Skip to content

SMR System 1360.50.820Recalls & reported problems

Limacorporate SPA · SMR System - Reverse Liner +6mm · Class 2 · Implant

All models of SMR System (45)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08033390273619), received between Feb 25, 2026 and May 18, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K110598); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Device Dislodged or Dislocated2

Patient problems

  • Joint Dislocation2
  • Loose body in joint1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 18, 2026Detachment of Device or Device Component; Device Dislodged or Dislocated

    REVISION SURGERY WAS PERFORMED ON (B)(6) 2026 DUE TO SHOULDER DISLOCATION. PREVIOUS SURGERY OCCURRED ON (B)(6) 2025 WHEN ORIGINAL ANATOMIC PROSTHESIS WAS CONVERSED TO A REVERSE CONFIGURATION DUE TO ROTATOR CUFF FAILURE (PREVIOUSLY REPORTED WITH MFR. 3008021110-2026-00073). PATIENT DISLOCATED AND THE SURGEON DECIDED TO PERFORM A REVISION SURGERY SINCE THEY COULD NOT REDUCE THE DISLOCATION OF THE REVERSE IN OTHER WAY. DURING REVISION, THE PRE-EXISTING SMR SMALL/STD CONNECTOR +2MM (PN 1374.15.322, LOT 2514593, STERILIZATION (B)(4), GLENOSPHERE Ø 36MM (PN 1374.09.111,LOT 2515422, STERILIZATION 2500135) AND SMR REVERSE LINER + 6 MM (PN 1360.50.820, LOT 23AT40J, STERILIZATION 2400057) WERE REMOVED AND REPLACED WITH NEW CONNECTOR, ECCENTRICAL GLENOSPHERE DIA40MM (PN 1376.09.041) AND REVERSE LINER RETENTIVE +3MM (PN 1365.50.816). REVERSE LINER WAS FOUND DISSOCIATED FROM HUMERAL BODY DURING REVISION. SURGERY WAS COMPLETED AS INTENDED. PATIENT IS MALE, DATE OF BIRTH (B)(6) 1951. THE EVENT OCCURRED IN THE UNITED STATES.

    Patient problems: Joint Dislocation

    MAUDE report 25200759 ↗
  • InjuryMay 4, 2026Device Dislodged or Dislocated

    REVISION SURGERY WAS PERFORMED ON (B)(6) 2026 DUE TO DISLOCATION. PATIENT CLINICAL HISTORY SUMMARY IS THE FOLLOWING: (B)(6) 2025: ORIGINAL SURGERY WITH SMR HUMERAL COMPONENTS (FINNED STEM DIA. 19 MM L. 80MM PN 1304.15.190; 140° HUMERAL BODY REVERSE PN 1352.15.011; EXTENSION FOR HUMERAL REVERSE BODY PN 1352.15.001; SMR REVERSE LINER +3MM PN 1360.50.815) AND DJO ALTIVATE GLENOID COMPONENTS. (B)(6) 2026: REVISION SURGERY FOR HUMERAL STEM LOOSENING (PREVIOUSLY REPORTED WITH REFERENCE MFR. 3008021110-2026-00183) WAS PERFORMED TO REPLACE ALL EXISTING COMPONENTS WITH NEW ONES (ON THE HUMERAL SIDE THE ABOVE MENTIONED COMPONENTS WERE REPLACED WITH FINNED STEM DIA. 20 MM L. 80MM PN 1304.15.200; 140° HUMERAL BODY REVERSE PN 1352.15.011, LOT 2535453, STERILIZATION (B)(6); AND SMR REVERSE LINER +6MM PN 1360.50.820, LOT 25AT1AB, STERILIZATION (B)(6)). (B)(6) 2026: REVISION SURGERY FOR SHOULDER DISLOCATION (HEREBY REPORTED) WAS PERFORMED REPLACING THE ABOVE MENTIONED 140° HUMERAL BODY REVERSE AND SMR REVERSE LINER +6MM WITH NEW EXTENSION FOR HUMERAL REVERSE BODY (PN 1352.15.001), 140° HUMERAL BODY FINNED REVERSE (PN 1352.15.051) AND SMR REVERSE LINER STANDARD (PN 1360.50.810). (B)(6) 2026: REVISI

    Patient problems: Joint Dislocation

    MAUDE report 25067089 ↗
  • InjuryFeb 25, 2026Detachment of Device or Device Component

    REVISION SURGERY WAS PERFORMED ON (B)(6) 2026 DUE TO DISASSEMBLY OF COMPONENTS. PREVIOUS SURGERY WAS PERFORMED ON (B)(6) 2025 AND FOLLOWING COMPONENTS WERE IMPLANTED: SMR REVERSE LINER + 6 MM (PART NUMBER: 1360.50.820, LOT. 24AT5P8, STERILIZATION N. 2500009). SMR CONNECTOR SMALL STD (PART NUMBER: 1374.15.310, LOT: 2500907, STERILIZATION N. 2500027). SMR GLENOSPHERE Ø 36MM (PART NUMBER: 1374.09.111, LOT: 2506966, STERILIZATION N. 2500066). SMR METAL-BACK GLENOID SMALL (PART NUMBER: 1375.21.020, LOT: 2435106, STERILIZATION N. 2500053). SMR CEMENTLESS FINNED STEM (PART NUMBER: 1304.15.210, LOT: 2102161, STERILIZATION N. 2100078). SOME MONTHS AFTER ORIGINAL SURGERY THE ABOVE-MENTIONED REVERSE POLY LINER WAS FOUND LOOSE. DURING REVISION SURGERY LINER WAS EXPLANTED AND REPLACED WITH A NEW ONE FROM SAME LOT. SURGERY WAS COMPLETED AS INTENDED. PATIENT IS A MALE, DATE OF BIRTH ON (B)(6) 1940. THE EVENT OCCURRED IN UNITES STATES.

    Patient problems: Loose body in joint

    MAUDE report 24442723 ↗

FDA clearance and classification

Cleared through 510(k) K110598 on Aug 18, 2011 (decision: Substantially Equivalent). Submitted by Lima Corporate S.P.A. as “Smr Reverse Shoulder System”. The FDA review took 169 days. FDA record ↗

The FDA regulates product code KWS (prosthesis, shoulder, semi-constrained, metal/polymer cemented) under 21 CFR 888.3660 as a Class II device. 21 CFR 888.3660 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K110598. One submission often covers many models, so check the product and lot details in each record.

  • Z-1711-2016 · Apr 8, 2016 · Terminated
    SMR glenosphere impactor/extractor; Used to implant the glenosphere in the SMR Reverse Shoulder replacement.

Common questions about the SMR System 1360.50.820

Has the SMR System 1360.50.820 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08033390273619), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the SMR System 1360.50.820?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Feb 25, 2026 and May 18, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (2) and “Device Dislodged or Dislocated” (2); the most common patient problems are “Joint Dislocation” (2) and “Loose body in joint” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SMR System 1360.50.820?

None of the reports are classified as deaths, 3 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SMR System 1360.50.820 FDA approved?

It was cleared through 510(k) K110598 on Aug 18, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the SMR System 1360.50.820 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the SMR System 1360.50.820, and is it still sold?

The FDA's device database lists Limacorporate SPA as the company responsible for it (brand SMR System, model 1360.50.820). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices

All Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page