OF THE REVISION SURGERY. EVENT HAPPENED IN AUSTRALIA.
Patient problems: Muscle/Tendon Damage
MAUDE report 24223086 ↗
Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page