Quantum Perfusion Dual Lumen Cannula 27 DL27F-V0Recalls & reported problems
Spectrum Medical Srl · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08051160300808), received between Jan 20, 2023 and Sep 29, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Crack” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Crack1
- Fracture1
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Air Embolism1
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
ON FRIDAY EVENING, CUSTOMER CALLED TO REPORT THAT EARLIER THAT DAY, THEY PLACED A PATIENT ON ECMO. THEY SELECTED THE QUANTUM 27FR DLC, BUT THEY WERE UNABLE TO FLOW. THE CUSTOMER REPORTED THAT THE CANNULA DID PRIME BACK WITH BLOOD, HOWEVER, WHEN THEY WENT TO CONNECT THE CANNULA TO THE CICUIT, THEY WERE UNABLE TO INITIATE AND ACHIEVE ANY FLOW. CUSTOMER REPORTED THAT THE ISSUE APPEARED TO BE ON THE DRAINAGE SIDE, AS THEY WERE NOT GETTING BLOOD BACK TO THE PUMP WHEN INITIATING AT 1500 RPMS OR WHEN SLOWLY INCREASING RPMS. DR. REDDY, THE CANNULATING SURGEON, HAD TO REMOVE THIS CANNULA, AND OVER WIRE TO EXCHANGE FOR A DIFFERENT ONE. THEY OPTED TO REPLACE OUR CANNULA WITH THE 29FR PROTEK DUO AND WERE ABLE TO SUCCESSFULLY DRAIN & FLOW TO THE PATIENT. THE CENTRIMAG CONSOLE AND PRECONNECTED PACK REMAINED THE SAME, THE ONLY CHANGE WAS THE CANNULA. THIS CUSTOMER DOES NOT MONITOR PRESSURES ON THEIR CENTRIMAG PACK, SO NONE TO REPORT. THE CUSTOMER DID SAVE THE CANNULA AND WOULD LIKE TO SEND IT BACK FOR INVESTIGATION. THE COORDINATOR, PERFUSIONIST AND SURGEON ALL VISUALLY INSPECTED THE CANNULA, BUT DID NOT SEE ANYTHING ABNORMAL.
Patient problems: Insufficient Information
MAUDE report 23166422 ↗MalfunctionMay 19, 2023Fracture; Fracture
RECEIVED A CALL THAT THEY EXPERIENCE A FRACTURE TO A 27FR DUAL LUMEN CANNULA. CANNULA HAD BEEN IN FOR 7 DAYS. PATIENT ALSO HAS A HEARTMATE 3 LVAD. WITH FRACTURE OF THE CANNULA, THE SYSTEM ENTRAINED AIR. CANNULA WAS PULLED AND PATIENT PLACED ON DOBUTAMINE FOR RIGHT SIDE SUPPORT. PATIENT ADEQUATELY SUPPORTED.
Patient problems: Air Embolism; No Clinical Signs, Symptoms or Conditions
MAUDE report 16960947 ↗MalfunctionJan 20, 2023Crack; Crack
27FR DUAL LUMEN CANNULA NOTED TO BE LEAKING BELOW BIFURCATION OF CANNULA. PATIENT HAD BEEN ON FOR ALMOST A MONTH. CANNULA WAS REPLACED WITH ANOTHER 27FR DUAL LUMEN CANNULA WITHOUT ISSUE. POSITIVE PRESSURE MEASURING 350 ON LEAKING CANNULA. NO SUTURE WAS AT THE AREA WHERE THE CANNULA STARTING LEAKING, PERFUSIONIST SAID IT LOOKED MORE LIKE A POINT OF FLEXION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16213214 ↗
Other Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY2V10R33Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page